In vivo gleno-humeral joint loads during forward flexion and abduction

被引:110
作者
Bergmann, G. [1 ]
Graichen, F. [1 ]
Bender, A. [1 ,2 ]
Rohlmann, A. [1 ]
Halder, A. [3 ]
Beier, A. [3 ]
Westerhoff, P. [1 ]
机构
[1] Charite Univ Med Berlin, Julius Wolff Inst, D-13353 Berlin, Germany
[2] Berlin Brandenburg Ctr Regenerat Therapies, Berlin, Germany
[3] Klin Endoprothet Sommerfeld, Beetz Sommerfeld, Germany
关键词
Shoulder joint; Load; Force; Moment; In vivo; Measurement; Telemetry; Gleno-humeral joint; Abduction; Forward flexion; Elevation; Friction; GLENOHUMERAL CONTACT FORCES; SHOULDER JOINT; MODEL; HIP; REPLACEMENT; MECHANISM; IMPLANTS;
D O I
10.1016/j.jbiomech.2011.02.142
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
To improve design and preclinical test scenarios of shoulder joint implants as well as computer-based musculoskeletal models, a precise knowledge of realistic loads acting in vivo is necessary. Such data are also helpful to optimize physiotherapy after joint replacement and fractures. This is the first study that presents forces and moments measured in vivo in the gleno-humeral joint of 6 patients during forward flexion and abduction of the straight arm. The peak forces and, even more, the maximum moments varied inter-individually to a considerable extent. Forces of up to 238%BW (percent of body weight) and moments up to 1.74%BWm were determined. For elevation angles of less than 90 degrees the forces agreed with many previous model-based calculations. At higher elevation angles, however, the measured loads still rose in contrast to the analytical results. When the exercises were performed at a higher speed, the peak forces decreased. The force directions relative to the humerus remained quite constant throughout the whole motion. Large moments in the joint indicate that friction in shoulder implants is high if the glenoid is not replaced. A friction coefficient of 0.1-0.2 seems to be realistic in these cases. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1543 / 1552
页数:10
相关论文
共 26 条
  • [1] Glenohumeral contact forces
    Anglin, C
    Wyss, UP
    Pichora, DR
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2000, 214 (H6) : 637 - 644
  • [2] In vivo glenohumeral contact forces -: Measurements in the first patient 7 months postoperatively
    Bergmann, G.
    Graichen, F.
    Bender, A.
    Kaab, M.
    Rohlmann, A.
    Westerhoff, P.
    [J]. JOURNAL OF BIOMECHANICS, 2007, 40 (10) : 2139 - 2149
  • [3] Hip contact forces and gait patterns from routine activities
    Bergmann, G
    Deuretzbacher, G
    Heller, M
    Graichen, F
    Rohlmann, A
    Strauss, J
    Duda, GN
    [J]. JOURNAL OF BIOMECHANICS, 2001, 34 (07) : 859 - 871
  • [4] Bergmann G., 2001, HIP98 LOADING HIP JO
  • [5] Tuberosity malposition and migration:: Reasons for poor outcomes after hemiarthroplasty for displaced fractures of the proximal humerus
    Boileau, P
    Krishnan, SG
    Tinsi, L
    Walch, G
    Coste, JS
    Molé, D
    [J]. JOURNAL OF SHOULDER AND ELBOW SURGERY, 2002, 11 (05) : 401 - 412
  • [6] FLOATING-SOCKET TOTAL SHOULDER REPLACEMENT - ANATOMICAL, BIOMECHANICAL, AND SURGICAL RATIONALE
    BUECHEL, FF
    PAPPAS, MJ
    DEPALMA, AF
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1978, 12 (01): : 89 - 114
  • [7] A BIOMECHANICAL MODEL TO QUANTIFY SHOULDER LOAD AT THE WORK PLACE
    DUL, J
    [J]. CLINICAL BIOMECHANICS, 1988, 3 (03) : 124 - 128
  • [8] Numerical modelling of the shoulder for clinical applications
    Favre, Philippe
    Snedeker, Jess G.
    Gerber, Christian
    [J]. PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 2009, 367 (1895): : 2095 - 2118
  • [9] Implantable 9-channel telemetry system for in vivo load measurements with orthopedic implants
    Graichen, Friedmar
    Arnold, Ruediger
    Rohlmann, Antonius
    Bergmann, Georg
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2007, 54 (02) : 253 - 261
  • [10] ESB clinical biomechanics award 2008: Complete data of total knee replacement loading for level walking and stair climbing measured in vivo with a follow-up of 6-10 months
    Heinlein, Bernd
    Kutzner, Ines
    Graichen, Friedmar
    Bender, Alwina
    Rohlmann, Antonius
    Halder, Andreas M.
    Beier, Alexander
    Bergmann, Georg
    [J]. CLINICAL BIOMECHANICS, 2009, 24 (04) : 315 - 326