Preliminary evaluation of UF-5000 Body Fluid Mode for automated cerebrospinal fluid cell counting

被引:16
作者
Seghezzi, Michela [1 ]
Manenti, Barbara [1 ]
Previtali, Giulia [1 ]
Alessio, Maria Grazia [1 ]
Dominoni, Paola [1 ]
Buoro, Sabrina [1 ]
机构
[1] Hosp Papa Giovanni XXIII, Clin Chem Lab, Piazza OMS 1, I-24128 Bergamo, Italy
关键词
Sysmex UF-5000; Automated cell counting; Cerebrospinal fluid; Body fluids; BLOOD-CELLS; XE-5000; MODULE;
D O I
10.1016/j.cca.2017.08.026
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Cellular analysis of cerebrospinal fluid (CSF) provides important diagnostic information in various medical conditions. The aim of this study was to evaluate the application of Sysmex UF-5000 body fluid mode in cytometric analysis of CSF compared to Light Microscopic (LM). Methods: Eighty-one consecutive CSF samples were analyzed by UF-5000 body fluid mode and by LM. The study also included the evaluation of: limit of Blank (LoB), limit of Detection (LoD), limit of Quantitation (LoQ), carryover and linearity. Results: For total nucleated cells (TNC-UF) and white blood cells (WBC-UF) LoB, LoD and LoQ were 1 x 10(6) cells/L, 1.8 x 10(6) cells/L and 1.9 x 10(6) cells/L respectively. For red blood cells (RBC) LoB was 2 x 10(6) cells/L, LoD was 3.5 x 10(6) cells/L and LoQ was 14 x 10(6) cells/L respectively. Linearity was excellent, carryover was negligible. The agreement between UF-5000 body fluid mode parameters and manual cell counts was good in all CSF samples with bias ranged between -0.5 and 25.1 x 10(6) cells/L. The ROC curve analysis showed an area under curve of 0.99 for both TNC-UF and WBC-UF parameters. Conclusions: The UF-5000 body fluid mode offers rapid and accurate counts in clinically relevant concentration ranges, replacing the LM for most samples. However, in samples with abnormal cell counts or with abnormal scattergram the need for microscopic review remains.
引用
收藏
页码:133 / 138
页数:6
相关论文
共 31 条
  • [11] Automated Cerebrospinal Fluid Cell Counts Using the New Body Fluid Mode of Sysmex UF-1000i
    Buoro, Sabrina
    Esposito, Sara Apassiti
    Alessio, MariaGrazia
    Crippa, Alberto
    Ottomano, Cosimo
    Lippi, Giuseppe
    [J]. JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2016, 30 (05) : 381 - 391
  • [12] Analytical evaluation of Sysrnex UF-1000i for flow cytometric analysis of peritoneal fluid
    Buoro, Sabrina
    Gustinetti, Rosanna
    Dominoni, Paola
    Alessio, Maria Grazia
    Filisetti, Patrizia
    Luca, Maria Grazia
    Fagiuoli, Stefano
    Ottomano, Cosimo
    Lippi, Giuseppe
    [J]. CLINICAL BIOCHEMISTRY, 2012, 45 (15) : 1263 - 1265
  • [13] A multicenter evaluation of the Iris iQ200 automated urine microscopy analyzer Body Fluids Module and comparison with hemacytometer cell counts
    Butch, Anthony W.
    Wises, Porntip K.
    Wah, David T.
    Gornet, Terrie G.
    Fritsche, Herbert A.
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2008, 129 (03) : 445 - 450
  • [14] Flow Cytometric Characterization of Cerebrospinal Fluid Cells
    de Graaf, Marieke T.
    de Jongste, Arjen H. C.
    Kraan, Jaco
    Boonstra, Joke G.
    Smitt, Peter A. E. Sillevis
    Gratama, Jan W.
    [J]. CYTOMETRY PART B-CLINICAL CYTOMETRY, 2011, 80B (05) : 271 - 281
  • [15] Evaluation of the new body fluid mode on the Sysmex XE-5000 for counting leukocytes and erythrocytes in cerebrospinal fluid and other body fluids
    de Jonge, Robert
    Brouwer, Rob
    de Graaf, Marieke T.
    Luitwieler, Ronald L.
    Fleming, Cherina
    de Frankrijker-Merkestijn, Magda
    Smitt, Peter A. E. Sillevis
    Boonstra, Joke G.
    Lindemans, Jan
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (05) : 665 - 675
  • [16] When and How to Investigate the Patient with Headache
    De Luca, G. C.
    Bartleson, J. D.
    [J]. SEMINARS IN NEUROLOGY, 2010, 30 (02) : 131 - 144
  • [17] Parasitoses of the human central nervous system
    Finsterer, J.
    Auer, H.
    [J]. JOURNAL OF HELMINTHOLOGY, 2013, 87 (03) : 257 - 270
  • [18] Emergency department management of meningitis and encephalitis
    Fitch, Michael T.
    Abrahamian, Fredrick M.
    Moran, Gregory J.
    Talan, David A.
    [J]. INFECTIOUS DISEASE CLINICS OF NORTH AMERICA, 2008, 22 (01) : 33 - 52
  • [19] Clinical relevance and contemporary methods for counting blood cells in body fluids suspected of inflammatory disease
    Fleming, Cherina
    Russcher, Henk
    Lindemans, Jan
    de Jonge, Robert
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2015, 53 (11) : 1689 - 1706
  • [20] Validation of the body fluid module on the new Sysmex XN-1000 for counting blood cells in cerebrospinal fluid and other body fluids
    Fleming, Cherina
    Brouwer, Rob
    Lindemans, Jan
    de Jonge, Robert
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2012, 50 (10) : 1791 - 1798