摘要
This paper presents a novel Peristaltic Micro Pump with a twofold intrinsic Normally-Closed mechanism (PMP-NC2). It is based on a modular concept with a low cost disposable microfluidic chip and a re-usable actuator unit. The normally-closed mechanism prevents fluid flow without power consumption in the stand-by mode for pressures up to 150 kPa. Together with the backpressure independent pumping performance up to 80 kPa this results in a fail-safe and precise micropump. The addressed range of pump rates is up to 100 mu L/min. Due to the symmetrical setup, bi-directional pumping is possible and also air bubble tolerance is demonstrated. This results from the very low dead volume per pump chamber of only 230 nL, which leads to a complete transport of the air bubbles through the pump.
引用
收藏
页码:234 / +
页数:2
相关论文
共 6 条
  • [1] [Anonymous], SENSORS ACTUATORS
  • [2] Design and simulation of an implantable medical drug delivery system using microelectromechanical systems technology
    Cao, L
    Mantell, S
    Polla, D
    [J]. SENSORS AND ACTUATORS A-PHYSICAL, 2001, 94 (1-2) : 117 - 125
  • [3] NORMALLY CLOSED MICROVALVE AND MICROPUMP FABRICATED ON A SILICON-WAFER
    ESASHI, M
    SHOJI, S
    NAKANO, A
    [J]. SENSORS AND ACTUATORS, 1989, 20 (1-2): : 163 - 169
  • [4] Shinohara J., 2000, Proceedings IEEE Thirteenth Annual International Conference on Micro Electro Mechanical Systems (Cat. No.00CH36308), P86, DOI 10.1109/MEMSYS.2000.838495
  • [5] PIEZOELECTRIC MICROPUMP WITH 3 VALVES WORKING PERISTALTICALLY
    SMITS, JG
    [J]. SENSORS AND ACTUATORS A-PHYSICAL, 1990, 21 (1-3) : 203 - 206
  • [6] Normally-closed peristaltic micropump with re-usable actuator and disposable fluidic chip
    Trenkle, F.
    Haeberle, S.
    Zengerle, R.
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2011, 154 (02): : 137 - 141