Retrieval of a leaflet escaped in a Tri-technologies bileaflet mechanical prosthetic valve

被引:13
作者
Cianciulli, Tomas F. [1 ,2 ]
Lax, Jorge A. [1 ,2 ]
Saccheri, Maria C. [1 ,2 ]
Guidoin, Robert [3 ,4 ]
Salvado, Cesar M. [1 ,2 ]
Fernandez, Adrian J. [1 ,2 ]
Prezioso, Horacio A. [1 ,2 ]
机构
[1] Hosp Gobierno, Dept Cardiol, Echocardiog Lab, Buenos Aires, DF, Argentina
[2] Hosp Gobierno, Dept Cardiovasc Surg, Buenos Aires, DF, Argentina
[3] Univ Laval, Dept Surg, Quebec City, PQ, Canada
[4] St Francis Assisi Hosp, Quebec Biomat Inst, Quebec City, PQ, Canada
来源
EUROPEAN JOURNAL OF ECHOCARDIOGRAPHY | 2008年 / 9卷 / 01期
关键词
leaflet escape; acute prosthetic valve dysfunction; tri-technologies; embolization; ultrasonography; computed tomography;
D O I
10.1016/j.euje.2006.08.018
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The escape of the prosthetic heart valve disc is one of the causes of prosthetic dysfunction that requires emergency surgery. The removal of the embolized disc should be carried out because of the risk of a progressive extrusion on the aortic wall. Several imaging techniques can be used for the detection of the missing disc localization. In this report we describe a 32-year-old man who underwent mitral. valve replacement with a Tri-technologies bileaflet valve three years ago, and was admitted in cardiogenic shock. Transesophageal echocardiography showed acute-onset massive mitral. regurgitation. The patient underwent emergency replacement of the prosthetic valve. Only one of the two leaflets remained in the removed prosthetic valve. The missing leaflet could not be found within the cardiac cavity. The abdominal fluoroscopic study and plain radiography were unable to detect the escaped leaflet. The abdominal computed tomography scan and the ultrasound showed the escaped leaflet in the terminal portion of the aortic bifurcation. To retrieve the embolized disc laparotomy and aortotomy were performed three months later. The escaped leaflet shows a fracture of one of the pivot systems caused by structural failure. This kind of failure mode is usually the result of high stress concentration.
引用
收藏
页码:65 / 68
页数:4
相关论文
共 4 条
  • [1] Leaflet escape in a new bileaflet mechanical valve - TRI technologies
    Bottio, T
    Casarotto, D
    Thiene, G
    Caprili, L
    Angelini, A
    Gerosa, G
    [J]. CIRCULATION, 2003, 107 (18) : 2303 - 2306
  • [2] Cianciulli TF, 2005, J HEART VALVE DIS, V14, P664
  • [3] CIANCIULLI TF, IN PRESS EUR J ECHOC
  • [4] Dikmengil M, 2004, J HEART VALVE DIS, V13, P638