Factors Influencing the Reliability of Autorefractometry After LASIK for Myopia and Myopic Astigmatism

被引:2
作者
Mirshahi, Alireza [3 ]
Wesemann, Wolfgang [4 ]
Buehren, Jens [1 ]
Kohnen, Thomas [1 ,2 ]
机构
[1] Goethe Univ Frankfurt, Dept Ophthalmol, D-60590 Frankfurt, Germany
[2] Baylor Coll Med, Cullen Eye Inst, Houston, TX 77030 USA
[3] Johannes Gutenberg Univ Mainz, Dept Ophthalmol, Med Ctr, Mainz, Germany
[4] Sch Optometry, Cologne, Germany
关键词
HIGHER-ORDER ABERRATIONS; IN-SITU KERATOMILEUSIS; SUBJECTIVE REFRACTION; PHOTOREFRACTIVE KERATECTOMY; INTRAOCULAR LENSES; LASER; CORNEA; DIAMETER; ERRORS; POWER;
D O I
10.1016/j.ajo.2010.06.027
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To study the factors influencing the reliability (accuracy) of autorefractometry before and after laser in situ keratomileusis (LASIK) for myopia and myopic astigmatism. DESIGN: Retrospective case series. METHODS: A total of 250 consecutive eyes (132 patients, mean age 37 years) were included from the Department of Ophthalmology, J.W. Goethe University, Frankfurt, Germany. The mean preoperative spherical equivalent of the subjective refraction (SR) was -6.59 diopters (D) ( -1.38 to -15.13). The patient data were reviewed before and 1 month after LASIK, including SR, objective automated refraction (AR), and visual acuity. Subgroup analysis was performed with regard to the preoperative myopia and the excimer laser optical zone (OZ). The Holm-Sidak and Wilcoxon matched-pairs tests were used for statistical analysis. RESULTS: The correlation coefficient between AR und SR is r = 0.98 before LASIK versus r = 0.79 afterwards (P < .001). The mean difference between the spherical equivalents (DSE) is +0.13 +/- 0.51 D preoperatively versus -0.30 +/- 0.58 D after LASIK (P < .001). With a small optical zone size (5.0-5.5 mm) the postoperative difference is -0.61 D, versus -0.36 D for OZ 5.6-6.0, versus -0.16 for OZ 6.1-7.0 mm. With a rising preoperative amount of myopia, the postoperative AR results become increasingly more myopic than the SR. CONCLUSIONS: Following LASIK, autorefractometry is less accurate than in nonoperated eyes. The reliability of the AR is influenced by the OZ and the preoperative amount of myopia, with a small OZ and high myopia resulting in a greater difference between AR and SR and with the AR determining more myopic results. (Am J Ophthalmol 2010;150:774-779. (C) 2010 by Elsevier Inc. All rights reserved.)
引用
收藏
页码:774 / 779
页数:6
相关论文
共 29 条
  • [1] The effect of optical zone decentration on lower- and higher-order aberrations after Photorefractive keratectomy in a cat model
    Buehren, Jens
    Yoon, Geunyoung
    Kenner, Shawn
    MacRae, Scott
    Huxlin, Krystel
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2007, 48 (12) : 5806 - 5814
  • [2] Factors affecting the change in lower-order and higher-order aberrations after wavefront-guided laser in situ keratomileusis for myopia with the Zyoptix 3.1 system
    Buehren, Jens
    Kohnen, Thomas
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (07) : 1166 - 1174
  • [3] Influence of pupil and optical zone diameter on higher-order aberrations after wavefront-guided myopic LASIK
    Bühren, J
    Kühne, C
    Kohnen, T
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (12) : 2272 - 2280
  • [4] DIEPES H, 2004, REFRAKTIONSBESTIMMUN, P184
  • [5] Night vision disturbances after corneal refractive surgery
    Fan-Paul, NI
    Li, J
    Miller, JS
    Florakis, GJ
    [J]. SURVEY OF OPHTHALMOLOGY, 2002, 47 (06) : 533 - 546
  • [6] Goss D A, 1996, J Am Optom Assoc, V67, P619
  • [7] A comparative analysis of five methods of determining corneal refractive power in eyes that have undergone myopic laser in situ keratomileusis
    Hamed, AH
    Wang, L
    Misra, M
    Koch, DD
    [J]. OPHTHALMOLOGY, 2002, 109 (04) : 651 - 658
  • [8] Wavefront aberration and its relationship to the accommodative stimulus-response function in myopic subjects
    Hazel, CA
    Cox, MJ
    Strang, NC
    [J]. OPTOMETRY AND VISION SCIENCE, 2003, 80 (02) : 151 - 158
  • [9] Comparison of corneal higher-order aberrations induced by myopic and hyperopic LASIK
    Kohnen, T
    Mahmoud, K
    Bühren, J
    [J]. OPHTHALMOLOGY, 2005, 112 (10) : 1692 - 1698
  • [10] Intraocular lenses for the correction of refraction errors. Part II. Phakic posterior chamber lenses and refractive lens exchange with posterior chamber lens implantation
    Kohnen, T
    Kasper, T
    Terzi, E
    [J]. OPHTHALMOLOGE, 2005, 102 (11): : 1105 - 1117