Investigation of dose reduction in neonatal radiography using specially designed phantoms and LiF:Mg,Cu,P TLDs

被引:19
作者
Duggan, L
Warren-Forward, H
Smith, T
Kron, T
机构
[1] Newcastle Mater Misericordiae Hosp, Dept Radiat Oncol, Waratah, NSW 2298, Australia
[2] Univ Newcastle, Sch Math & Phys Sci Phys, Callaghan, NSW 2308, Australia
[3] Univ Newcastle, Sch Hlth Sci Med Radiat Sci, Callaghan, NSW 2308, Australia
关键词
D O I
10.1259/bjr/79291075
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Many departments still do not use recommended radiographic parameters to X-ray neonates. Direct, accurate dose measurements of individual examinations may assist a department in justifying technique modifications that provide a substantial dose reduction without a significant loss of image quality. The aim of this study was to investigate dose reduction techniques for neonates in the intensive care unit. Alterations in beam energy (kVp and filtration) and collimation were investigated using specially designed phantoms mimicking a 700 g and 2000 g neonate, and ultrasensitive LiF:Mg,Cu,P thermoluminescence dosimeters (TLDs). Differences in entrance surface dose (ESD) and dose at depth (3 cm or 5 cm) were compared for two, overlapping fields centred individually on the chest and abdomen (Technique 1) and one large chest-abdomen field (Technique 2 or babygram). The large phantom was irradiated at 54 kVp, 60 kVp and 70 kVp without additional filtration and at 66 kVp and 70 kVp with a rare-earth hafnium filter. Focus-film distance (FFD) and mAs were adjusted to maintain optical density (OD) on each radiograph. The baseline dose at 54 kVp and 100 cm FFD was (46+/-2) pGy. Increasing the tube potential from 54 kVp to 70 kVp without additional filtration reduced the ESD by 27%. However, the addition of a 0.05 mm hafnium filter at 66 kVp further reduced the radiation dose by 13%, to produce an ESD of (28+/-2) pGy. All contrast details were observable at 66 kVp with hafnium filtration. Technique 1 may lead to an increase in effective dose due to field overlap, which diverges at depth, and increased scatter at the periphery of the fields.
引用
收藏
页码:232 / 237
页数:6
相关论文
共 20 条
[1]  
Adams J, 1990, Radiogr Today, V56, P11
[2]   THE RISK OF X-RAY EXAMINATIONS OF THE LUNGS IN NEONATES [J].
ARROE, M .
ACTA PAEDIATRICA SCANDINAVICA, 1991, 80 (05) :489-493
[3]   REEVALUATION OF X-RAY ATOMIC ENERGY LEVELS [J].
BEARDEN, JA ;
BURR, AF .
REVIEWS OF MODERN PHYSICS, 1967, 39 (01) :125-&
[4]  
*COLL RAD I PHYS S, 1992, NAT PROT PAT DOS MEA
[5]   DOSE REDUCTION BY THE USE OF ERBIUM FILTRATION IN A GENERAL RADIOGRAPHIC ROOM [J].
CRANAGE, RW ;
HOWARD, CJ ;
WELSH, AD .
BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (771) :232-237
[6]   Suitability of LiF:Mg,Cu,P and Al2O3:C for low dose measurements in medical imaging [J].
Duggan, L ;
Sathiakumar, C ;
Warren-Forward, H ;
Symonds, M ;
McConnell, P ;
Smith, T ;
Kron, T .
RADIATION PROTECTION DOSIMETRY, 1999, 85 (1-4) :425-428
[7]  
*ENV PROT AG, 1999, RAD GUID 2, V6
[8]   RADIATION-DOSE TO NEONATES ON A SPECIAL CARE BABY UNIT [J].
FAULKNER, K ;
BARRY, JL ;
SMALLEY, P .
BRITISH JOURNAL OF RADIOLOGY, 1989, 62 (735) :230-233
[9]   Advanced LiF technology for the assessment of patient exposure in diagnostic radiology [J].
Fill, UA ;
Regulla, DF .
RADIATION PROTECTION DOSIMETRY, 1998, 80 (1-3) :225-229
[10]   DOSE REDUCTION IN PEDIATRIC RADIOLOGY USING RARE-EARTH FILTRATION [J].
FLEAY, RF ;
FOX, RA ;
SPRAGUE, PL ;
ADAMS, JP .
PEDIATRIC RADIOLOGY, 1984, 14 (05) :332-334