The reproducibility of measurements using a standardization phantom for the evaluation of fractional anisotropy (FA) derived from diffusion tensor imaging (DTI)

被引:11
作者
Kimura, Mitsuhiro [1 ]
Yabuuchi, Hidetake [1 ]
Matsumoto, Ryoji [2 ]
Kobayashi, Koji [2 ]
Yamashita, Yasuo [2 ]
Nagatomo, Kazuya [1 ]
Mikayama, Ryoji [1 ]
Kamitani, Takeshi [3 ]
Sagiyama, Koji [3 ]
Yamasaki, Yuzo [3 ]
机构
[1] Kyushu Univ, Grad Sch Med Sci, Dept Hlth Sci, Higashi Ku, 3-1-1 Maidashi, Fukuoka 8128582, Japan
[2] Kyushu Univ Hosp, Dept Med Technol, Div Radiol, Higashi Ku, 3-1-1 Maidashi, Fukuoka 8128582, Japan
[3] Kyushu Univ, Grad Sch Med Sci, Dept Clin Radiol, Higashi Ku, 3-1-1 Maidashi, Fukuoka 8128582, Japan
基金
日本学术振兴会;
关键词
Magnetic resonance imaging; Diffusion tensor imaging; Standardization; Anisotropy; Quantitative evaluation; WHITE-MATTER; MRI; CONSTRUCTION; DESIGN;
D O I
10.1007/s10334-019-00776-w
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives It is necessary to standardize the examination procedure and diagnostic criteria of diffusion tensor imaging (DTI). Thus, the purpose of this study was to examine the reproducibility of measurements using a standardization phantom composed of different fibre materials with different fibre densities (FDs) for the evaluation of fractional anisotropy (FA) derived from DTI. Materials and methods Two types of fibre materials wrapped in heat-shrinkable tubes were used as fibre phantoms. We designed fibre phantoms with three different FDs of each fibre material. The standardization phantom was examined using DTI protocol six times a day, and each examination session was repeated once a month for 7 consecutive months. Fibre tracking was performed by setting regions of interest in the FA map, and FA was measured in each fibre phantom. Coefficients of variation (CVs) were used to evaluate the inter-examination reproducibility of FA values. Furthermore, Bland-Altman plots were used to evaluate the intra-operator reproducibility of FA measurements. Results All CVs for each fibre phantom were within 2% throughout the 7-month study of repeated DTI sessions. The high intra-operator reproducibility of the FA measurement was confirmed. Discussion High reproducibility of measurements using a standardization phantom for the evaluation of FA was achieved.
引用
收藏
页码:293 / 298
页数:6
相关论文
共 23 条
  • [1] Diffusion tensor imaging (DTI)-based white matter mapping in brain research: A review
    Assaf, Yaniv
    Pasternak, Ofer
    [J]. JOURNAL OF MOLECULAR NEUROSCIENCE, 2008, 34 (01) : 51 - 61
  • [2] MR DIFFUSION TENSOR SPECTROSCOPY AND IMAGING
    BASSER, PJ
    MATTIELLO, J
    LEBIHAN, D
    [J]. BIOPHYSICAL JOURNAL, 1994, 66 (01) : 259 - 267
  • [3] Inferring microstructural features and the physiological state of tissues from diffusion-weighted images
    Basser, PJ
    [J]. NMR IN BIOMEDICINE, 1995, 8 (7-8) : 333 - 344
  • [4] Diffusion-tensor MRI: theory, experimental design and data analysis - a technical review
    Basser, PJ
    Jones, DK
    [J]. NMR IN BIOMEDICINE, 2002, 15 (7-8) : 456 - 467
  • [5] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [6] Boujraf S, 2001, MAGN RESON MATER PHY, V13, P82
  • [7] Clinical applications of diffusion tensor imaging
    Dong, Q
    Welsh, RC
    Chenevert, TL
    Carlos, RC
    Maly-Sundgren, P
    Gomez-Hassan, DM
    Mukherji, SK
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 19 (01) : 6 - 18
  • [8] Novel multisection design of anisotropic diffusion phantoms
    Farrher, Ezequiel
    Kaffanke, Joachim
    Celik, A. Avdo
    Stoecker, Tony
    Grinberg, Farida
    Shah, N. Jon
    [J]. MAGNETIC RESONANCE IMAGING, 2012, 30 (04) : 518 - 526
  • [9] Simulation and experimental verification of the diffusion in an anisotropic fiber phantom
    Fieremans, Els
    De Deene, Yves
    Delputte, Steven
    Ozdemir, Mahir S.
    D'Asseler, Yves
    Vlassenbroeck, Jelle
    Deblaere, Karel
    Achten, Eric
    Lemahieu, Ignace
    [J]. JOURNAL OF MAGNETIC RESONANCE, 2008, 190 (02) : 189 - 199
  • [10] Quantitative evaluation of 10 tractography algorithms on a realistic diffusion MR phantom
    Fillard, Pierre
    Descoteaux, Maxime
    Goh, Alvina
    Gouttard, Sylvain
    Jeurissen, Ben
    Malcolm, James
    Ramirez-Manzanares, Alonso
    Reisert, Marco
    Sakaie, Ken
    Tensaouti, Fatima
    Yo, Ting
    Mangin, Jean-Francois
    Poupon, Cyril
    [J]. NEUROIMAGE, 2011, 56 (01) : 220 - 234