Nanoband electrode for high-performance in-channel amperometric detection in dual-channel microchip capillary electrophoresis

被引:19
作者
Chen, Chuanpin [1 ]
Teng, Wei [1 ]
Hahn, Jong Hoon [1 ]
机构
[1] Pohang Univ Sci & Technol, BK Sch Mol Sci, Dept Chem, Pohang 790784, South Korea
关键词
Capillary electrophoresis; Dual-channel; Electrochemical detection; In-channel detection; Microchip; BAND ELECTRODES; FARADAIC ELECTROCHEMISTRY; MICROELECTRODE ARRAYS; NANOELECTRODE ARRAYS; FABRICATION; ULTRAMICROELECTRODES; POLY(DIMETHYLSILOXANE); MICROCYLINDER; VOLTAMMETRY; DIMENSIONS;
D O I
10.1002/elps.201000661
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A nanoband electrode detector integrated with a dual-channel polydimethylsiloxane microchip is proposed for in-channel amperometric detection in microchip capillary electrophoresis. Gold nanoband electrodes, which were fabricated on SU-8 substrates with a 100-nm-width gold layer, were introduced into the dual-channel microchip to be an electrochemical detector. Due to the nano-sized width of the detector, the noise of the amperometric detection was significantly reduced, and a high separation resolution was achieved for monitoring the analytes. The detection sensitivity of the system was improved by high signal-to-noise ratio, and a low detection limit on microchip was obtained for p-aminophenol (2.09 nM). Because of the high resolution in measuring half-peak width, the plate number that is used to evaluate the separation efficiency was 1.5-fold higher than that using 50-mu m-width electrochemical detector. The effect of sample injection time and data acquisition time on separation efficiency was investigated, and an attractive separation efficiency was achieved with a plate number up to 17 500.
引用
收藏
页码:838 / 843
页数:6
相关论文
共 38 条
[1]  
AMATORE CA, 1986, J ELECTROANAL CHEM, V206, P23, DOI 10.1016/0022-0728(86)90253-6
[2]   Nanoelectrodes, nanoelectrode arrays and their applications [J].
Arrigan, DWM .
ANALYST, 2004, 129 (12) :1157-1165
[3]   DIGITAL-SIMULATION OF THE MEASURED ELECTROCHEMICAL RESPONSE OF REVERSIBLE REDOX COUPLES AT MICROELECTRODE ARRAYS - CONSEQUENCES ARISING FROM CLOSELY SPACED ULTRAMICROELECTRODES [J].
BARD, AJ ;
CRAYSTON, JA ;
KITTLESEN, GP ;
SHEA, TV ;
WRIGHTON, MS .
ANALYTICAL CHEMISTRY, 1986, 58 (11) :2321-2331
[4]   CONSTRUCTION AND USE OF PAIRED AND TRIPLE BAND MICROELECTRODES IN SOLUTIONS OF LOW IONIC-STRENGTH [J].
BARTELT, JE ;
DEAKIN, MR ;
AMATORE, C ;
WIGHTMAN, RM .
ANALYTICAL CHEMISTRY, 1988, 60 (19) :2167-2169
[5]   ELECTROCHEMISTRY IN ORGANIC-SOLVENTS WITHOUT SUPPORTING ELECTROLYTE USING PLATINUM MICROELECTRODES [J].
BOND, AM ;
FLEISCHMANN, M ;
ROBINSON, J .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1984, 168 (1-2) :299-312
[6]   THEORY AND EXPERIMENTAL CHARACTERIZATION OF LINEAR GOLD MICROELECTRODES WITH SUBMICROMETER THICKNESS [J].
BOND, AM ;
HENDERSON, TLE ;
THORMANN, W .
JOURNAL OF PHYSICAL CHEMISTRY, 1986, 90 (13) :2911-2917
[7]   Characteristics of nanoscopic Au band electrodes [J].
Caston, SL ;
McCarley, RL .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2002, 529 (02) :124-134
[8]   Dual-channel method for interference-free in-channel amperometric detection in microchip capillary electrophoresis [J].
Chen, Chuanpin ;
Hahn, Jong Hoon .
ANALYTICAL CHEMISTRY, 2007, 79 (18) :7182-7186
[9]   Design, fabrication, characterisation and application of nanoelectrode arrays [J].
Compton, Richard G. ;
Wildgoose, Gregory G. ;
Rees, Neil V. ;
Streeter, Ian ;
Baron, Ronan .
CHEMICAL PHYSICS LETTERS, 2008, 459 (1-6) :1-17
[10]  
Crevillén AG, 2009, LAB CHIP, V9, P346, DOI 10.1039/b809963d