Determination of uranium and plutonium in mixed oxide samples by X-ray methods and their comparison with potentiometry

被引:9
作者
Bootharajan, M. [1 ]
Kumar, G. V. S. Ashok [1 ]
Sreenivasulu, B. [1 ]
Senthilvadivu, R. [1 ]
Garnayak, P. M. [3 ]
Pandey, A. [4 ]
Kelkar, A. [4 ]
Sathe, D. B. [4 ]
Bhatt, R. B. [4 ]
Rao, C. V. S. Brahmananda [1 ]
Sundararajan, K. [1 ,2 ]
Jayaraman, V. [1 ,2 ]
Sivaraman, N. [1 ,2 ]
机构
[1] Indira Gandhi Ctr Atom Res, Mat Chem & Met Fuel Cycle Grp, Kalpakkam, India
[2] Indira Gandhi Ctr Atom Res, Homi Bhabha Natl Inst HBNI, Kalpakkam 603102, India
[3] Bhabha Atom Res Ctr, Nucl Recycle Board, FRFCF, Kalpakkam 603102, India
[4] Bhabha Atom Res Ctr, Nucl Recycle Board, Fuel Fabricat, Integrated Nucl Recycling Plant O, Tarapur 401 502, India
关键词
Uranium; Plutonium; KED; EDXRF; Potentiometry; COMPOSITIONAL CHARACTERIZATION;
D O I
10.1016/j.sab.2022.106481
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Uranium and plutonium in nuclear fuels are generally determined by conventional potentiometric methods which generates large quantities of radioactive waste. In the present study, K-edge densitometry (KED) and Energy dispersive X-ray fluorescence (EDXRF) have been examined for the determination of uranium and plutonium in solutions as an alternative to potentiometric methods. In the present study, uranium plutonium mixed oxide (MOX) pellets were dissolved in nitric acid medium containing small quantities of HF. Uranium and plutonium in MOX solutions were determined by KED and EDXRF methods and the results are compared with the one obtained by conventional potentiometric methods. The relative standard deviation (RSD) of the analytical results for U and Pu using KED were found to be better than & PLUSMN;1% and & PLUSMN; 0.9% respectively. Similarly, RSD for U and Pu by EDXRF were found to be better than & PLUSMN;1.6% and & PLUSMN; 0.8%, respectively. The absolute difference between KED and EDXRF with respect to potentiometric data for uranium and plutonium were calculated and it was found to be marginal indicating that KED and EDXRF methods can be employed for the routine analysis of U and Pu in solutions. X-ray methods employed in the present study are non-consumptive, direct and offer precise determination of U and Pu and generate no waste.
引用
收藏
页数:6
相关论文
共 25 条
[1]  
[Anonymous], 1985, KARLSR INT C AN CHEM
[2]  
[Anonymous], 1983, P SERIES
[3]   Development and validation of hybrid K-edge/K-XRF densitometer for assay of nuclear fuel from reprocessing plants [J].
Bootharajan, M. ;
Senthilvadivu, R. ;
Sundararajan, K. ;
Kumar, R. .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2020, 324 (02) :623-633
[4]   DETERMINATION OF PLUTONIUM AND URANIUM IN MIXED-OXIDE FUELS BY SEQUENTIAL REDOX TITRATION [J].
CHADWICK, PH ;
MCGOWAN, IR .
TALANTA, 1972, 19 (11) :1335-&
[5]   Recovery of plutonium from uranium analytical waste solution containing phosphate using 2-ethylhexyl 2-ethylhexyl phosphonic acid (PC88A) [J].
Chetty, K. Venugopal ;
Gamare, Jayashree S. ;
Godbole, A. G. ;
Vaidya, V. N. .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2006, 269 (01) :15-20
[6]   THE DETERMINATION OF URANIUM AND PLUTONIUM IN FUEL MATERIALS BY TITRATION AND BY K-EDGE ABSORPTION [J].
CROMBOOM, O ;
KOCH, L ;
WELLUM, R .
JOURNAL OF NUCLEAR MATERIALS, 1991, 178 (2-3) :249-253
[7]  
Davies W., 1964, Talanta, V11, P1203, DOI DOI 10.1016/0039-9140(64)80171-5
[8]   Development of a microanalytical energy dispersive X-ray fluorescence method for compositional characterization of (U, Pu)O2 samples [J].
Dhara, Sangita ;
Kumar, S. Sanjay ;
Jayachandran, Kavitha ;
Karnat, J. V. ;
Kumar, Ashok ;
Radhakrishna, J. ;
Misra, N. L. .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2017, 131 :124-129
[9]   Betterment in EDXRF analytical results for compositional characterization of mixed uranium thorium oxide samples with bead specimens compared with pressed pellet specimens [J].
Dhara, Sangita ;
Kumar, S. Sanjay ;
Misra, N. L. .
X-RAY SPECTROMETRY, 2016, 45 (05) :268-273
[10]   POTENTIOMETRIC DETERMINATION OF PLUTONIUM BY ARGENTIC OXIDATION FERROUS REDUCTION AND DICHROMATE TITRATION [J].
DRUMMOND, JL ;
GRANT, RA .
TALANTA, 1966, 13 (03) :477-&