Preliminary evaluation of UF-5000 Body Fluid Mode for automated cerebrospinal fluid cell counting

被引:16
作者
Seghezzi, Michela [1 ]
Manenti, Barbara [1 ]
Previtali, Giulia [1 ]
Alessio, Maria Grazia [1 ]
Dominoni, Paola [1 ]
Buoro, Sabrina [1 ]
机构
[1] Hosp Papa Giovanni XXIII, Clin Chem Lab, Piazza OMS 1, I-24128 Bergamo, Italy
关键词
Sysmex UF-5000; Automated cell counting; Cerebrospinal fluid; Body fluids; BLOOD-CELLS; XE-5000; MODULE;
D O I
10.1016/j.cca.2017.08.026
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Cellular analysis of cerebrospinal fluid (CSF) provides important diagnostic information in various medical conditions. The aim of this study was to evaluate the application of Sysmex UF-5000 body fluid mode in cytometric analysis of CSF compared to Light Microscopic (LM). Methods: Eighty-one consecutive CSF samples were analyzed by UF-5000 body fluid mode and by LM. The study also included the evaluation of: limit of Blank (LoB), limit of Detection (LoD), limit of Quantitation (LoQ), carryover and linearity. Results: For total nucleated cells (TNC-UF) and white blood cells (WBC-UF) LoB, LoD and LoQ were 1 x 10(6) cells/L, 1.8 x 10(6) cells/L and 1.9 x 10(6) cells/L respectively. For red blood cells (RBC) LoB was 2 x 10(6) cells/L, LoD was 3.5 x 10(6) cells/L and LoQ was 14 x 10(6) cells/L respectively. Linearity was excellent, carryover was negligible. The agreement between UF-5000 body fluid mode parameters and manual cell counts was good in all CSF samples with bias ranged between -0.5 and 25.1 x 10(6) cells/L. The ROC curve analysis showed an area under curve of 0.99 for both TNC-UF and WBC-UF parameters. Conclusions: The UF-5000 body fluid mode offers rapid and accurate counts in clinically relevant concentration ranges, replacing the LM for most samples. However, in samples with abnormal cell counts or with abnormal scattergram the need for microscopic review remains.
引用
收藏
页码:133 / 138
页数:6
相关论文
共 31 条
  • [1] [Anonymous], EP17A2A CLSI
  • [2] [Anonymous], CLIN CHEM LAB MED
  • [3] [Anonymous], 2006, H56A CLSI
  • [4] [Anonymous], 2003, EP06A CLSI
  • [5] [Anonymous], 2013, J MED BIOCHEM
  • [6] Evaluation of the XE-5000 for the automated analysis of blood cells in cerebrospinal fluid
    Boer, Klas
    Deufel, Thomas
    Reinhoefer, Mike
    [J]. CLINICAL BIOCHEMISTRY, 2009, 42 (7-8) : 684 - 691
  • [7] ICSH guidelines for the verification and performance of automated cell counters for body fluids
    Bourner, G.
    De la Salle, B.
    George, T.
    Tabe, Y.
    Baum, H.
    Culp, N.
    Keng, T. B.
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 (06) : 598 - 612
  • [8] Automated cerebrospinal fluid cell count - New reference ranges and evaluation of its clinical use in central nervous system infections
    Bremell, Daniel
    Mattsson, Niklas
    Wallin, Fredrik
    Henriksson, John
    Wall, Mariann
    Blennow, Kaj
    Zetterberg, Henrik
    Hagberg, Lars
    [J]. CLINICAL BIOCHEMISTRY, 2014, 47 (1-2) : 25 - 30
  • [9] Brown Wendy, 2003, Lab Hematol, V9, P155
  • [10] Evaluation of Mindray BC-6800 body fluid mode for automated cerebrospinal fluid cell counting
    Buoro, Sabrina
    Seghezzi, Michela
    Mecca, Tommaso
    Vavassori, Mauro
    Crippa, Alberto
    La Gioia, Antonio
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2016, 54 (11) : 1799 - 1810