Chronic recording and electrochemical performance of Utah microelectrode arrays implanted in rat motor cortex

被引:44
作者
Black, Bryan J. [1 ]
Kanneganti, Aswini [1 ]
Joshi-Imre, Alexandra [1 ]
Rihani, Rashed [1 ]
Chakraborty, Bitan [1 ]
Abbott, Justin [1 ]
Pancrazio, Joseph J. [1 ]
Cogan, Stuart F. [1 ]
机构
[1] Univ Texas Dallas, Dept Bioengn, 800 W Campbell Rd,Bioengn & Sci Bldg 13-601, Richardson, TX 75080 USA
关键词
chronic recording; electrical stimulation; rodent models; Utah electrode array; INTRACORTICAL MICROSTIMULATION; BRAIN-TISSUE; STIMULATION; RECOVERY; STROKE; ADULT; NEUROSTIMULATION; PLASTICITY; IMPEDANCE; BLIND;
D O I
10.1152/jn.00181.2018
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Multisite implantable electrode arrays serve as a tool to understand cortical network connectivity and plasticity. Furthermore, they enable electrical stimulation to drive plasticity, study motor/sensory mapping, or provide network input for controlling brain-computer interfaces. Neurobehavioral rodent models are prevalent in studies of motor cortex injury and recovery as well as restoration of auditory/visual cues due to their relatively low cost and ease of training. Therefore, it is important to understand the chronic performance of relevant electrode arrays in rodent models. In this report, we evaluate the chronic recording and electrochemical performance of 16-channel Utah electrode arrays. the current state-of-the-art in pre-/clinical cortical recording and stimulation, in rat motor cortex over a period of 6 mo. The single-unit active electrode yield decreased from 52.8 +/- 10.0 (week 1) to 13.4 +/- 5.1% (week 24). Similarly, the total number of single units recorded on all electrodes across all arrays decreased from 106 to 15 over the same time period. Parallel measurements of electrochemical impedance spectra and cathodic charge storage capacity exhibited significant changes in electrochemical characteristics consistent with development of electrolyte leakage pathways over time. Additionally, measurements of maximum cathodal potential excursion indicated that only a relatively small fraction of electrodes (10-35% at 1 and 24 wk postimplantation) were capable of delivering relevant currents (20 mu A at 4 nC/ph) without exceeding negative or positive electrochemical potential limits. In total, our findings suggest mainly abiotic failure modes, including mechanical wire breakage as well as degradation of conducting and insulating substrates. NEW & NOTEWORTHY Multisite implantable electrode arrays serve as a tool to record cortical network activity and enable electrical stimulation to drive plasticity or provide network feedback. The use of rodent models in these fields is prevalent. We evaluated chronic recording and electrochemical performance of 16-channel Utah electrode arrays in rat motor cortex over a period of 6 mo. We primarily observed abiotic failure modes suggestive of mechanical wire breakage and/or degradation of insulation.
引用
收藏
页码:2083 / 2090
页数:8
相关论文
共 40 条
[11]   Temporal plasticity in auditory cortex improves neural discrimination of speech sounds [J].
Engineer, Crystal T. ;
Shetake, Jai A. ;
Engineer, Navzer D. ;
Vrana, Will A. ;
Wolf, Jordan T. ;
Kilgard, Michael P. .
BRAIN STIMULATION, 2017, 10 (03) :543-552
[12]   Speech sound discrimination training improves auditory cortex responses in a rat model of autism [J].
Engineer, Crystal T. ;
Centanni, Tracy M. ;
lm, Kwok W. ;
Kilgard, Michael P. .
FRONTIERS IN SYSTEMS NEUROSCIENCE, 2014, 8
[13]   Animal models of ischemic stroke and their application in clinical research [J].
Fluri, Felix ;
Schuhmann, Michael K. ;
Kleinschnitz, Christoph .
DRUG DESIGN DEVELOPMENT AND THERAPY, 2015, 9 :3445-3454
[14]   Cognitive recovery in the aged rat after stroke and anti-Nogo-A immunotherapy [J].
Gillani, Rebecca L. ;
Tsai, Shih-Yen ;
Wallace, Douglas G. ;
O'Brien, Timothy E. ;
Arhebamen, Ebinehita ;
Tole, Mateo ;
Schwab, Martin E. ;
Kartje, Gwendolyn L. .
BEHAVIOURAL BRAIN RESEARCH, 2010, 208 (02) :415-424
[15]   Comparative study on the insertion Behavior of cerebral microprobes [J].
Hosseini, Neda Haj ;
Hoffmann, Ruediger ;
Kisban, Sebastian ;
Stieglitz, Thomas ;
Paul, Oliver ;
Ruther, Patrick .
2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, :4711-+
[16]   Virtual typing by people with tetraplegia using a self-calibrating intracortical brain-computer interface [J].
Jarosiewicz, Beata ;
Sarma, Anish A. ;
Bacher, Daniel ;
Masse, Nicolas Y. ;
Simeral, John D. ;
Sorice, Brittany ;
Oakley, Erin M. ;
Blabe, Christine ;
Pandarinath, Chethan ;
Gilja, Vikash ;
Cash, Sydney S. ;
Eskandar, Emad N. ;
Friehs, Gerhard ;
Henderson, Jaimie M. ;
Shenoy, Krishna V. ;
Donoghue, John P. ;
Hochberg, Leigh R. .
SCIENCE TRANSLATIONAL MEDICINE, 2015, 7 (313)
[17]   Progress towards biocompatible intracortical microelectrodes for neural interfacing applications [J].
Jorfi, Mehdi ;
Skousen, John L. ;
Weder, Christoph ;
Capadona, Jeffrey R. .
JOURNAL OF NEURAL ENGINEERING, 2015, 12 (01)
[18]   Electrical Performance of Penetrating Microelectrodes Chronically Implanted in Cat Cortex [J].
Kane, Sheryl R. ;
Cogan, Stuart F. ;
Ehrlich, Julia ;
Plante, Timothy D. ;
McCreery, Douglas B. ;
Troyk, Philip R. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2013, 60 (08) :2153-2160
[19]   Relationship between intracortical electrode design and chronic recording function [J].
Karumbaiah, Lohitash ;
Saxena, Tarun ;
Carlson, David ;
Patil, Ketki ;
Patkar, Radhika ;
Gaupp, Eric A. ;
Betancur, Martha ;
Stanley, Garrett B. ;
Carin, Lawrence ;
Bellamkonda, Ravi V. .
BIOMATERIALS, 2013, 34 (33) :8061-8074
[20]   Vagus Nerve Stimulation Delivered During Motor Rehabilitation Improves Recovery in a Rat Model of Stroke [J].
Khodaparast, Navid ;
Hays, Seth A. ;
Sloan, Andrew M. ;
Fayyaz, Tabbassum ;
Hulsey, Daniel R. ;
Rennaker, Robert L., II ;
Kilgard, Michael P. .
NEUROREHABILITATION AND NEURAL REPAIR, 2014, 28 (07) :698-706