A microstructure based fiberoptic pressure sensor for measurements in lumbar intervertebral discs.

被引:7
作者
Höjer, S [1 ]
Krantz, M [1 ]
Ekström, L [1 ]
Kaigle, A [1 ]
Holm, S [1 ]
机构
[1] Samba Sensors AB, S-41328 Gothenburg, Sweden
来源
BIOMEDICAL SENSORS, FIBERS, AND OPTICAL DELIVERY SYSTEMS, PROCEEDINGS | 1999年 / 3570卷
关键词
micromechanics; fiberoptic; pressure; intervertebral disc;
D O I
10.1117/12.336921
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
A fiberoptic system with a microstructure sensor element was used for measuring lumbar intervertebral disc pressure in a porcine model. The fiberoptic pressure sensor was inserted in the disc using a guiding needle. A reference sensor was also introduced into the same area of an adjacent disc. The fiberoptic sensor detected pressures from 0.7-8 bar in the disc. Dynamic measurements were carried out at frequencies between 2 and 10 Hz. No phase lag was observed between the applied force and the measured pressures. Sensitivity, dynamic response and available pressure range are all important design characteristics for which this fiberoptic sensor has a competitive edge.
引用
收藏
页码:115 / 122
页数:8
相关论文
共 12 条
[1]   THE EFFECT OF POSTURE ON THE ROLE OF THE APOPHYSEAL JOINTS IN RESISTING INTERVERTEBRAL COMPRESSIVE FORCES [J].
ADAMS, MA ;
HUTTON, WC .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1980, 62 (03) :358-362
[2]   THE MECHANICAL FUNCTION OF THE LUMBAR APOPHYSEAL JOINTS [J].
ADAMS, MA ;
HUTTON, WC .
SPINE, 1983, 8 (03) :327-330
[3]  
HOLM S, 1982, CLIN ORTHOP RELAT R, V170, P296
[4]  
HOLM S, 1997, CLIN ORTHOP RELAT R, P178
[5]   HYOMEX - A MINIATURE UNIVERSAL TESTING MACHINE FOR IN-VIVO BIOMECHANICAL STUDIES [J].
HULT, E ;
EKSTROM, L ;
HOLM, S ;
HANSSON, T .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1994, 18 (05) :169-172
[6]   THE DEPENDENCE OF INTERVERTEBRAL-DISK MECHANICAL-PROPERTIES ON PHYSIOLOGICAL CONDITIONS [J].
KELLER, TS ;
HOLM, SH ;
HANSSON, TH ;
SPENGLER, DM .
SPINE, 1990, 15 (08) :751-761
[7]  
KROHN DA, 1992, FIBER OPTIC SENSORS, pCH8
[8]   INTERNAL INTERVERTEBRAL-DISK MECHANICS AS REVEALED BY STRESS PROFILOMETRY [J].
MCNALLY, DS ;
ADAMS, MA .
SPINE, 1992, 17 (01) :66-73
[9]   MECHANICAL-PROPERTIES OF LUMBAR SPINE MOTION SEGMENTS UNDER LARGE LOADS [J].
MILLER, JAA ;
SCHULTZ, AB ;
WARWICK, DN ;
SPENCER, DL .
JOURNAL OF BIOMECHANICS, 1986, 19 (01) :79-84
[10]   DISK PRESSURE MEASUREMENTS [J].
NACHEMSON, AL .
SPINE, 1981, 6 (01) :93-97