Facile Design of Phase Separation for Microfluidic Droplet-Based Liquid Phase Microextraction as a Front End to Electrothermal Vaporization-ICPMS for the Analysis of Trace Metals in Cells

被引:16
作者
Yu, Xiaoxiao [1 ]
Chen, Beibei [1 ]
He, Man [1 ]
Wang, Han [1 ]
Tian, Songbai [1 ]
Hu, Bin [1 ]
机构
[1] Wuhan Univ, Dept Chem, Minist Educ, Key Lab Analyt Chem Biol & Med, Wuhan 430072, Hubei, Peoples R China
关键词
PLASMA-MASS SPECTROMETRY; SOLVENT-EXTRACTION; BIOLOGICAL SAMPLES; HEPG2; CELLS; GLASS CHIP; SYSTEM; MS; INTEGRATION; MICROCHIP; KINETICS;
D O I
10.1021/acs.analchem.8b03078
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The issue of quantifying trace metals in cells has drawn widespread attention but is threatened with insufficient sensitivity of the instruments, complex cellular matrix and limited cell consumption. In this study, micro-fluidic droplet-based liquid phase microextraction (LPME), as a miniaturized platform, was developed and combined with electrothermal vaporization (ETV)-inductively coupled plasma mass spectrometry (ICPMS) for the analysis of trace Cd, Hg, Pb, and Bi in cells. A novel and facile design of phase separation region was proposed, which made the phase separation very easily for subsequent ETV-ICPMS detection. Mechanism of the phase separation was carefully discussed using the incompressible formulation of the Navier-Stokes equations. The developed microfluidic droplet-based LPME system exhibited much higher extraction efficiency to target metals than microfluidic stratified flow-based LPME. Under the optimized conditions, the limits of detection of the proposed microfluidic droplet-based LPME-ETV-ICPMS system were 2.5, 3.9, 5.5, and 3.4 ng L-1 for Cd, Hg, Pb, and Bi, respectively. The accuracy of the developed method was well validated by analyzing the target metals in Certified Reference Materials of GBW07601a human hair. Finally, the proposed method was successfully applied to the analysis of target metals in HeLa and HepG2 cells with the recoveries for the spiked samples ranging from 83.5 to 112.3%. Overall, the proposed design is a simple and reliable solution for the phase separation on droplet-chip and the microfluidic droplet-based LPME-ETV-ICPMS combination strategy shows great promise for trace elements analysis in cells.
引用
收藏
页码:10078 / 10086
页数:9
相关论文
共 43 条
  • [1] Microscale tipstreaming in a microfluidic flow focusing device
    Anna, Shelley L.
    Mayer, Hans C.
    [J]. PHYSICS OF FLUIDS, 2006, 18 (12)
  • [2] Formation of dispersions using "flow focusing" in microchannels
    Anna, SL
    Bontoux, N
    Stone, HA
    [J]. APPLIED PHYSICS LETTERS, 2003, 82 (03) : 364 - 366
  • [3] Intracellular signaling pathways involved in mediating the effects of lead on the transcription factor Sp1
    Atkins, DS
    Basha, MR
    Zawia, NH
    [J]. INTERNATIONAL JOURNAL OF DEVELOPMENTAL NEUROSCIENCE, 2003, 21 (05) : 235 - 244
  • [4] Inductively Coupled Plasma Mass Spectrometry
    Beauchemin, Diane
    [J]. ANALYTICAL CHEMISTRY, 2010, 82 (12) : 4786 - 4810
  • [5] Magnetic solid phase microextraction on a microchip combined with electrothermal vaporization-inductively coupled plasma mass spectrometry for determination of Cd, Hg and Pb in cells
    Chen, Beibei
    Heng, Shujing
    Peng, Hanyong
    Hu, Bin
    Yu, Xu
    Zhang, Zhiling
    Pang, Daiwen
    Yue, Xin
    Zhu, Ying
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2010, 25 (12) : 1931 - 1938
  • [6] A novel droplet dosing strategy-based versatile microscale biosensor for detection of DNA, protein and ion
    Chen, Jinyang
    Liu, Yufei
    Ye, Tai
    Xiang, Xia
    Ji, Xinghu
    He, Zhike
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2015, 215 : 206 - 214
  • [7] Fast extraction and enrichment of rare earth elements from waste water via microfluidic-based hollow droplet
    Chen, Zhuo
    Wang, Wen-Ting
    Sang, Fu-Ning
    Xu, Jian-Hong
    Luo, Guang-Sheng
    Wang, Yun-Dong
    [J]. SEPARATION AND PURIFICATION TECHNOLOGY, 2017, 174 : 352 - 361
  • [8] Molecular mechanisms of methylmercury-induced cell death in human HepG2 cells
    Cuello, Susana
    Goya, Luis
    Madrid, Yolanda
    Campuzano, Susana
    Pedrero, Maria
    Bravo, Laura
    Camara, Carmen
    Ramos, Sonia
    [J]. FOOD AND CHEMICAL TOXICOLOGY, 2010, 48 (05) : 1405 - 1411
  • [9] Ordered and disordered patterns in two-phase flows in microchannels
    Dreyfus, R
    Tabeling, P
    Willaime, H
    [J]. PHYSICAL REVIEW LETTERS, 2003, 90 (14) : 4
  • [10] Liquid-liquid extraction of uranium(VI) with Aliquat® 336 from HCl media in microfluidic devices: Combination of micro-unit operations and online ICP-MS determination
    Helle, Gwendolyne
    Mariet, Clarisse
    Cote, Gerard
    [J]. TALANTA, 2015, 139 : 123 - 131