Development and validation of a thermal desorber gas chromatography method for determination of residual solvents in drug loaded albumin

被引:3
作者
Asfaw, Adissu Alemayehu [1 ,2 ]
Wolfs, Kris [1 ]
Van Schepdael, Ann [1 ]
Adams, Erwin [1 ]
机构
[1] Univ Leuven, Dept Pharmaceut & Pharmacol Sci, Pharmaceut Anal, KU Leuven, Herestr 49,O&N2,PB 923, B-3000 Leuven, Belgium
[2] Mekelle Univ, Coll Hlth Sci, Sch Pharm, Mekelle, Ethiopia
关键词
Thermal desorber; Gas chromatography; Albumin; Residual solvents; DELIVERY; WATER;
D O I
10.1016/j.jpba.2019.113032
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The conventional approach for residual solvent (RS) analysis is headspace-gas chromatography (HS-GC). This starts from a homogenous sample solution and is based on the equilibrium of the analyte between the sample and the gas phase. Unfortunately, aqueous solutions of albumin form irreversible hydrophobic aggregates when heated above 50 degrees C. Consequently, the use of HS-GC for RS analysis in albumin becomes problematic due to the presence of an additional solid phase in the HS vial. In this work, a method using a thermal desorber (TD) combined with GC was developed for the determination of RS in drug loaded albumin. Samples were immobilized between two double layers of quartz filter (QF) in a polytetrafluoroethylene (PTFE) insert which was placed in an empty desorption tube prior to TD-GC analysis. The liquid standard mix consisted of ethanol (EtOH), acetone (Ace), dichloromethane (DCM) and chloroform (Chl) dissolved in toluene. Offline liquid calibration (OLC) was applied by introducing 2 mu L of the standard mix under counter flow of an inert gas into the TD tube containing a mixed bed of mesoporous silica (MPSi) immobilized between two double layers of QF. The OLC results were verified using the inline liquid calibration (ILC) approach based on a heated GC injector installed on the TD. The validation results revealed that the proposed method has good recovery (> 98 %). R-2-values (>= 0.998) indicated good linearity over a wide range. The limit of detection (LOD) and limit of quantification (LOQ) were found to be 0.01 and 0.04 mu g on tube, respectively. Repeatability of the method was reported as RSD-values and they were lower than 3 %. A method based on the complete enzymatic digestion of albumin combined with conventional HS-GC was developed to verify the completeness of release of the RS from the albumin. Both the TD-GC and HS-GC methods were applied for the determination of EtOH and DCM in two different albumin samples loaded with experimental drugs. Statistical comparison indicated that there was no significant difference (p > 0.05) between the two methods. However, the HS-GC method following enzymatic degradation is much more expensive and time consuming. (C) 2019 Elsevier B.V. All rights reserved.
引用
收藏
页数:7
相关论文
共 22 条
  • [1] [Anonymous], 2009, GUID RES SOLV
  • [2] [Anonymous], 2017, EUROPEAN PHARMACOPOE
  • [3] Exploration of the problems and solutions related to reference introduction prior to calibration of thermal desorber-gas chromatography
    Asfaw, Adissu Alemayehu
    Van der Veken, Matthias
    Wolfs, Kris
    Van Schepdael, Ann
    Adams, Erwin
    [J]. JOURNAL OF SEPARATION SCIENCE, 2019, 42 (17) : 2816 - 2825
  • [4] Determination of residual dimethylsulphoxide in drug loaded gelatin using thermal desorber - gas chromatography
    Asfaw, Adissu Alemayehu
    Wolfs, Kris
    Van Schepdael, Ann
    Adams, Erwin
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2018, 153 : 193 - 198
  • [5] Thermal desorption-Gas chromatographic methodology for the determination of residual solvents in mesoporous silica
    Asfaw, Adissu Alemayehu
    Wolfs, Kris
    Van Schepdael, Ann
    Adams, Erwin
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2017, 1500 : 160 - 166
  • [6] Kinetics of Thermal Denaturation and Aggregation of Bovine Serum Albumin
    Borzova, Vera A.
    Markossian, Kira A.
    Chebotareva, Natalia A.
    Kleymenov, Sergey Yu.
    Poliansky, Nikolay B.
    Muranov, Konstantin O.
    Stein-Margolina, Vita A.
    Shubin, Vladimir V.
    Markov, Denis I.
    Kurganov, Boris I.
    [J]. PLOS ONE, 2016, 11 (04):
  • [7] A generic static headspace gas chromatography method for determination of residual solvents in drug substance
    Cheng, Chang
    Liu, Shaorong
    Mueller, Bradford J.
    Yan, Zimeng
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (41) : 6413 - 6421
  • [8] Chick H, 1911, J PHYSIOL-LONDON, V43, P1
  • [9] Liquid paraffin as new dilution medium for the analysis of high boiling point residual solvents with static headspace-gas chromatography
    D'Autry, Ward
    Zheng, Chao
    Bugalama, John
    Wolfs, Kris
    Hoogmartens, Jos
    Adams, Erwin
    Wang, Bochu
    Van Schepdael, Ann
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2011, 55 (05) : 1017 - 1023
  • [10] Albumin-based nanoparticles as potential controlled release drug delivery systems
    Elzoghby, Ahmed O.
    Samy, Wael M.
    Elgindy, Nazik A.
    [J]. JOURNAL OF CONTROLLED RELEASE, 2012, 157 (02) : 168 - 182