Reevaluation of intrapartum fetal monitoring using fetal oximetry: A review

被引:4
|
作者
Uchida, Toshiyuki [1 ]
Kanayama, Naohiro [1 ]
Kawai, Kenta [1 ]
Mukai, Mari [1 ]
Suzuki, Kazunao [1 ]
Itoh, Hiroaki [1 ]
Niwayama, Masatsugu [2 ]
机构
[1] Hamamatsu Univ Sch Med, Dept Obstet & Gynecol, Hamamatsu, Shizuoka, Japan
[2] Shizuoka Univ, Dept Elect & Elect Engn, Hamamatsu, Shizuoka, Japan
基金
日本学术振兴会;
关键词
examiner's mounting probe; fetal monitoring; fetal pulse oximetry; near-infrared spectroscopy; tissue oximetry; TISSUE OXYGEN-SATURATION; PULSE OXIMETRY; CONTROLLED-TRIAL; LABOR; COMPLICATION; ACCURACY; DELIVERY; MULTICENTER; MANAGEMENT; 2ND-STAGE;
D O I
10.1111/jog.13761
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Aim Although several studies reported the measurement of fetal oxygen saturation using fetal pulse oximetry (FPO) for evaluation of the fetal intrapartum condition, a systematic review of the seven randomized controlled trials (RCTs) provided no evidence to support FPO for intrapartum fetal monitoring. In the present review, we re-evaluate an overview for the use of FPO and seven RCTs of FPO. Methods Results We reviewed numerous previous reports on FPO and seven RCTs of intrapartum FPO. RCTs were conducted with the main outcome measure being a reduction in the cesarean section rate. The largest trial with 5341 entries failed to show any reduction. The negative result from this RCT may be explained by the use of a different cutoff value for fetal oxygen saturation compared to the other RCT; in addition, there were differences in the indications for cesarean section due to dystocia and in the definition of non-reassuring fetal status (NRFS). An abnormal FPO value, defined as the fetal oxygen saturation value <30% for at least 10 min, is useful for making a diagnosis of fetal acidosis. A newly developed device, an examiner's finger-mounted tissue oximetry, accurately measures tissue oxygen saturation while overcoming the drawbacks of FPO, such as infection risk and slipping off of the sensor during descent of the fetal head. Conclusion FPO (including the new device) with fetal heart rate monitoring in selected cases of NRFS may reduce the cesarean section rate.
引用
收藏
页码:2127 / 2134
页数:8
相关论文
共 50 条
  • [41] PROBLEMS IN THE CLINICAL USE OF INTRAPARTUM FETAL ECG MONITORING
    COCKBURN, JE
    PEARCE, JM
    CHAMBERLAIN, GVP
    JOURNAL OF PERINATAL MEDICINE, 1994, 22 (03) : 195 - 204
  • [42] National survey of routines for intrapartum fetal monitoring in Norway
    Kaasen, Anne
    Aanstad, Kristin J.
    Pay, Aase S. D.
    Okland, Inger
    Blix, Ellen
    ACTA OBSTETRICIA ET GYNECOLOGICA SCANDINAVICA, 2019, 98 (03) : 390 - 395
  • [43] Intrapartum fetal scalp lactate sampling for fetal assessment in the presence of a non-reassuring fetal heart rate trace
    East, Christine E.
    Leader, Leo R.
    Sheehan, Penelope
    Henshall, Naomi E.
    Colditz, Paul B.
    Lau, Rosalind
    COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2015, (05):
  • [44] Fetal pulse oximetry and fetal heart rate monitoring during stage II of labour
    Langer, B
    Carbonne, B
    Goffinet, F
    LeGoueff, F
    Berkane, N
    Laville, M
    EUROPEAN JOURNAL OF OBSTETRICS & GYNECOLOGY AND REPRODUCTIVE BIOLOGY, 1997, 72 : S57 - S61
  • [45] Feasibility of simultaneous application of fetal electrocardiography and fetal pulse oximetry
    Luttkus, AK
    Stupin, JH
    Callsen, TA
    Dudenhausen, JW
    ACTA OBSTETRICIA ET GYNECOLOGICA SCANDINAVICA, 2003, 82 (05) : 443 - 448
  • [46] The fetal admission test and intrapartum fetal death
    Golditch, BD
    Ahn, MO
    Phelan, JP
    AMERICAN JOURNAL OF PERINATOLOGY, 1998, 15 (04) : 273 - 276
  • [47] Fetal pulse oximetry for fetal assessment in labour
    East, Christine E.
    Begg, Lisa
    Colditz, Paul B.
    Lau, Rosalind
    COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2014, (10):
  • [48] Simultaneous use of intrapartum fetal pulse oximetry and amnioinfusion in meconium stained amniotic fluid
    Halvax, L
    Szabó, I
    Vizer, M
    Csermely, T
    Ertl, T
    EUROPEAN JOURNAL OF OBSTETRICS GYNECOLOGY AND REPRODUCTIVE BIOLOGY, 2002, 104 (02): : 105 - 108
  • [49] THE EFFECT OF MECONIUM ON NEONATAL AND FETAL REFLECTANCE PULSE OXIMETRY
    JOHNSON, N
    JOHNSON, VA
    BANNISTER, J
    MCNAMARA, H
    JOURNAL OF PERINATAL MEDICINE, 1990, 18 (05) : 351 - 355
  • [50] Predictive value of pulse oximetry for the development of fetal acidosis
    Nonnenmacher, Andreas
    Hopp, Hartmut
    Dudenhausen, Joachim
    JOURNAL OF PERINATAL MEDICINE, 2010, 38 (01) : 83 - 86