Effect of the geometrical parameters of an HPGe detector on efficiency calculations using Monte Carlo methods

被引:8
作者
Subercaze, Alexandre [1 ]
Sauzedde, Thibault [2 ]
Domergue, Christophe [1 ]
Destouches, Christophe [1 ]
Philibert, Herve [1 ]
Fausser, Clement [1 ]
Thiollay, Nicolas [1 ]
Gregoire, Gilles [1 ]
Zoia, Andrea [3 ]
机构
[1] CEA, DES, IRESNE, DER,Lab Dosimetrie Capteurs & Instrumentat, F-13108 St Paul Les Durance, France
[2] CEA, DES, IRESNE, DER,Lab Projets Nucl, F-13108 St Paul Les Durance, France
[3] Univ Paris Saclay, CEA, Serv Etud Reacteurs & Math Appl, F-91191 Gif Sur Yvette, France
关键词
Gamma spectroscopy; Monte Carlosimulation; Efficiency measurements; HPGe detector modelisation; COINCIDENCE-SUMMING CORRECTIONS; GAMMA-RAY; DEAD-LAYER; SIMULATION; CALIBRATION; WELL; CASCADE; MODEL;
D O I
10.1016/j.nima.2022.167096
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This paper describes a procedure for constructing a Monte Carlo model of an HPGe detector used to measure activity. The purpose of this parametric study is to obtain a single model that effectively reproduces the detection efficiency for both small (< 1 cm) and large (> 40 cm) source-detector distances. We therefore investigated the effect of the HPGe detector's geometrical parameters on the Monte Carlo efficiency calculations. The results highlight detector parameters with the greatest effect on the calculations within a wide range of gamma energy levels (60-1900 keV) and source-detector distances (1 to 50 cm): the air gap, the dead layers thicknesses and the crystal height (see Figure 1). Analysis of these results has made it possible to determine the key parameters, together with the optimal energy and distance ranges, significantly reducing the computational time required. A broad range of calibrated point-like radioactive sources were used to ensure a high level of experimental efficiency when constructing the detector model. A single Monte Carlo model for this detector was thus optimised by minimising the difference between the experimental efficiency and the calculations. The single HPGe model resulting from this optimisation process led to an average deviation between experimental and simulated efficiency of 3% for both small and large source-detector distances.
引用
收藏
页数:12
相关论文
共 40 条
[11]  
GEANT4 collaboration, 2003, NUCL INSTRUM METHO A, V506, P250, DOI DOI 10.1016/S0168-9002(03)01368-8
[12]   INITIAL MCNP6 RELEASE OVERVIEW [J].
Goorley, T. ;
James, M. ;
Booth, T. ;
Brown, F. ;
Bull, J. ;
Cox, L. J. ;
Durkee, J. ;
Elson, J. ;
Fensin, M. ;
Forster, R. A. ;
Hendricks, J. ;
Hughes, H. G. ;
Johns, R. ;
Kiedrowski, B. ;
Martz, R. ;
Mashnik, S. ;
McKinney, G. ;
Pelowitz, D. ;
Prael, R. ;
Sweezy, J. ;
Waters, L. ;
Wilcox, T. ;
Zukaitis, T. .
NUCLEAR TECHNOLOGY, 2012, 180 (03) :298-315
[13]   Method for developing HPGe detector model in Monte Carlo simulation codes [J].
Haj-Heidari, M. T. ;
Safari, M. J. ;
Afarideh, H. ;
Rouhi, H. .
RADIATION MEASUREMENTS, 2016, 88 :1-6
[14]   The use of Monte Carlo calculations in the determination of a Ge detector efficiency curve [J].
Helmer, RG ;
Hardy, JC ;
Iacob, VE ;
Sanchez-Vega, M ;
Neilson, RG ;
Nelson, J .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 511 (03) :360-381
[15]  
Hugot F.-X., 2018, RPSD2018
[16]   Monte Carlo simulation of the response of a germanium detector for low-level spectrometry measurements using GEANT4 [J].
Hurtado, S ;
García-León, M ;
García-Tenorio, R .
APPLIED RADIATION AND ISOTOPES, 2004, 61 (2-3) :139-143
[17]  
Incerti S, 2020, CERN WIKI GEANT4
[18]   Self-attenuation corrections for Pb-210 in gamma-ray spectrometry using well and coaxial HPGe detectors [J].
Iurian, A. R. ;
Millward, G. E. ;
Sima, O. ;
Taylor, A. ;
Blake, W. .
APPLIED RADIATION AND ISOTOPES, 2018, 134 :151-156
[19]   Cascade summing effects in close geometry gamma-ray spectrometry [J].
Johnston, Peter N. ;
Hult, Mikael ;
Gasparro, Joel .
APPLIED RADIATION AND ISOTOPES, 2006, 64 (10-11) :1323-1328
[20]   Application of PENELOPE code to the efficiency calibration of coaxial germanium detectors [J].
Jurado Vargas, M. ;
Guerra, A. L. .
APPLIED RADIATION AND ISOTOPES, 2006, 64 (10-11) :1319-1322