An In Vitro Biomechanical Comparison of Single-Rod, Dual-Rod, and Dual-Rod With Transverse Connector in Anterior Thoracolumbar Instrumentation

被引:9
|
作者
Reddy, Chandan G. [1 ]
Magnetta, Michael [2 ]
Dahdaleh, Nader S. [1 ]
Demmer, Matthew [2 ]
Iyamah, Kingsley Abode [1 ]
Lim, Tae-Hong [2 ]
Torner, James C. [3 ]
Hitchon, Patrick W. [1 ]
机构
[1] Univ Iowa, Dept Neurosurg, Iowa City, IA USA
[2] Univ Iowa, Dept Biomed Engn, Iowa City, IA USA
[3] Univ Iowa, Dept Epidemiol, Iowa City, IA USA
关键词
Anterior fixation; Biomechanics; Corpectomy; Thoracolumbar spine; BURST FRACTURES; LUMBAR SPINE; CADAVERIC SPINE; PEDICLE SCREW; POSTERIOR; DECOMPRESSION; FUSION; STABILIZATION; SCOLIOSIS; FIXATION;
D O I
10.1227/NEU.0b013e31823cf254
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND: After thoracolumbar corpectomy, standard anterolateral instrumentation may consist of dual rods with cross-connectors. However, when the vertebral bodies are small or involved with disease, only 1 rod may be possible. OBJECTIVE: To compare the biomechanics of an in vitro L1 corpectomy model using 1 rod, 2 rods, or 2 rods with 2 cross-connectors. METHODS: Eight fresh frozen human cadaveric spines were potted from T9 to L3. Pure moments of 1.5, 3, and 4.5 Nm were applied, and the motion of the spine was measured using 3 infrared cameras. Loads were applied in flexion and extension, right and left lateral bending, and right and left axial rotation. Each spine was first tested in the intact state. After performing an L1 corpectomy and replacement with a carbon fiber reinforced polymer cage, 3 constructs were tested: single rod (1R), dual rod (2R), and dual rod with 2 transverse connectors (CC). RESULTS: Analysis of variance suggests significant main effects of load (P < .0001), axis (P = .022), construct (P = .0019), and individual spine (P < .0001). Overall, the single-rod construct is significantly less rigid than the intact spine in axial rotation. There is no significant difference between the intact spine and either the dual-rod construct or the dual-rod cross-connector construct. CONCLUSION: In our in vitro model of anterior spinal stabilization after corpectomy and grafting, a single-rod construct is significantly less rigid than the intact spine. Addition of a second rod returns the rigidity of the spine to the intact state. A dual-rod cross-connector construct is significantly more rigid than a single-rod construct.
引用
收藏
页码:1017 / 1023
页数:7
相关论文
共 50 条
  • [41] Sawbones comparison of single-rod and double-rod derotation techniques for adolescent idiopathic scoliosis
    Hassanzadeh, Hamid
    Stein, Benjamin E.
    Neubauer, Philip R.
    Logan, Catherine A.
    Kebaish, Khaled M.
    Ain, Michael C.
    CURRENT ORTHOPAEDIC PRACTICE, 2014, 25 (03): : 257 - 260
  • [42] Analytical method for design a TEM00 mode resonator of a dual-rod Nd:YAG laser with full birefringence compensation
    Hua, RZ
    Wada, S
    Tashiro, H
    OPTICS COMMUNICATIONS, 2004, 232 (1-6) : 333 - 341
  • [43] Single screw-rod anterior instrumentation for thoracolumbar burst fractures with incomplete neurological deficit
    Sharma, Siddhartha
    Singh, Dara
    Singh, Manjeet
    Kohli, Arvind
    Singh, Gurjit
    Arora, Mohit
    JOURNAL OF ORTHOPAEDIC SURGERY, 2013, 21 (01): : 71 - 76
  • [44] LD double-end pumped dual-rod acousto-optic Q-switched Tm:LuAG laser
    Fan, Z.
    Wen, Y.
    Shi, X.
    Wu, C. T.
    Chen, X. Y.
    Jiang, Y.
    Dai, T. Y.
    INFRARED PHYSICS & TECHNOLOGY, 2019, 102
  • [45] Dual and Single Memory Rod Construct Comparison in an Animal Study
    Newton, Peter O.
    Farnsworth, Christine L.
    Upasani, Vidyadhar V.
    Chambers, Reid
    Yoon, Seung Hwan
    Firkins, Paul
    SPINE, 2011, 36 (14) : E904 - E913
  • [46] 36.5 W high beam quality 532 nm green laser based on dual-rod AO Q-switched resonator
    Liu, Q.
    Chen, H.
    Yan, X.
    Gong, M.
    OPTICS COMMUNICATIONS, 2011, 284 (13) : 3383 - 3386
  • [47] High repetition rate dual-rod acousto-optics Q-switched composite Nd:YVO4 laser
    Yan, Xingpeng
    Liu, Qiang
    Fu, Xing
    Chen, Hailong
    Gong, Mali
    Wang, Dongsheng
    OPTICS EXPRESS, 2009, 17 (24): : 21956 - 21968
  • [48] Biomechanical comparison of different anchors (foundations) for the pediatric dual growing rod technique
    Mahar, Andrew Todd
    Bagheri, Ramin
    Oka, Richard
    Kostial, Patricia
    Akbarnia, Behrooz A.
    SPINE JOURNAL, 2008, 8 (06): : 933 - 939
  • [49] Anterior thoracolumbar instrumentation: Stiffness and load sharing characteristics of plate and rod systems
    Brodke, DS
    Gollogly, S
    Bachus, KN
    Mohr, RA
    Nguyen, BKN
    SPINE, 2003, 28 (16) : 1794 - 1801
  • [50] Anterior single rod instrumentation for thoracolumbar adolescent idiopathic scoliosis with and without the use of structural interbody support
    Lowe, TG
    Alongi, PR
    Smith, DAB
    O'Brien, MF
    Mitchell, SL
    Pinteric, RJ
    SPINE, 2003, 28 (19) : 2232 - 2241