In-vivo X-Ray micro-imaging and micro-CT with the Medipix2 semiconductor detector at UniAndes

被引:0
作者
Caicedo, I. [1 ]
Avila, C. [1 ]
Gomez, B. [1 ]
Bula, C. [1 ]
Roa, C. [1 ]
Sanabria, J. [1 ]
机构
[1] Univ Los Andes, Bogota, Colombia
来源
IX LATIN AMERICAN SYMPOSIUM ON NUCLEAR PHYSICS AND APPLICATIONS | 2012年 / 1423卷
关键词
Medipix; Pixel detectors; X-ray Imaging; In vivo X-ray micro-Imaging; Micro-CT;
D O I
10.1063/1.3688831
中图分类号
O59 [应用物理学];
学科分类号
摘要
This poster contains the procedure to obtain micro-CTs and to image moving samples using the Medipix2 detector, with its corresponding results. The high granularity of the detector makes it suitable for micro-CT. We used commercial software (Octopus) to do the 3D reconstruction of the samples in the first place, and we worked on modifying free reconstruction software afterwards. Medipix has a very fast response ( similar to hundreds of nanoseconds) and high sensibility. These features allow obtaining nearly in-vivo high resolution (55m * 55m) images. We used an exposure time of 0.1 s for each frame, and the resulting images were animated. The High Energy Physics Group at UniAndes is a member of the Medipix3 collaboration. Its research activities are focused on developing set-ups for biomedical applications and particle tracking using the Medipix2 and Timepix detectors, and assessing the feasibility of the Medipix3 detector for future applications.
引用
收藏
页数:4
相关论文
共 50 条
[41]   Micro-imaging performance of multilayers used as monochromators for coherent hard X-ray synchrotron radiation [J].
Rack, A. ;
Weitkamp, T. ;
Riotte, M. ;
Rack, T. ;
Dietsch, R. ;
Holz, T. ;
Kraemer, M. ;
Siewert, F. ;
Meduna, M. ;
Morawe, Ch. ;
Cloetens, P. ;
Ziegler, E. .
ADVANCES IN X-RAY/EUV OPTICS AND COMPONENTS V, 2010, 7802
[42]   Development of two-dimensional beam hardening correction for X-ray micro-CT [J].
Davis, Graham R. .
JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2022, 30 (05) :863-874
[43]   Application of X-ray micro-CT for micro-structural characterization of APCVD deposited SiC coatings on graphite conduit [J].
Agrawal, A. K. ;
Sarkar, P. S. ;
Singh, B. ;
Kashyap, Y. S. ;
Rao, P. T. ;
Sinha, A. .
APPLIED RADIATION AND ISOTOPES, 2016, 108 :133-142
[44]   Micro-CTvlab : A web based virtual gallery of biological specimens using X-ray microtomography (micro-CT) [J].
Keklikoglou, Kleoniki ;
Faulwetter, Sarah ;
Chatzinikolaou, Eva ;
Michalakis, Nikitas ;
Filiopoulou, Irene ;
Minadakis, Nikos ;
Panteri, Emmanouela ;
Perantinos, George ;
Gougousis, Alexandros ;
Arvanitidis, Christos .
BIODIVERSITY DATA JOURNAL, 2016, 4
[45]   Development of 36M-pixel x-ray detector for large field of view and high-resolution micro-CT [J].
Umetani, Keiji ;
Kawata, Yoshiki ;
Niki, Noboru .
OPTOELECTRONIC IMAGING AND MULTIMEDIA TECHNOLOGY IV, 2016, 10020
[46]   Material Decomposition in Low-Energy Micro-CT Using a Dual-Threshold Photon Counting X-Ray Detector [J].
Solem, Rasmus ;
Dreier, Till ;
Goncalves, Isabel ;
Bech, Martin .
FRONTIERS IN PHYSICS, 2021, 9 (09)
[47]   In vivo micro-CT imaging of liver lesions in small animal models [J].
Martiniova, Lucia ;
Schimel, Daniel ;
Lai, Edwin W. ;
Limpuangthip, Andrea ;
Kvetnansky, Richard ;
Pacak, Karel .
METHODS, 2010, 50 (01) :20-25
[48]   Exploring VivoVist™ for Ex Vivo Micro-CT Imaging of Mouse Vasculature [J].
Badea, C. T. ;
Nadkarni, R. ;
Allphin, A. ;
Qi, Y. ;
Clark, D. P. ;
Hainfeld, J. F. .
MEDICAL IMAGING 2025: CLINICAL AND BIOMEDICAL IMAGING, 2025, 13410
[49]   Low-dose, 4D myocardial perfusion with x-ray micro-CT [J].
Clark, D. P. ;
Badea, C. T. .
MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
[50]   Local variations in bone mineral density: A comparison of OCT vs X-ray micro-CT [J].
Uggryumova, Nadya ;
Stevens-Smith, Jerma ;
Scutt, Andrew ;
Matcher, Stephen J. .
COHERENCE DOMAIN OPTICAL METHODS AND OPTICAL COHERENCE TOMOGRAPHY IN BIOMEDICINE XII, 2008, 6847