Technical Note: Unified imaging and robotic couch quality assurance

被引:6
|
作者
Cook, Molly C. [1 ]
Roper, Justin [2 ]
Elder, Eric S. [2 ]
Schreibmann, Eduard [2 ]
机构
[1] Georgia Inst Technol, Med Phys Program, 770 State St, Atlanta, GA 30332 USA
[2] Emory Univ, Winship Canc Inst, Dept Radiat Oncol, 1365-C Clifton Rd NE, Atlanta, GA 30322 USA
关键词
quality assurance; image guided radiation therapy; robotic couch; 6D couch; image registration; ACCELERATOR; RADIOSURGERY; RADIOTHERAPY; ISOCENTER; ROTATION;
D O I
10.1118/1.4960369
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To introduce a simplified quality assurance (QA) procedure that integrates tests for the linac's imaging components and the robotic couch. Current QA procedures for evaluating the alignment of the imaging system and linac require careful positioning of a phantom at isocenter before image acquisition and analysis. A complementary procedure for the robotic couch requires an initial displacement of the phantom and then evaluates the accuracy of repositioning the phantom at isocenter. We propose a two-in-one procedure that introduces a custom software module and incorporates both checks into one motion for increased efficiency. Methods: The phantom was manually set with random translational and rotational shifts, imaged with the in-room imaging system, and then registered to the isocenter using a custom software module. The software measured positioning accuracy by comparing the location of the repositioned phantom with a CAD model of the phantom at isocenter, which is physically verified using the MV port graticule. Repeatability of the custom software was tested by an assessment of internal marker location extraction on a series of scans taken over differing kV and CBCT acquisition parameters. Results: The proposed method was able to correctly position the phantom at isocenter within acceptable 1 mm and 1 degrees SRS tolerances, verified by both physical inspection and the custom software. Residual errors for mechanical accuracy were 0.26 mm vertically, 0.21 mm longitudinally, 0.55 mm laterally, 0.21 degrees in pitch, 0.1 degrees in roll, and 0.67 degrees in yaw. The software module was shown to be robust across various scan acquisition parameters, detecting markers within 0.15 mm translationally in kV acquisitions and within 0.5 mm translationally and 0.3 degrees rotationally across CBCT acquisitions with significant variations in voxel size. Agreement with vendor registration methods was well within 0.5 mm; differences were not statistically significant. Conclusions: As compared to the current two-step approach, the proposed QA procedure streamlines the workflow, accounts for rotational errors in imaging alignment, and simulates a broad range of variations in setup errors seen in clinical practice. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:5080 / 5083
页数:4
相关论文
共 50 条
  • [1] An Unified Imaging and Robotic Couch Quality Assurance Procedure, and Automation Software
    Schreibmann, E.
    Elder, E.
    MEDICAL PHYSICS, 2015, 42 (06) : 3357 - 3357
  • [2] Daily Quality Assurance for a 6D Robotic Couch System
    Detorie, N. C.
    Scinta, L.
    Dragojevic, I.
    Song, W.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [3] Technical Note: Independent component analysis for quality assurance in functional MRI
    Astrakas, Loukas G.
    Kallistis, Nikolaos S.
    Kalef-Ezra, John A.
    MEDICAL PHYSICS, 2016, 43 (02) : 983 - 992
  • [4] Radioluminescence imaging feasibility for robotic radiosurgery field size quality assurance
    Shakarami, Zahra
    Broggi, Sara
    Del Vecchio, Antonella
    Fiorino, Claudio
    Spinelli, Antonello E.
    MEDICAL PHYSICS, 2022, 49 (10) : 6588 - 6598
  • [5] TECHNICAL QUALITY ASSURANCE IN HISTOPATHOLOGY
    BARR, WT
    WILLIAMS, ED
    MEDICAL LABORATORY SCIENCES, 1983, 40 (03): : 257 - 261
  • [6] Technical Note: Quality assurance and relative dosimetry testing of a 60Co total body irradiator using optical imaging
    Tendler, Irwin I.
    Bredfeldt, Jeremy S.
    Zhang, Rongxiao
    Bruza, Petr
    Jermyn, Michael
    Pogue, Brian W.
    Gladstone, David J.
    MEDICAL PHYSICS, 2019, 46 (08) : 3674 - 3678
  • [7] Technical Note: Proof of concept for radiomics-based quality assurance for computed tomography
    Branco, Luciano R. F.
    Ger, Rachel B.
    Mackin, Dennis S.
    Zhou, Shouhao
    Court, Laurence E.
    Layman, Rick R.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (11): : 199 - 205
  • [8] Sensitivity of a Surface Imaging System in Assessing Robotic Couch Isocentricity
    El-Sherif, O.
    Remmes, N.
    Kruse, J.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [9] QUALITY ASSURANCE AND LANGUAGES IN TECHNICAL EDUCATION
    Greculescu, Anca
    Solomon, Gheorghe
    ENVIRONMENTAL ENGINEERING AND MANAGEMENT JOURNAL, 2009, 8 (04): : 715 - 719
  • [10] Technical Assurance for Construction Quality Management
    Jiang, Haohao
    FRONTIERS OF GREEN BUILDING, MATERIALS AND CIVIL ENGINEERING, PTS 1-8, 2011, 71-78 : 3397 - 3400