Radiation doses in interventional radiology procedures: The RAD-IR Study - Part I: Overall measures of dose

被引:309
作者
Miller, DL
Balter, S
Cole, PE
Lu, HT
Schueler, BA
Geisinger, M
Berenstein, A
Albert, R
Georgia, JD
Noonan, PT
Cardella, JF
George, JS
Russell, EJ
Malisch, TW
Vogelzang, RL
Miller, GL
Anderson, J
机构
[1] Natl Naval Med Ctr, Dept Radiol, Bethesda, MD 20889 USA
[2] Uniformed Serv Univ Hlth Sci, F Edward Hebert Sch Med, Dept Radiol & Nucl Med, Bethesda, MD 20814 USA
[3] Natl Canc Inst, Med Oncol Clin Res Unit, Canc Res Ctr, Bethesda, MD USA
[4] US FDA, Ctr Devices & Radiol Hlth, Rockville, MD 20857 USA
[5] Lenox Hill Hosp, Dept Med, New York, NY 10021 USA
[6] Beth Israel Deaconess Med Ctr, Hyman Newman Inst Neurol & Neurosurg, Ctr Endovasc Surg, New York, NY 10003 USA
[7] SUNY Upstate Med Univ, Dept Radiol, Syracuse, NY USA
[8] Yale Univ, Sch Med, Dept Diagnost Radiol, New Haven, CT 06510 USA
[9] Mayo Clin, Dept Radiol, Rochester, MN USA
[10] Cleveland Clin Fdn, Dept Radiol, Cleveland, OH 44195 USA
[11] Northwestern Univ, Feinberg Sch Med, Northwestern Mem Hosp, Dept Radiol, Chicago, IL 60611 USA
[12] Univ Texas, SW Med Ctr, Dept Radiol, Dallas, TX 75230 USA
关键词
D O I
10.1097/01.RVI.0000079980.80153.4B
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To determine patient radiation doses for interventional radiology and neuroradiology procedures, to identify procedures associated with higher radiation doses, and to determine the effects of various parameters on patient doses. MATERIALS AND METHODS: A prospective observational study was performed at seven academic medical centers. Each site contributed demographic and radiation dose data for subjects undergoing specific procedures in fluoroscopic suites equipped with built-in cumulative dose (CD) and dose-area-product (DAP) measurement capability compliant with International Electrotechnical Commission standard 60601-2-43. The accuracy of the dosimetry was confirmed by comprehensive measurements and by frequent consistency checks performed over the course of the study. RESULTS: Data were collected on 2,142 instances of interventional radiology procedures, 48 comprehensive physics evaluations, and 581 periodic consistency checks from the 12 fluoroscopic units in the study. There were wide variations in dose and statistically significant differences in fluoroscopy time, number of images, DAP, and CD for different instances of the same procedure, depending on the nature of the lesion, its anatomic location, and the complexity of the procedure. For the 2,142 instances, observed CD and DAP correlate well overall (r = 0.83, P < .000001), but correlation in individual instances is poor. The same is true for the correlation between fluoroscopy time and CD (r = 0.79, P < .000001). The correlation between fluoroscopy time and DAP (r = 0.60, P < .000001) is not as good. In 6% of instances (128 of 2,142), which were principally embolization procedures, transjugular intrahepatic portosystemic shunt (TIPS) procedures, and renal/visceral artery stent placements, CD was greater than 5 Gy. CONCLUSIONS: Most procedures studied can result in clinically significant radiation dose to the patient, even when performed by trained operators with use of dose-reducing technology and modem fluoroscopic equipment. Embolization procedures, TIPS creation, and renal/visceral artery stent placement are associated with a substantial likelihood of clinically significant patient dose. At minimum, patient dose data should be recorded in the medical record for these three types of procedures. These data should include indicators of the risk of deterministic effects as well as the risk of stochastic effects.
引用
收藏
页码:711 / 727
页数:17
相关论文
共 52 条
  • [1] Uterine arterial embolization: Factors influencing patient radiation exposure
    Andrews, RT
    Brown, PH
    [J]. RADIOLOGY, 2000, 217 (03) : 713 - 722
  • [2] [Anonymous], BRIT MED J
  • [3] Evaluation of vascular and interventional procedures with time-action analysis: A pilot study
    Bakker, NH
    Tanase, D
    Reekers, JA
    Grimbergen, CA
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2002, 13 (05) : 483 - 488
  • [4] Balter S., 2001, Interventional fluoroscopy. Physics, technology
  • [5] BERGERON P, 1994, AM J NEURORADIOL, V15, P1809
  • [6] ASSESSMENT OF ORGAN RADIATION-DOSES AND ASSOCIATED RISK FOR DIGITAL BIFEMORAL ARTERIOGRAPHY
    CASTELLANO, IA
    MCNEILL, JG
    THORP, NC
    DANCE, DR
    RAPHAEL, MJ
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (809) : 502 - 507
  • [7] Estimation of effective dose in some digital angiographic and interventional procedures
    Cruces, RR
    García-Granados, J
    Romero, FJD
    Armas, JH
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (841) : 42 - 47
  • [8] Deterministic effects in interventional radiology
    Faulkner, K
    Vañó, E
    [J]. RADIATION PROTECTION DOSIMETRY, 2001, 94 (1-2) : 95 - 98
  • [9] Patient dosimetry measurement methods
    Faulkner, K
    Broadhead, DA
    Harrison, RM
    [J]. APPLIED RADIATION AND ISOTOPES, 1999, 50 (01) : 113 - 123
  • [10] Dose displays and record keeping
    Faulkner, K
    [J]. RADIATION PROTECTION DOSIMETRY, 2001, 94 (1-2) : 143 - 145