Technical Note: Unified imaging and robotic couch quality assurance

被引:6
|
作者
Cook, Molly C. [1 ]
Roper, Justin [2 ]
Elder, Eric S. [2 ]
Schreibmann, Eduard [2 ]
机构
[1] Georgia Inst Technol, Med Phys Program, 770 State St, Atlanta, GA 30332 USA
[2] Emory Univ, Winship Canc Inst, Dept Radiat Oncol, 1365-C Clifton Rd NE, Atlanta, GA 30322 USA
关键词
quality assurance; image guided radiation therapy; robotic couch; 6D couch; image registration; ACCELERATOR; RADIOSURGERY; RADIOTHERAPY; ISOCENTER; ROTATION;
D O I
10.1118/1.4960369
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To introduce a simplified quality assurance (QA) procedure that integrates tests for the linac's imaging components and the robotic couch. Current QA procedures for evaluating the alignment of the imaging system and linac require careful positioning of a phantom at isocenter before image acquisition and analysis. A complementary procedure for the robotic couch requires an initial displacement of the phantom and then evaluates the accuracy of repositioning the phantom at isocenter. We propose a two-in-one procedure that introduces a custom software module and incorporates both checks into one motion for increased efficiency. Methods: The phantom was manually set with random translational and rotational shifts, imaged with the in-room imaging system, and then registered to the isocenter using a custom software module. The software measured positioning accuracy by comparing the location of the repositioned phantom with a CAD model of the phantom at isocenter, which is physically verified using the MV port graticule. Repeatability of the custom software was tested by an assessment of internal marker location extraction on a series of scans taken over differing kV and CBCT acquisition parameters. Results: The proposed method was able to correctly position the phantom at isocenter within acceptable 1 mm and 1 degrees SRS tolerances, verified by both physical inspection and the custom software. Residual errors for mechanical accuracy were 0.26 mm vertically, 0.21 mm longitudinally, 0.55 mm laterally, 0.21 degrees in pitch, 0.1 degrees in roll, and 0.67 degrees in yaw. The software module was shown to be robust across various scan acquisition parameters, detecting markers within 0.15 mm translationally in kV acquisitions and within 0.5 mm translationally and 0.3 degrees rotationally across CBCT acquisitions with significant variations in voxel size. Agreement with vendor registration methods was well within 0.5 mm; differences were not statistically significant. Conclusions: As compared to the current two-step approach, the proposed QA procedure streamlines the workflow, accounts for rotational errors in imaging alignment, and simulates a broad range of variations in setup errors seen in clinical practice. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:5080 / 5083
页数:4
相关论文
共 50 条
  • [41] Automated quality assurance in mammographic Imaging
    Blot, L
    Zwiggelaar, R
    Hobulinka, M
    Davis, A
    IWDM 2000: 5TH INTERNATIONAL WORKSHOP ON DIGITAL MAMMOGRAPHY, 2001, : 641 - 647
  • [42] Quality Assurance Phantoms in Diagnostic Imaging
    Rubert, N.
    MEDICAL PHYSICS, 2017, 44 (06) : 3230 - 3230
  • [43] Quality assurance of imaging for stereotactic treatments
    Novotny, J
    Novotny, JJ
    RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S95 - S95
  • [44] Implementation of a Daily Quality Assurance Phantom for Six Degrees-Of-Freedom Couch Testing
    Gaballa, H.
    O'Brien, M.
    Riegel, A.
    Jamshidi, A.
    Klein, E.
    MEDICAL PHYSICS, 2016, 43 (06) : 3571 - 3571
  • [45] Technical aspects of quality assurance for intravitreal injections (IVI)
    Bucher, F.
    Mussinghoff, P.
    Kuehn, T.
    Stahl, A.
    Boehringer, D.
    OPHTHALMOLOGE, 2020, 117 (04): : 307 - 312
  • [46] Harmonisation and quality assurance of criminal technical report in the Switzerland
    Pfefferli, Peter W.
    KRIMINALISTIK, 2007, 61 (8-9): : 572 - 574
  • [47] ORGANIZATIONAL ASPECTS OF QUALITY ASSURANCE FOR TECHNICAL CONDITIONS OF TRAINS
    Spijunovic, Zeljko
    Mirkovic, Radosav
    Radojicic, Miroslav
    Nesic, Zoran
    METALURGIA INTERNATIONAL, 2012, 17 (09): : 115 - 119
  • [48] The Role of Accreditation in Promoting Quality Assurance of Technical Education
    Natarajan, R.
    INTERNATIONAL JOURNAL OF ENGINEERING EDUCATION, 2000, 16 (02) : 85 - 96
  • [49] INTERNAL QUALITY ASSURANCE SYSTEMS AT THE TECHNICAL UNIVERSITY OF MADRID
    Diaz, D.
    Collado, M.
    Herreros, M. A.
    Leo, T.
    Perez, R.
    Mora, E.
    Moran, J. L.
    Rodriguez, M. C.
    de lara, J.
    ICERI2015: 8TH INTERNATIONAL CONFERENCE OF EDUCATION, RESEARCH AND INNOVATION, 2015, : 6482 - 6486
  • [50] Technical editing as quality assurance: Adding value to content
    Corbin, M
    Moell, P
    Boyd, M
    TECHNICAL COMMUNICATION, 2002, 49 (03) : 286 - 300