Intrasession repeatability of ocular anatomical measurements obtained with a multidiagnostic device in healthy eyes

被引:13
作者
Pinero, David P. [1 ]
Cabezos, Inmaculada [1 ]
Lopez-Navarro, Alberto [2 ]
de Fez, Dolores [1 ]
Caballero, Maria T. [1 ]
Camps, Vicente J. [1 ]
机构
[1] Univ Alicante, Dept Opt Pharmacol & Anat, Grp Opt & Visual Percept, Crta San Vicente Raspeig S-N, Alicante 03016, Spain
[2] Univ Alicante, Prevent Serv, Optometry Clin, Alicante, Spain
来源
BMC OPHTHALMOLOGY | 2017年 / 17卷
关键词
Anterior chamber depth; Pachymetry; White-to-white; Irido-corneal angle; Scheimpflug imaging; VX120; system; LOW-COHERENCE REFLECTOMETRY; PHOTOGRAPHY-BASED SYSTEM; INTRASUBJECT REPEATABILITY; ANTERIOR SEGMENT; BIOMETER REPEATABILITY; AGREEMENT; CORNEAL; REPRODUCIBILITY; TOPOGRAPHER; KERATOCONUS;
D O I
10.1186/s12886-017-0589-x
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: To evaluate the intrasession repeatability of anterior chamber depth (ACD), central (CCT) and peripheral corneal thickness (PCT), white-to-white diameter (WTW), and irido-corneal angle (IA) measurements obtained with a multidiagnostic device in healthy eyes. Methods: A total of 107 eyes of 107 patients ranging in age from 23 to 65 years were examined with the VX120 system (Visionix-Luneau Technologies). Three consecutive measurements were obtained with this device to assess the intrasession repeatability of ACD, CCT, PCT at different nasal and temporal locations, WTW, and nasal and temporal IA. Data analysis included the calculation of within-subject standard deviation (S-w), intrasubject precision (1.96xS(w)), coefficient of variation (CV) and intraclass correlation coefficient (ICC). Results: The S-w and CV for ACD was 0.03 mm and 1.16%, respectively, with an ICC of 0.992. The Sw values for central and peripheral pachymetric measurements were below 9 mu m, with CV of less than 1.6% and ICC of 0.976 or higher. For IA measurements, S-w values of 0.84 or lower were found, with a CV between 1 and 2%, and an ICC of more than 0.970. The S-w for WTW was 0.24 mm and the CV was 1.95%. No statistically significant correlations were found between any anatomical parameter evaluated and their S-w and CV values associated (-0.220 <= r <= 0.204, p >= 0.125). Conclusions: The VX120 system is able to provide repeatable measurements of anatomical parameters in healthy eyes. Inter-observer repeatability should be evaluated in future studies.
引用
收藏
页数:8
相关论文
共 23 条
  • [1] Asgari S, 2016, INT OPHTHALMOL
  • [2] Intrasubject repeatability of corneal power, thickness, and wavefront aberrations with a new version of a dual rotating Scheimpflug-Placido system
    Cervino, Alejandro
    Dominguez-Vicent, Alberto
    Ferrer-Blasco, Teresa
    Garcia-Lazaro, Santiago
    Albarran-Diego, Cesar
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (01) : 186 - 192
  • [3] Scheimpflug-Placido topographer and optical low-coherence reflectometry biometer: Repeatability and agreement
    Chen, Wuhe
    McAlinden, Colm
    Pesudovs, Konrad
    Wang, Qinmei
    Lu, Fan
    Feng, Yifan
    Chen, Jie
    Huang, Jinhai
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (09) : 1626 - 1632
  • [4] Repeatability, Reproducibility, and Agreement Between Three Different Scheimpflug Systems in Measuring Corneal and Anterior Segment Biometry
    Hernandez-Camarena, Julio C.
    Chirinos-Saldana, Patricia
    Navas, Alejandro
    Ramirez-Miranda, Arturo
    de la Mota, Angie
    Jimenez-Corona, Aida
    Graue-Hernandez, Enrique O.
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2014, 30 (09) : 616 - U44
  • [5] Repeatability and reproducibility of ocular biometry using a new noncontact optical low-coherence interferometer
    Huang, Jinhai
    Savini, Giacomo
    Wu, Fan
    Yu, Xinxin
    Yang, Jing
    Yu, Ayong
    Yu, Ye
    Wang, Qinmei
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (10) : 2233 - 2241
  • [6] Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry
    Kunert, Kathleen S.
    Peter, Monika
    Blum, Marcus
    Haigis, Wolfgang
    Sekundo, Walter
    Schuetze, Juliane
    Bueehren, Tobias
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (01) : 76 - 83
  • [7] Biometry with a new swept-source optical coherence tomography biometer: Repeatability and agreement with an optical low-coherence reflectometry device
    Kurian, Mathew
    Negalur, Nikhil
    Das, Sudeep
    Puttaiah, Narendra K.
    Haria, Dhawal
    Tejal, S. J.
    Thakkar, Maithil M.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (04) : 577 - 581
  • [8] Precision of higher-order aberration measurements with a new Placido-disk topographer and Hartmann-Shack wavefront sensor
    Lopez-Miguel, Alberto
    Martinez-Almeida, Loreto
    Gonzalez-Garcia, Maria J.
    Coco-Martin, Maria B.
    Sobrado-Calvo, Paloma
    Maldonado, Miguel J.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (02) : 242 - 249
  • [9] Repeatability and Intrasession Reproducibility Obtained by the Sirius Anterior Segment Analysis System
    Masoud, Muhannad
    Livny, Eitan
    Bahar, Irit
    [J]. EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2015, 41 (02): : 107 - 110
  • [10] Pachymetric measurements with a new Scheimpflug photography-based system Intraobserver repeatability and agreement with optical coherence tomography pachymetry
    Milla, Myriam
    Pinero, David P.
    Amparo, Francisco
    Alio, Jorge L.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (02) : 310 - 316