Comparison of two different methods for physiologic dead space measurements in ventilated dogs in a clinical setting

被引:9
作者
Mosing, Martina [1 ]
Staub, Lukas [2 ]
Moens, Yves [3 ]
机构
[1] Univ Liverpool, Fac Vet Sci, Leahurst, Neston, England
[2] Univ Bern, MEM Inst Evaluat Res Orthoped Surg, CH-3012 Bern, Switzerland
[3] Vet Univ Vienna, Clin Anaesthesiol & Perioperat Intens Care, A-1030 Vienna, Austria
关键词
dead space; dog; single breath test; ventilation; VOLUMETRIC CAPNOGRAPHY; PULMONARY-EMBOLISM; DEADSPACE; THORACOSCOPY; PRESSURE; CHILDREN;
D O I
10.1111/j.1467-2995.2010.00548.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Objective To compare physiologic dead space (V(D)) and physiologic dead space to tidal volume (V(T)) ratio (V(D)/V(T)) obtained by an automated single breath test for carbon dioxide (CO(2)) (method SBT) and a manual calculation (method MC) in ventilated healthy dogs. Study design Prospective clinical study. Animals Twenty client-owned dogs, ASA I and II undergoing anaesthesia for clinical purposes. Methods Following pre-medication, induction of anaesthesia, and intubation of the trachea, intermittent positive pressure ventilation was commenced. Mixed expired CO(2) partial pressure (PeCO(2)) was measured by two methods: method MC by analysis, using an infrared capnograph, of the expired gas collected in a mixing box and method SBT which calculated it automatically by a device consisting of a mainstream capnograph and a pneumotachograph. At four time points arterial partial pressure of CO(2) (PaCO(2)) was measured. Physiologic dead space variables (V(D) and V(D)/V(T)) were calculated manually (method MC) or automatically (method SBT) using the Bohr-Enghoff equation. Method MC and SBT were compared using Bland-Altman plots and linear regression. Intra-class correlation coefficient (ICC) was used to measure consistency of each method. Results Four measurement pairs were obtained in all 20 dogs for method SBT and MC. The bias was -1.15 mmHg, 7.97 mL and 0.02 for PeCO(2), V(D) and V(D)/V(T), respectively. Linear regression analysis revealed a correlation coefficient (r2) of 0.79, 0.94, and 0.83 for PeCO(2), V(D) and V(D)/V(T), respectively. The ICC revealed an excellent consistency for both methods. Conclusions The single breath test (SBT) can be used for clinical evaluation of V(D) and V(D)/V(T) in anaesthetized ventilated dogs. Clinical relevance Through measuring V(D) and V(D)/V(T) important information about lung ventilation can be obtained and the SBT is an easy method to use for this purpose.
引用
收藏
页码:393 / 400
页数:8
相关论文
共 24 条
[1]   On the fluctuation in the composition of the alveolar air during the respiratory cycle in muscular exercise. [J].
Aitken, RS ;
Clark-Kennedy, AE .
JOURNAL OF PHYSIOLOGY-LONDON, 1928, 65 (04) :389-411
[2]   RESPIRATORY DEADSPACE MEASUREMENTS IN NEONATES WITH CONGENITAL DIAPHRAGMATIC-HERNIA [J].
ARNOLD, JH ;
BOWER, LK ;
THOMPSON, JE .
CRITICAL CARE MEDICINE, 1995, 23 (02) :371-375
[3]   INSPIRATORY FLOW-RATE AND DEAD SPACE IN DOGS [J].
BERDINE, GG ;
JOHNSON, JE ;
DALE, D ;
LEHR, JL .
JOURNAL OF APPLIED PHYSIOLOGY, 1990, 68 (03) :1228-1232
[4]  
Blanch L, 2006, MINERVA ANESTESIOL, V72, P577
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]  
Bohr C, 1887, CENTRALBLATT PHYSL, V1, P236
[7]  
Enghoff Henrik, 1938, UPSALA LAKAREFOREN FORHANDL, V44, P191
[8]   DIAGNOSIS OF PULMONARY-EMBOLISM BASED UPON ALVEOLAR DEAD SPACE ANALYSIS [J].
ERIKSSON, L ;
WOLLMER, P ;
OLSSON, CG ;
ALBRECHTSSON, U ;
LARUSDOTTIR, H ;
NILSSON, R ;
SJOGREN, A ;
JONSON, B .
CHEST, 1989, 96 (02) :357-362
[9]  
Fletcher R, 1990, Acta Anaesthesiol Scand Suppl, V94, P46