The effects of incomplete breath-holding on 3D MR image quality

被引:65
作者
Maki, JH [1 ]
Chenevert, TL [1 ]
Prince, MR [1 ]
机构
[1] Univ Michigan Hosp, Dept Radiol, Div MRI, Ann Arbor, MI 48109 USA
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1997年 / 7卷 / 06期
关键词
MRI; phase-encoding order; motion correction; MR artifact; MR angiography; 3D imaging;
D O I
10.1002/jmri.1880070628
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to investigate how fast three-dimensional (3D) MR image quality is affected by breath-holding and to develop an optimal breath-holding strategy that minimizes artifact in the event of an incomplete breath-hold. A computer model was developed to study variable-duration breath-holds during fast 3D imaging. Modeling was validated by 3D gradient-echo imaging performed on 10 volunteers. Signal-to-noise ratio (SNR) and image blur were measured for both simulated and clinical images. Insights gained were applied to clinical 3D gadolinium-enhanced MR angiography. Breath-holding significantly improved abdominal 3D MR image quality. Most of this benefit could be achieved with a breath-hold fraction of 50% if it occurred during acquisition of central k space. Breath-holding during peripheral k-space acquisition, however, had no significant benefit. Respiratory motion artifact on fast 3D MRI occurring when a patient fails to suspend respiration for the entire scan duration can be minimized by collecting central k space first (centric acquisition) so that premature breathing affects only the acquisition of peripheral k space.
引用
收藏
页码:1132 / 1139
页数:8
相关论文
共 22 条
  • [1] MULTISLICE, BREATHHOLD IMAGING OF THE LUNG WITH SUBMILLISECOND ECHO TIMES
    ALSOP, DC
    HATABU, H
    BONNET, M
    LISTERUD, J
    GEFTER, W
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (05) : 678 - 682
  • [2] CENTRIC PHASE-ENCODING ORDER IN 3-DIMENSIONAL MP-RAGE SEQUENCES - APPLICATION TO ABDOMINAL IMAGING
    BAMPTON, AEH
    RIEDERER, SJ
    KORIN, HW
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (03): : 327 - 334
  • [3] THE EFFECT OF RESPIRATION ON THE CONTRAST AND SHARPNESS OF LIVER-LESIONS IN MRI
    BUTTS, K
    RIEDERER, SJ
    EHMAN, RL
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (01) : 1 - 7
  • [4] ADAPTIVE TECHNIQUE FOR HIGH-DEFINITION MR IMAGING OF MOVING STRUCTURES
    EHMAN, RL
    FELMLEE, JP
    [J]. RADIOLOGY, 1989, 173 (01) : 255 - 263
  • [5] Foo T. F., 1996, P SOC MAGN RES 1996, P453
  • [6] HAACKE E M, 1986, Magnetic Resonance Imaging, V4, P359, DOI 10.1016/0730-725X(86)91046-5
  • [7] PSEUDO-GATING - ELIMINATION OF PERIODIC MOTION ARTIFACTS IN MAGNETIC-RESONANCE-IMAGING WITHOUT GATING
    HAACKE, EM
    LENZ, GW
    NELSON, AD
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1987, 4 (02) : 162 - 174
  • [8] MEASUREMENT OF SIGNAL INTENSITIES IN THE PRESENCE OF NOISE IN MR IMAGES
    HENKELMAN, RM
    [J]. MEDICAL PHYSICS, 1985, 12 (02) : 232 - 233
  • [9] Breath-hold ultrafast three-dimensional gadolinium-enhanced MR angiography of the aorta and the renal and other visceral abdominal arteries
    Holland, GA
    Dougherty, L
    Carpenter, JP
    Golden, MA
    Gilfeather, M
    Slossman, F
    Schnall, MD
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 166 (04) : 971 - 981
  • [10] THE IMPORTANCE OF PHASE-ENCODING ORDER IN ULTRA-SHORT TR SNAPSHOT MR IMAGING
    HOLSINGER, AE
    RIEDERER, SJ
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1990, 16 (03) : 481 - 488