Wide-Range Calibration of Corneal Backscatter Analysis by In Vivo Confocal Microscopy

被引:12
作者
Hillenaar, Toine [1 ,2 ]
Sicam, Victor Arni D. P. [2 ]
Vermeer, Koenraad A. [2 ]
Braaf, Boy [2 ]
Remeijer, Lies [1 ,2 ]
Cals, Roger H. H. [1 ]
de Boer, Johannes F. [2 ,3 ]
机构
[1] Rotterdam Eye Hosp, Cornea & External Dis Serv, NL-3000 LM Rotterdam, Netherlands
[2] Rotterdam Ophthalm Inst ROI, Rotterdam, Netherlands
[3] Vrije Univ Amsterdam, Dept Phys & Astron, Amsterdam, Netherlands
关键词
LASER PHOTOREFRACTIVE KERATECTOMY; AUTOMATED ENDOTHELIAL KERATOPLASTY; EPITHELIAL BASEMENT-MEMBRANE; OPTICAL COHERENCE TOMOGRAPHY; LIGHT-SCATTERING; SITU KERATOMILEUSIS; STROMAL CHANGES; PENETRATING KERATOPLASTY; OBJECTIVE MEASUREMENT; EXCIMER KERATECTOMY;
D O I
10.1167/iovs.10-6314
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To report intra- and interinstrument calibration methods for corneal backscatter analysis by in vivo confocal microscopy. METHODS. Applicability of two reference standards was evaluated for corneal backscatter calibration. Repeated measurements of four concentrations of AMCO Clear (GFS Chemicals, Inc., Powell, OH) suspension and three transparencies (26%, 49%, and 65%) of polymethylmethacrylate (PMMA) slabs were performed to assess image intensity acquisition in a wide backscatter range. Intra- and intersession repeatability and lot-to-lot variation were determined for both standards. The effect of light intensity (LI) variation on image intensity acquisition was evaluated by examination of PMMA slabs with nonreference (60% and 80%) and reference (72%) LIs. Both reference standards were implemented in the protocol. Intrainstrument calibration was verified by measuring three normal corneas with 60%, 72%, and 80% LIs. Interinstrument calibration was tested by measuring PMMA slabs on a second, similar confocal microscope. RESULTS. AMCO Clear was used to express image intensity in absolute scatter units (SU), whereas the 49% transparent PMMA slab showed best repeatability, without image saturation, to adjust for LI variation. Intrainstrument calibration for LI variation reduced mean differences from -38.3% to 1.7% (60% LI) and from 33.9% to -0.6% (80% LI). The mean difference between similar microscopes decreased from 18.4% to 1.2%, after calibration of the second microscope. CONCLUSIONS. Large interinstrument differences necessitate calibration of corneal backscatter measurements. With AMCO Clear suspension and PMMA slabs, standardization was achieved in a wide backscatter range corresponding to normal and opaque corneas. These methods can easily be applied in ophthalmic practice. (Invest Ophthalmol Vis Sci. 2011;52:2136-2146) DOI: 10.1167/iovs.10-6314
引用
收藏
页码:2136 / 2146
页数:11
相关论文
共 63 条
[1]   HIGH-FREQUENCY ULTRASOUND QUANTITATIVE-ANALYSES OF CORNEAL SCARRING FOLLOWING EXCIMER-LASER KERATECTOMY [J].
ALLEMANN, N ;
CHAMON, W ;
SILVERMAN, RH ;
AZAR, DT ;
REINSTEIN, DZ ;
STARK, WJ ;
COLEMAN, J .
ARCHIVES OF OPHTHALMOLOGY, 1993, 111 (07) :968-973
[2]  
Andrade H A, 1990, Refract Corneal Surg, V6, P346
[3]  
Auzerie O, 2002, J FR OPHTALMOL, V25, P9
[4]   Scheimpflug anterior segment photography assessment of wound healing after myopic excimer laser photorefractive keratectomy [J].
Binder, PS ;
Bosem, M ;
Weinreb, RN .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 (02) :205-212
[5]  
Braunstein RE, 1996, OPHTHALMOLOGY, V103, P439
[6]   Central corneal thickness - Z-ring corneal confocal microscopy versus ultrasound pachymetry [J].
Brugin, Erica ;
Ghirlando, Alessandra ;
Gambato, Catia ;
Midena, Edoardo .
CORNEA, 2007, 26 (03) :303-307
[7]  
Busin M, 1989, Refract Corneal Surg, V5, P155
[8]  
Chang SS, 1996, J REFRACT SURG, V12, P412
[9]   COMPUTERIZED SCHEIMPFLUG DENSITOMETRY AS A MEASURE OF CORNEAL OPACIFICATION FOLLOWING EXCIMER-LASER SURGERY [J].
CHERNY, M ;
STASIUK, R ;
KELLY, P ;
LEE, S ;
GOLEMBO, G ;
TAYLOR, H .
OPHTHALMIC RESEARCH, 1994, 26 :48-54
[10]   Confocal Microscopy in Ophthalmology [J].
Erie, Jay C. ;
McLaren, Jay W. ;
Patel, Sanjay V. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 148 (05) :639-646