Field map estimation with a region growing scheme for iterative 3-point water-fat decomposition

被引:185
作者
Yu, HZ
Reeder, SB
Shimakawa, A
Brittain, JH
Pelc, NJ
机构
[1] Stanford Univ, Dept Elect Engn, Stanford, CA 94305 USA
[2] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[3] HE Healthcare, Appl Sci Lab W, Menlo Pk, CA USA
关键词
water-fat decomposition; field map; region growing; phase ambiguity; IDEAL;
D O I
10.1002/mrm.20654
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Robust fat suppression techniques are required for many clinical applications. Multi-echo water-fat separation methods are relatively insensitive to B-0 field inhomogeneity compared to the fat saturation method. Estimation of this field inhomogeneity, or field map, is an essential and important step, which is well known to have ambiguity. For an iterative water-fat decomposition method recently proposed, ambiguities still exist, but are more complex in nature. They were studied by analytical expressions and simulations. To avoid convergence to incorrect field map solutions, an initial guess closer to the true field map is necessary. This can be achieved using a region growing process, which correlates the estimation among neighboring pixels. Further improvement in stability is achieved using a low-resolution reconstruction to guide the selection of the starting pixels for the region growing. The proposed method was implemented and shown to significantly improve the algorithm's immunity to field inhomogeneity.
引用
收藏
页码:1032 / 1039
页数:8
相关论文
共 18 条
[1]   Two-point dixon technique for water-fat signal decomposition with B-o inhomogeneity correction [J].
Coombs, BD ;
Szumowski, J ;
Coshow, W .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (06) :884-889
[2]   SIMPLE PROTON SPECTROSCOPIC IMAGING [J].
DIXON, WT .
RADIOLOGY, 1984, 153 (01) :189-194
[3]  
Ghiglia D. C., 1998, 2 DIMENSIONAL PHASE
[4]   MULTIPOINT DIXON TECHNIQUE FOR WATER AND FAT PROTON AND SUSCEPTIBILITY IMAGING [J].
GLOVER, GH .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1991, 1 (05) :521-530
[5]   3-POINT DIXON TECHNIQUE FOR TRUE WATER FAT DECOMPOSITION WITH BO INHOMOGENEITY CORRECTION [J].
GLOVER, GH ;
SCHNEIDER, E .
MAGNETIC RESONANCE IN MEDICINE, 1991, 18 (02) :371-383
[6]   RARE IMAGING - A FAST IMAGING METHOD FOR CLINICAL MR [J].
HENNIG, J ;
NAUERTH, A ;
FRIEDBURG, H .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (06) :823-833
[7]   Inhomogeneity correction using an estimated linear field map [J].
Irarrazabal, P ;
Meyer, CH ;
Nishimura, DG ;
Macovski, A .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (02) :278-282
[8]   Real-time interactive MRI on a conventional scanner [J].
Kerr, AB ;
Pauly, JM ;
Hu, BS ;
Li, KC ;
Hardy, CJ ;
Meyer, CH ;
Macovski, A ;
Nishimura, DG .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (03) :355-367
[9]  
Nash SG., 1996, LINEAR NONLINEAR PRO
[10]   Cramer-Rao bounds for three-point decomposition of water and fat [J].
Pineda, AR ;
Reeder, SB ;
Wen, ZF ;
Pelc, NJ .