Two-Site Evaluation of the Repeatability and Precision of an Automated Dual-Column Hydrogen/Deuterium Exchange Mass Spectrometry Platform
被引:24
作者:
Cummins, David J.
[1
]
Espada, Alfonso
[2
]
Novick, Scott J.
[3
]
Molina-Martin, Manuel
[2
]
Stites, Ryan E.
[1
]
Felix Espinosa, Juan
[2
]
Broughton, Howard
[2
]
Goswami, Devrishi
[3
]
Pascal, Bruce D.
[4
]
Dodge, Jeffrey A.
[1
]
Chalmers, Michael J.
[1
]
Griffin, Patrick R.
[3
]
机构:
[1] Eli Lilly & Co, Lilly Res Labs, Lilly Corp Ctr, Indianapolis, IN 46285 USA
[2] Ctr Invest Lilly SA, Analyt Technol Dept, Ave Ind 30, Alcobendas 28108, Spain
[3] Scripps Res Inst, Dept Mol Therapeut, 130 Scripps Way, Jupiter, FL 33458 USA
[4] Scripps Res Inst, Informat Core, 130 Scripps Way, Jupiter, FL 33458 USA
基金:
美国国家卫生研究院;
关键词:
PROTEIN-LIGAND INTERACTIONS;
HYDROGEN-EXCHANGE;
BINDING DOMAIN;
STRUCTURAL BASIS;
RECEPTOR-ALPHA;
HDX;
MS;
DYNAMICS;
AGONIST;
COMPLEX;
D O I:
10.1021/acs.analchem.6b01650
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
Hydrogen/deuterium exchange coupled with mass spectrometry (HDX-MS) is an information-rich biophysical method for the characterization of protein dynamics. Successful applications of differential HDX-MS include the characterization of protein ligand binding. A single differential HDX-MS data set (protein +/- ligand) is often comprised of more than 40 individual HDX-MS experiments. To eliminate laborious manual processing of samples, and to minimize random and gross errors, automated systems for HDX-MS analysis have become routine in many laboratories. However, an automated system, while less prone to random errors introduced by human operators, may have systematic errors that go unnoticed without proper detection. Although the application of automated (and manual) HDX-MS has become common, there are only a handful of studies reporting the systematic evaluation of the performance of HDX-MS experiments, and no reports have been published describing a cross-site comparison of HDX-MS experiments. Here, we describe an automated HDX-MS platform that operates with a parallel, two-trap, two-column configuration that has been installed in two remote laboratories. To understand the performance of the system both within and between laboratories, we have designed and completed a test-retest repeatability study for differential HDX-MS experiments implemented at each of two laboratories, one in Florida and the other in Spain. This study provided sufficient data to do both within and between laboratory variability assessments. Initial results revealed a systematic run-order effect within one of the two systems. Therefore, the study was repeated, and this time the conclusion was that the experimental conditions were successfully replicated with minimal systematic error.
引用
收藏
页码:6607 / 6614
页数:8