Stereoscopic particle image velocimetry of the impinging venous needle jet during hemodialysis

被引:2
|
作者
Fulker, David [1 ]
Forster, Kyle [1 ]
Simmons, Anne [1 ]
Barber, Tracie [1 ]
机构
[1] Univ New South Wales, Sch Mech & Mfg Engn, Sydney, NSW, Australia
关键词
Stereo particle image velocimetry; Impinging jet; Venous needle; Hemodialysis; ARTERIOVENOUS-FISTULA; INTIMAL HYPERPLASIA; FLOW; SHEAR;
D O I
10.1016/j.ijheatfluidflow.2017.07.005
中图分类号
O414.1 [热力学];
学科分类号
摘要
Stereoscopic particle image velocimetry (S-P1V) was used to quantify the dynamic flow field caused by the venous needle jet (VNJ) in an idealised model of hemodialysis cannulation. Scaling based on Reynolds number ensured dynamic similarity with physiological conditions. Measurements taken along the center plane indicate the presence of a steady secondary flow region, which develops downstream of the impingement region. Upon impingement, a wall jet forms on the floor of the vein and spreads along the curvature of the vessel. Circulating flow forms due to the interaction between the jet spreading and the wall jet. This secondary flow region represents a potential site of stenosis on the roof of the vein where flow is reversed. The effects of the circulating flows can be minimized by using shallow needle angles, low needle flow rates and placement of the needle away from the walls of the vein. (C) 2017 Elsevier Inc. All rights reserved.
引用
收藏
页码:59 / 68
页数:10
相关论文
共 50 条
  • [21] Distortion compensation for generalized stereoscopic particle image velocimetry
    Soloff, S.M.
    Adrian, R.J.
    Liu, Z.-C.
    Measurement Science & Technology, 1997, 8 (12): : 1441 - 1454
  • [22] A simple calibration technique for stereoscopic particle image velocimetry
    Andreas Fouras
    Jonathan Dusting
    Kerry Hourigan
    Experiments in Fluids, 2007, 42 : 799 - 810
  • [23] A simple calibration technique for stereoscopic particle image velocimetry
    Fouras, Andreas
    Dusting, Jonathan
    Hourigan, Kerry
    EXPERIMENTS IN FLUIDS, 2007, 42 (05) : 799 - 810
  • [24] Stereoscopic particle image velocimetry using telecentric lenses
    Fournel, T
    Coudert, S
    Fournier, C
    Ducottet, C
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2003, 14 (04) : 494 - 499
  • [25] Stereoscopic particle image velocimetry in a transonic turbine stage
    Lang, H
    Morck, T
    Woisetschläger, J
    EXPERIMENTS IN FLUIDS, 2002, 32 (06) : 700 - 709
  • [26] Stereoscopic particle image velocimetry in a transonic turbine stage
    H. Lang
    T. Mørck
    J. Woisetschläger
    Experiments in Fluids, 2002, 32 : 700 - 709
  • [27] Comparison of stereoscopic particle image velocimetry and volumetric particle tracking velocimetry in the wake of a ducted propeller
    Friedhoff, B.
    Roettig, F.
    Wennemar, K.
    Hoyer, K.
    Beslac, R.
    Hesseling, C.
    Beck, T.
    OCEAN ENGINEERING, 2021, 234 (234)
  • [28] Particle image velocimetry applied to a spray jet
    Brandt, Andreas
    Merzkirch, Wolfgang
    Particle and Particle Systems Characterization, 1994, 11 (02): : 156 - 158
  • [29] PARTICLE IMAGE VELOCIMETRY APPLIED TO A SPRAY JET
    BRANDT, A
    MERZKIRCH, W
    PARTICLE & PARTICLE SYSTEMS CHARACTERIZATION, 1994, 11 (02) : 156 - 158
  • [30] XPIV-multi-plane stereoscopic particle image velocimetry
    Liberzon, A
    Gurka, R
    Hetsroni, G
    EXPERIMENTS IN FLUIDS, 2004, 36 (02) : 355 - 362