Evaluation of the Sinmed Mastercouch® as replacement for a standard couch

被引:23
作者
Gillis, S [1 ]
Bral, S [1 ]
De Wagter, C [1 ]
Derie, C [1 ]
Paelinck, L [1 ]
Van Vaerenbergh, K [1 ]
Coghe, M [1 ]
De Meerleer, G [1 ]
De Neve, W [1 ]
机构
[1] Ghent Univ Hosp, Dept Radiotherapy, B-9000 Ghent, Belgium
关键词
carbon-fibre; treatment couch; geometric comparison; attenuation; build-up; skin dose;
D O I
10.1016/j.radonc.2005.04.013
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Introducing into practice a new carbon-fibre couch necessitates its evaluation with regard to the present clinical situation. Materials and methods: In this study, a geometric and dosimetric evaluation has been made for the. Sinmed Mastercouch((R)) as a replacement for the Elekta C-arm couch with Mylar((R))-tennis racket combination. Geometrically, feasible gantry angles with regard to collision and beam intersection were discriminated as function of isocentric table rotation for 10 treatment isocentres. Dosimetrically, the build-up distortion and attenuation by the aforementioned tabletops and a carbon-fibre tabletop of an Elekta Precise simulator was measured. Finally, the clinical implications of these influences were assessed for a 3-field prostate treatment in three configurations: Mastercouch((R)), C-arm couch and no-intersection situation. Results: With regard to collision, the largest advantages are observed for the Mastercouch((R)) with the U-shaped treatment module compared to the C-arm couch for isocentres located in the upper part of the body, thanks to its shape and the absence of any metal. Dosimetrically, one has to take into account the build-up loss and attenuation by beam intersection with Mastercouch((R)) and the carbon-fibre edges of the tennis racket (C-arm couch). The clinical relevance of these dosimetric aspects depends on the dose delivered by the compromised beams. (c) 2005 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:227 / 236
页数:10
相关论文
共 20 条
[1]   A collision prevention software tool for complex three-dimensional isocentric set-ups [J].
Beange, I ;
Nisbet, A .
BRITISH JOURNAL OF RADIOLOGY, 2000, 73 (869) :537-541
[2]   Intensity-modulated radiotherapy as primary treatment for prostate cancer: Acute toxicity in 114 patients [J].
De Meerleer, G ;
Vakaet, L ;
Meersschout, S ;
Villeirs, G ;
Verbaeys, A ;
Oosterlinck, W ;
De Neve, W .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (03) :777-787
[3]   Radiotherapy of prostate cancer with or without intensity modulated beams: A planning comparison [J].
De Meerleer, GO ;
Vakaet, LAML ;
De Gersem, WRT ;
De Wagter, C ;
De Naeyer, B ;
De Neve, W .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 47 (03) :639-648
[4]  
De Neve W, 2002, FRONT RADIAT THER ON, V37, P163
[5]   Clinical delivery of intensity modulated conformal radiotherapy for relapsed or second-primary head and neck cancer using a multileaf collimator with dynamic control [J].
De Neve, W ;
De Gersem, W ;
Derycke, S ;
De Meerleer, G ;
Moerman, M ;
Bate, MT ;
Van Duyse, B ;
Vakaet, L ;
De Deene, Y ;
Mersseman, B ;
De Wagter, C .
RADIOTHERAPY AND ONCOLOGY, 1999, 50 (03) :301-314
[6]   FACTORS INFLUENCING THE DEGREE OF ERYTHEMATOUS SKIN REACTIONS IN HUMANS [J].
DENHAM, JW ;
HAMILTON, CS ;
SIMPSON, SA ;
OSTWALD, PM ;
OBRIEN, M ;
KRON, T ;
JOSEPH, DJ ;
DEAR, KBG .
RADIOTHERAPY AND ONCOLOGY, 1995, 36 (02) :107-120
[7]   The effect of carbon fibre inserts on the build-up and attenuation of high energy photon beams [J].
DeOst, B ;
Vanregemorter, J ;
Schaeken, B ;
VandenWeyngaert, D .
RADIOTHERAPY AND ONCOLOGY, 1997, 45 (03) :275-277
[8]  
Dutreix J, 1969, Ann N Y Acad Sci, V161, P33
[9]  
GILLIS S, IN PRESS INTER CTR
[10]  
HUM JL, 1995, INT J RADIAT ONCOL, V5, P1101