Technical Note: Unified imaging and robotic couch quality assurance

被引:6
|
作者
Cook, Molly C. [1 ]
Roper, Justin [2 ]
Elder, Eric S. [2 ]
Schreibmann, Eduard [2 ]
机构
[1] Georgia Inst Technol, Med Phys Program, 770 State St, Atlanta, GA 30332 USA
[2] Emory Univ, Winship Canc Inst, Dept Radiat Oncol, 1365-C Clifton Rd NE, Atlanta, GA 30322 USA
关键词
quality assurance; image guided radiation therapy; robotic couch; 6D couch; image registration; ACCELERATOR; RADIOSURGERY; RADIOTHERAPY; ISOCENTER; ROTATION;
D O I
10.1118/1.4960369
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To introduce a simplified quality assurance (QA) procedure that integrates tests for the linac's imaging components and the robotic couch. Current QA procedures for evaluating the alignment of the imaging system and linac require careful positioning of a phantom at isocenter before image acquisition and analysis. A complementary procedure for the robotic couch requires an initial displacement of the phantom and then evaluates the accuracy of repositioning the phantom at isocenter. We propose a two-in-one procedure that introduces a custom software module and incorporates both checks into one motion for increased efficiency. Methods: The phantom was manually set with random translational and rotational shifts, imaged with the in-room imaging system, and then registered to the isocenter using a custom software module. The software measured positioning accuracy by comparing the location of the repositioned phantom with a CAD model of the phantom at isocenter, which is physically verified using the MV port graticule. Repeatability of the custom software was tested by an assessment of internal marker location extraction on a series of scans taken over differing kV and CBCT acquisition parameters. Results: The proposed method was able to correctly position the phantom at isocenter within acceptable 1 mm and 1 degrees SRS tolerances, verified by both physical inspection and the custom software. Residual errors for mechanical accuracy were 0.26 mm vertically, 0.21 mm longitudinally, 0.55 mm laterally, 0.21 degrees in pitch, 0.1 degrees in roll, and 0.67 degrees in yaw. The software module was shown to be robust across various scan acquisition parameters, detecting markers within 0.15 mm translationally in kV acquisitions and within 0.5 mm translationally and 0.3 degrees rotationally across CBCT acquisitions with significant variations in voxel size. Agreement with vendor registration methods was well within 0.5 mm; differences were not statistically significant. Conclusions: As compared to the current two-step approach, the proposed QA procedure streamlines the workflow, accounts for rotational errors in imaging alignment, and simulates a broad range of variations in setup errors seen in clinical practice. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:5080 / 5083
页数:4
相关论文
共 50 条
  • [31] Mazor X Stealth Robotic Technology: A Technical Note
    O'Connor, Timothy E.
    O'Hehir, Mary Margaret
    Khan, Asham
    Mao, Jennifer Z.
    Levy, Lauren C.
    Mullin, Jeffrey P.
    Pollina, John
    WORLD NEUROSURGERY, 2021, 145 : 435 - 442
  • [32] The quality assurance of volumetric modulated arc therapy (VMAT) plans for early stage prostate cancer: a technical note
    Elith, Craig A.
    Dempsey, Shane E.
    Cao, Fred
    Farshadi, Afrooz
    Warren-Forward, Helen M.
    JOURNAL OF MEDICAL RADIATION SCIENCES, 2014, 61 (04): : 261 - 266
  • [33] Technical Note: Evaluation of plastic scintillator detector for small field stereotactic patient-specific quality assurance
    Qin, Yujiao
    Gardner, Stephen J.
    Kim, Joshua
    Huang, Yimei
    Wen, Ning
    Doemer, Anthony
    Chetty, Indrin J.
    MEDICAL PHYSICS, 2017, 44 (10) : 5509 - 5516
  • [34] Technical Note: Rapid prototyping of 3D grid arrays for image guided therapy quality assurance
    Kittle, David
    Holshouser, Barbara
    Slater, James M.
    Guenther, Bob D.
    Pitsianis, Nikos P.
    Pearlstein, Robert D.
    MEDICAL PHYSICS, 2008, 35 (12) : 5708 - 5712
  • [35] Technical Note: Flattening filter free beam from Halcyon linac: Evaluation of the profile parameters for quality assurance
    Fogliata, A.
    Cayez, R.
    Garcia, R.
    Khamphan, C.
    Reggiori, G.
    Scorsetti, M.
    Cozzi, L.
    MEDICAL PHYSICS, 2020, 47 (08) : 3669 - 3674
  • [36] Technical Note: Validation of an automatic ACR phantom quality assurance tool for an MR-guided radiotherapy system
    Gao, Yu
    Lotey, Rajiv
    Low, Daniel A.
    Hu, Peng
    Yang, Yingli
    MEDICAL PHYSICS, 2021, 48 (04) : 1540 - 1545
  • [37] Quality Assurance and Precision Analysis of robotic clipped Radiosurgery
    Blanck, O.
    Poppe, B.
    Hildebrandt, G.
    Dunst, J.
    Schweikard, A.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2014, 190 : 7 - 7
  • [38] Diagnostic Ultrasound Imaging Quality Assurance
    Zagzebski, J.
    MEDICAL PHYSICS, 2016, 43 (06) : 3800 - 3800
  • [39] Commentary: Quality assurance in ophthalmic imaging
    Vaidya, Harshit
    Bhende, Muna
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2019, 67 (08) : 1288 - 1288
  • [40] Machine-specific quality assurance procedure for stereotactic treatments with dynamic couch rotations
    Wilson, Byron
    Gete, Ermias
    MEDICAL PHYSICS, 2017, 44 (12) : 6529 - 6537