Intensity of quartz-tungsten-halogen light-curing units used in private practice in Toronto

被引:28
作者
El-Mowafy, O
El-Badrawy, W
Lewis, DW
Shokati, B
Kermalli, J
Soliman, O
Encioiu, A
Zawi, R
Rajwani, F
机构
[1] Univ Toronto, Fac Dent, Toronto, ON M5G 1G6, Canada
[2] Univ Toronto, Fac Arts & Sci, Toronto, ON M5G 1G6, Canada
[3] Univ Toronto, Fac Appl Dent, Toronto, ON M5G 1G6, Canada
[4] Univ Toronto, Fac Life Sci, Toronto, ON M5G 1G6, Canada
[5] Univ Toronto, Fac Sci, Toronto, ON M5G 1G6, Canada
关键词
intensity; quartz-tungsten-halogen light units; resin-based composite;
D O I
10.14219/jada.archive.2005.0260
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Background. The authors conducted a study to determine light intensity and heat/glare measurements of quartz-tungsten-halogen (QTH) light polymerization units used in dental offices. Methods. Research assistants visited 100 dental offices and assessed 214 QTH light units. They recorded each unit's model age, service history, light intensity and heat/glare emissions. Results. Mean light intensity was 526 milliwates per square centimeter (120-1,000 mW/cm(2)), with 26 units having intensity less than 300 mW/cm(2). The mean light unit age was 5.6 years. Light units older than three years had significantly lower output intensities that those that were one, two or three years old. The authors found a wide range of heat/glare measurements (3-300 mW/cm(2)), with 4.6 percent of the units having values greater than 50 mW/cm(2), including three with values of more than 200 mW/cm(2). The mean light intensity of units serviced in the preceding year was 539 mW/cm(2); it was 418 mW/cm(2) for units serviced from one to six years previously. Conclusions. Light intensity and heat/glare values varied among the 214 units; some units had values well outside the recommended levels. Each unit's age and service history significantly affected its intensity. An awareness campaign is needed to promote testing, repair or replacement of light polymerization inits. Periodic testing of light polymerization units should be considered by regulatory bodies to ensure optimum quality of composite restorations. Clinical Implications. Light polymerization units in some private dental offices in Toronto had intensities that may result in composites restorations with inferior properties. Dentists need to regularly monitor the intensity of light polymerization units and maintain the units to ensure quality composite restorations.
引用
收藏
页码:766 / 773
页数:8
相关论文
共 33 条
[1]  
ASMUSSEN E, 1982, SCAND J DENT RES, V90, P490
[2]   EFFECT OF EXPOSURE TIME ON THE DEPTH OF POLYMERIZATION OF A VISIBLE LIGHT-CURED COMPOSITE RESIN [J].
BAHARAV, H ;
ABRAHAM, D ;
CARDASH, HS ;
HELFT, M .
JOURNAL OF ORAL REHABILITATION, 1988, 15 (02) :167-172
[3]   EVALUATING INTENSITY OUTPUT OF CURING LIGHTS IN PRIVATE DENTAL OFFICES [J].
BARGHI, N ;
BERRY, T ;
HATTON, C .
JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 1994, 125 (07) :992-996
[4]   VISIBLE LIGHT-CURED COMPOSITE RESTORATIVE - CLINICAL OPEN ASSESSMENT [J].
BASSIOUNY, MA ;
GRANT, AA .
BRITISH DENTAL JOURNAL, 1978, 145 (11) :327-330
[5]   CORRELATION OF CYTOTOXICITY, FILLER LOADING AND CURING TIME OF DENTAL COMPOSITES [J].
CAUGHMAN, WF ;
CAUGHMAN, GB ;
SHIFLETT, RA ;
RUEGGEBERG, F ;
SCHUSTER, GS .
BIOMATERIALS, 1991, 12 (08) :737-740
[6]   SPECTRAL DISTRIBUTIONS OF DENTAL PHOTO-POLYMERIZATION SOURCES [J].
COOK, WD .
JOURNAL OF DENTAL RESEARCH, 1982, 61 (12) :1436-1438
[7]  
Costa CAD, 2003, OPER DENT, V28, P365
[8]   Comparative efficiency of plasma and halogen light sources on composite micro-hardness in different curing conditions [J].
Dietschi, D ;
Marret, N ;
Krejci, I .
DENTAL MATERIALS, 2003, 19 (06) :493-500
[9]  
El-Mowafy O, 2002, J DENT RES, V81, pA323
[10]  
El-Mowafy O, 2002, J CAN DENT ASSOC, V68, P568