Tympanic membrane temperature as a measure of core temperature

被引:89
作者
Childs, C [1 ]
Harrison, R [1 ]
Hodkinson, C [1 ]
机构
[1] Booth Hall Childrens Hosp, Reg Paediat Burns Unit, Manchester M9 7AA, Lancs, England
关键词
tympanic membrane temperature; measurement error; temperature; burns;
D O I
10.1136/adc.80.3.262
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Background-Ear thermometers are becoming popular as a method for measuring deep body (core) temperature. Aim-To determine the variability of a single user's tympanic membrane tear) temperature measurements. Subjects-Forty two, afebrile, healthy children, and 20 febrile children with acute burns. Results-In afebrile children measurements made in both ears (and within just a few minutes of each other) differed by as much as 0.6 degrees C. Operator measurement error, s(w) of three consecutive measurements, in the same ear, was 0.13 degrees C. In the group of febrile, burned children, core temperature was measured hourly at a number of sites (ear, rectum, axilla, bladder). A peak in core temperature occurred approximately 10-12 hours after the burn. Measurement error was calculated in 14 febrile, burned children with a peak temperature in excess of 38 degrees C. For the left ear, measurement error was 0.19 degrees C and for the right ear, 0.11 degrees C. In the febrile children agreement between the ears was poor. The Limits of agreement were 0.4 degrees C to -0.8 degrees C. It was not possible to predict the occasions when the temperature differences between the ears would be large or small. Conclusions-The measurement error of one recording from the next is probably acceptable at about 0.1 to 0.2 degrees C. To limit the variations in temperature of one ear to the other, measurements should be restricted to one of the ears whenever possible and the same ear used throughout the temperature monitoring period. Nurses and parents should take more than one temperature reading from the same ear whenever possible.
引用
收藏
页码:262 / 266
页数:5
相关论文
共 29 条
  • [1] [Anonymous], ENERGY BALANCE TEMPE
  • [2] HEAT REGULATION - HOMEOSTASIS OF CENTRAL TEMPERATURE IN MAN
    BENZINGER, TH
    [J]. PHYSIOLOGICAL REVIEWS, 1969, 49 (04) : 671 - +
  • [3] Benzinger TH, 1961, SCI AM, V203, P134
  • [4] Measurement error
    Bland, JM
    Altman, DG
    [J]. BRITISH MEDICAL JOURNAL, 1996, 312 (7047) : 1654 - 1654
  • [5] BLAND JM, 1986, BRIT MED J, V1, P307
  • [6] EVALUATION OF A TYMPANIC THERMOMETER IN CHILDREN
    BROGAN, P
    CHILDS, C
    PHILLIPS, BM
    MOULTON, C
    [J]. LANCET, 1993, 342 (8883) : 1364 - 1365
  • [7] FEVER IN BURNED CHILDREN
    CHILDS, C
    [J]. BURNS, 1988, 14 (01) : 1 - 6
  • [8] ACUTE CHANGES IN OXYGEN-CONSUMPTION AND BODY-TEMPERATURE AFTER BURN INJURY
    CHILDS, C
    LITTLE, RA
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD, 1994, 71 (01) : 31 - 34
  • [9] CLARKE RP, 1985, MAN HIS THERMAL ENV, P36
  • [10] THE GENIUS INFRARED TYMPANIC THERMOMETER - AN EVALUATION FOR CLINICAL USE
    EDGE, G
    MORGAN, M
    [J]. ANAESTHESIA, 1993, 48 (07) : 604 - 607