Noninvasive Sensor for Continuous intraocular Pressure Monitoring

被引:1
作者
Lei, Shujing [1 ,2 ]
Zhang, Lei [1 ,2 ]
Fei, Gao [3 ]
Fan, Lei [1 ,2 ]
Dong, Helei [1 ,2 ]
Tan, Qiulin [1 ,2 ]
Xiong, Jijun [1 ,2 ]
Shen, Sanmin [1 ,2 ]
机构
[1] North Univ China, Minist Educ, Key Lab Instrumentat Sci & Dynam Measurement, Taiyuan 030051, Shanxi, Peoples R China
[2] North Univ China, Sci & Technol Elect Test & Measurement Lab, Taiyuan 030051, Shanxi, Peoples R China
[3] PLAGF, Xianyang Off, Xian Mil Representat Bur, Xinjiang, Peoples R China
关键词
IOP; LC resonant; inductor coils; PDMS; MEMS;
D O I
10.18494/SAM.2018.1991
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Pathological elevated intraocular pressure (lOP) is a primary index for glaucoma. In this paper, we present a biocompatible IOP sensor that operates based on the principle of inductor-capacitor (LC) resonance. The IOP sensor is composed solely of a spiral inductor that can sense deformations of corneal curvature. The inductor is fabricated by embedding gold coils into polydimethylsiloxane (PDMS), a soft, biologically compatible material, via MEMS technology. The resonant frequency of the sensor is designed to vary when the IOP is changed in experiments. The frequency response of the sensor is tested using a model eye, a tube, and a network analyzer. The results show that there is approximately linearity between the frequency response of the sensor and the IOP. The sensitivity of the sensor is 417.29 kHz/mmHg, when the pressure ranges from 7.5 to 36.88 mmHg. Consequently, an unstable IOP could be accurately tracked, using the calibrated relationship between frequency response and the IOP.
引用
收藏
页码:2811 / 2821
页数:11
相关论文
共 19 条
[11]  
Haque R. M., 2011, IMPLANTABLE MICROSYS
[12]   24-hour monitoring of intraocular pressure in glaucoma management: A retrospective review [J].
Hughes, E ;
Spry, P ;
Diamond, J .
JOURNAL OF GLAUCOMA, 2003, 12 (03) :232-236
[13]   Design and Optimization of Printed Spiral Coils for Efficient Transcutaneous Inductive Power Transmission [J].
Jow, Uei-Ming ;
Ghovanloo, Maysam .
IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2007, 1 (03) :193-202
[14]  
Kang Hao, 2013, Chinese Journal of Sensors and Actuators, V26, P498, DOI 10.3969/j.issn.1004-1699.2013.04.012
[15]  
Laukhin V., 2010, SENSOR ACTUAT A-PHYS, V170, P36
[16]  
Luo Tao, 2014, Chinese Journal of Sensors and Actuators, V27, P327, DOI 10.3969/j.issn.1004-1699.2014.03.010
[17]   Chronically Implanted Pressure Sensors: Challenges and State of the Field [J].
Yu, Lawrence ;
Kim, Brian J. ;
Meng, Ellis .
SENSORS, 2014, 14 (11) :20620-20644
[18]   Physical modeling of spiral inductors on silicon [J].
Yue, CP ;
Wong, SS .
IEEE TRANSACTIONS ON ELECTRON DEVICES, 2000, 47 (03) :560-568
[19]  
Zeng P., 2016, SENSORS 2016 IEEE, P1