Chronic recording and electrochemical performance of Utah microelectrode arrays implanted in rat motor cortex

被引:44
作者
Black, Bryan J. [1 ]
Kanneganti, Aswini [1 ]
Joshi-Imre, Alexandra [1 ]
Rihani, Rashed [1 ]
Chakraborty, Bitan [1 ]
Abbott, Justin [1 ]
Pancrazio, Joseph J. [1 ]
Cogan, Stuart F. [1 ]
机构
[1] Univ Texas Dallas, Dept Bioengn, 800 W Campbell Rd,Bioengn & Sci Bldg 13-601, Richardson, TX 75080 USA
关键词
chronic recording; electrical stimulation; rodent models; Utah electrode array; INTRACORTICAL MICROSTIMULATION; BRAIN-TISSUE; STIMULATION; RECOVERY; STROKE; ADULT; NEUROSTIMULATION; PLASTICITY; IMPEDANCE; BLIND;
D O I
10.1152/jn.00181.2018
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Multisite implantable electrode arrays serve as a tool to understand cortical network connectivity and plasticity. Furthermore, they enable electrical stimulation to drive plasticity, study motor/sensory mapping, or provide network input for controlling brain-computer interfaces. Neurobehavioral rodent models are prevalent in studies of motor cortex injury and recovery as well as restoration of auditory/visual cues due to their relatively low cost and ease of training. Therefore, it is important to understand the chronic performance of relevant electrode arrays in rodent models. In this report, we evaluate the chronic recording and electrochemical performance of 16-channel Utah electrode arrays. the current state-of-the-art in pre-/clinical cortical recording and stimulation, in rat motor cortex over a period of 6 mo. The single-unit active electrode yield decreased from 52.8 +/- 10.0 (week 1) to 13.4 +/- 5.1% (week 24). Similarly, the total number of single units recorded on all electrodes across all arrays decreased from 106 to 15 over the same time period. Parallel measurements of electrochemical impedance spectra and cathodic charge storage capacity exhibited significant changes in electrochemical characteristics consistent with development of electrolyte leakage pathways over time. Additionally, measurements of maximum cathodal potential excursion indicated that only a relatively small fraction of electrodes (10-35% at 1 and 24 wk postimplantation) were capable of delivering relevant currents (20 mu A at 4 nC/ph) without exceeding negative or positive electrochemical potential limits. In total, our findings suggest mainly abiotic failure modes, including mechanical wire breakage as well as degradation of conducting and insulating substrates. NEW & NOTEWORTHY Multisite implantable electrode arrays serve as a tool to record cortical network activity and enable electrical stimulation to drive plasticity or provide network feedback. The use of rodent models in these fields is prevalent. We evaluated chronic recording and electrochemical performance of 16-channel Utah electrode arrays in rat motor cortex over a period of 6 mo. We primarily observed abiotic failure modes suggestive of mechanical wire breakage and/or degradation of insulation.
引用
收藏
页码:2083 / 2090
页数:8
相关论文
共 40 条
  • [1] Adult Cortical Plasticity Studied with Chronically Implanted Electrode Arrays
    Abe, Hiroshi
    McManus, Justin N. J.
    Ramalingam, Nirmala
    Li, Wu
    Marik, Sally A.
    Borgloh, Stephan Meyer zum Alten
    Gilbert, Charles D.
    [J]. JOURNAL OF NEUROSCIENCE, 2015, 35 (06) : 2778 - 2790
  • [2] Scanning electron microscopy of chronically implanted intracortical microelectrode arrays in non-human primates
    Barrese, James C.
    Aceros, Juan
    Donoghue, John P.
    [J]. JOURNAL OF NEURAL ENGINEERING, 2016, 13 (02)
  • [3] Failure mode analysis of silicon-based intracortical microelectrode arrays in non-human primates
    Barrese, James C.
    Rao, Naveen
    Paroo, Kaivon
    Triebwasser, Corey
    Vargas-Irwin, Carlos
    Franquemont, Lachlan
    Donoghue, John P.
    [J]. JOURNAL OF NEURAL ENGINEERING, 2013, 10 (06)
  • [4] Review: Human Intracortical Recording and Neural Decoding for Brain-Computer Interfaces
    Brandman, David M.
    Cash, Sydney S.
    Hochberg, Leigh R.
    [J]. IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2017, 25 (10) : 1687 - 1696
  • [5] Unique electrophysiological and impedance signatures between encapsulation types: An analysis of biological Utah array failure and benefit of a biomimetic coating in a rat model
    Cody, Patrick A.
    Eles, James R.
    Lagenaur, Carl F.
    Kozai, Takashi D. Y.
    Cui, X. Tracy
    [J]. BIOMATERIALS, 2018, 161 : 117 - 128
  • [6] Neural stimulation and recording electrodes
    Cogan, Stuart F.
    [J]. ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 2008, 10 : 275 - 309
  • [7] The influence of electrolyte composition on the in vitro charge-injection limits of activated iridium oxide (AIROF) stimulation electrodes
    Cogan, Stuart F.
    Troyk, Philip R.
    Ehrlich, Julia
    Gasbarro, Christina M.
    Plante, Timothy D.
    [J]. JOURNAL OF NEURAL ENGINEERING, 2007, 4 (02) : 79 - 86
  • [8] Crawley JN, 2003, COMPARATIVE MED, V53, P140
  • [9] ARTIFICIAL VISION FOR THE BLIND BY ELECTRICAL-STIMULATION OF THE VISUAL-CORTEX
    DOBELLE, WH
    QUEST, DO
    ANTUNES, JL
    ROBERTS, TS
    GIRVIN, JP
    [J]. NEUROSURGERY, 1979, 5 (04) : 521 - 527
  • [10] FACTORS INFLUENCING THE BIOCOMPATIBILITY OF INSERTABLE SILICON MICROSHAFTS IN CEREBRAL-CORTEX
    EDELL, DJ
    TOI, VV
    MCNEIL, VM
    CLARK, LD