Cup-To-Neck Contact and Range of Motion after Total Hip Arthroplasty with Large Head Diameters: An Original Three-Dimensional Combined Gait and Videofluoroscopy Analysis

被引:1
作者
Belvedere, Claudio [1 ]
Lullini, Giada [1 ]
Ortolani, Maurizio [1 ]
Ensini, Andrea [2 ]
Durante, Stefano [3 ]
Ruberto, Francesco [4 ]
Turner, Emma [1 ]
Leardini, Alberto [1 ]
机构
[1] IRCCS Ist Ortoped Rizzoli, Movement Anal Lab, Via Barbiano 1-10, I-40136 Bologna, Italy
[2] IRCCS Ist Ortoped Rizzoli, Orthopaed & Traumatol Clin 1, Via Barbiano 1-10, I-40136 Bologna, Italy
[3] IRCCS Ist Ortoped Rizzoli, Nursing Tech & Rehabil Assistance Serv, I-40136 Bologna, Italy
[4] Italian Univ Sport & Movement Foro Italico, Dept Movement Human & Hlth Sci, I-00135 Rome, Italy
来源
APPLIED SCIENCES-BASEL | 2020年 / 10卷 / 08期
关键词
total hip arthroplasty; gait analysis; 3D videofluoroscopy; stem-to-liner impingement; daily living activities; RESURFACING ARTHROPLASTY; FEMORAL HEADS; IN-VITRO; REPLACEMENT; DISLOCATION; KINEMATICS; JOINT; KNEE; IMPINGEMENT; OSTEOARTHRITIS;
D O I
10.3390/app10082695
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
After a total hip arthroplasty, a limited range of motion and lower-limb disability continue to be observed, with these being mainly associated with the implant design and the head-to-neck ratio. Larger diameters of the head bearings were assumed to provide better stability, a larger range of motion, and smaller risks of dislocation and stem-to-liner impingement. However, these claims have never been demonstrated in real patients. The specific aim of this study was to assess, via multi-instrumental analysis, whether the range of motion of a replaced hip is limited by the stem-to-liner contact in patients with large femoral head diameters. Twenty-three patients with a total hip arthroplasty were evaluated at their one-year follow-ups using clinical and instrumental examinations. A combined three-dimensional gait analysis of the full body and videofluoroscopy analysis of the replaced hip were performed during the execution of standard, i.e., daily living, and more demanding motor tasks. The latter were meant to reach the extreme range of motion at the replaced hip site, thus revealing possible stem-to-liner contact. An original technique based on imaging and computer-aided design (CAD) models of the prosthesis components was developed to calculate the stem-to-liner distance. Excellent clinical scores were observed in the study. The gait analysis showed that the range of motion of the replaced hip in the sagittal plane, averaged over all patients, ranged from 28 degrees to 78 degrees in standard activities. In more demanding tasks, single peaks were as high as 110 degrees, 39 degrees, and 60 degrees in the sagittal, frontal, and transverse anatomical planes, respectively. In all motor tasks, the stem-to-liner distances ranged from 8.7 to 13.0 mm on average, with one outlier minimum distance being 2.2 mm. This study shows that, even in demanding motor tasks and with an extreme range of motion, the hip joint replaced with large femoral head diameters did not experience impingement between the prosthesis components.
引用
收藏
页数:15
相关论文
共 56 条
[1]  
Agarwala S., 2017, OPEN J ORTHOP, DOI [10.4236/ojo.2017.712044, DOI 10.4236/OJO.2017.712044]
[2]  
Amstutz HC, 2019, BONE JOINT J, V101B, P1186, DOI 10.1302/0301-620X.101B10.BJJ-2018-1535.R2
[3]   TREATMENT OF PRIMARY OSTEOARTHRITIS OF THE HIP - A COMPARISON OF TOTAL JOINT AND SURFACE REPLACEMENT ARTHROPLASTY [J].
AMSTUTZ, HC ;
THOMAS, BJ ;
JINNAH, R ;
KIM, W ;
GROGAN, T ;
YALE, C .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1984, 66A (02) :228-241
[4]   Biomechanical changes and recovery of gait function after total hip arthroplasty for osteoarthritis: a systematic review and meta-analysis [J].
Bahl, J. S. ;
Nelson, M. J. ;
Taylor, M. ;
Solomon, L. B. ;
Arnold, J. B. ;
Thewlis, D. .
OSTEOARTHRITIS AND CARTILAGE, 2018, 26 (07) :847-863
[5]   Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy [J].
Banks, SA ;
Hodge, WA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1996, 43 (06) :638-649
[6]  
BELLAMY N, 1988, J RHEUMATOL, V15, P1833
[7]   Three-dimensional motion analysis of the human knee joint: comparison between intra- and post-operative measurements [J].
Belvedere, C. ;
Tamarri, S. ;
Notarangelo, D. P. ;
Ensini, A. ;
Feliciangeli, A. ;
Leardini, A. .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2013, 21 (10) :2375-2383
[8]   Less range of motion with resurfacing arthroplasty than with total hip arthroplasty In vitro examination of 8 designs [J].
Bengs, Benjamin C. ;
Sangiorgio, Sophia N. ;
Ebramzadeh, Edward .
ACTA ORTHOPAEDICA, 2008, 79 (06) :755-762
[9]   Excellent results of large-diameter ceramic-on-ceramic bearings in total hip arthroplasty IS SQUEAKING RELATED TO HEAD SIZE? [J].
Blakeney, W. G. ;
Beaulieu, Y. ;
Puliero, B. ;
Lavigne, M. ;
Roy, A. ;
Masse, V ;
Vendittoli, P-A .
BONE & JOINT JOURNAL, 2018, 100B (11) :1434-1441
[10]   Impingement in total hip replacement: mechanisms and consequences [J].
Brown, Thomas D. ;
Callaghan, John J. .
CURRENT ORTHOPAEDICS, 2008, 22 (06) :376-391