The beat-to-beat relationship between pulse transit time and systolic blood pressure

被引:0
作者
Poon, Carmen C. Y. [1 ]
Zhang, Yuan-Ting [1 ]
Wong, George [2 ]
Poon, Wai Sang [2 ]
机构
[1] Chinese Univ Hong Kong, Dept Elect Engn, Joint Res Ctr Biomed Engn JCBME, Hong Kong, Hong Kong, Peoples R China
[2] Prince Wales Hosp, Div Neurosurg, Hong Hom, Hong Kong, Peoples R China
来源
2008 INTERNATIONAL SPECIAL TOPIC CONFERENCE ON INFORMATION TECHNOLOGY AND APPLICATIONS IN BIOMEDICINE, VOLS 1 AND 2 | 2008年
关键词
wearable medical devices; cuffless blood pressure measurement; pulse transit time; m-health;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Non-invasive measurement of beat-to-beat arterial blood pressure (BP) is an interesting but currently unsolved topic in long-term physiological monitoring by wearable devices and body sensor networks. Although it is suggested that pulse transit time (PTT) is related to and can be used to estimate BP, few studies showed how their relationship changes over time. This study reported the beat-to-beat correlation between PTT and invasive BP of short segments for 4 subjects during an 8-hour period. The results of the study found that the median correlation of beat-to-beat PTT and invasive BP decreased from -0.73 to -0.63 when the number of consecutive bents used to calculate the correlation increased from 15 to 360, suggesting that PTT-based techniques may need frequent calibration in order to estimate beat-to-beat BP.
引用
收藏
页码:357 / +
页数:2
相关论文
共 5 条
  • [1] Continuous estimation of systolic blood pressure using the pulse arrival time and intermittent calibration
    Chen, W
    Kobayashi, T
    Ichikawa, S
    Takeuchi, Y
    Togawa, T
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2000, 38 (05) : 569 - 574
  • [2] An evaluation of a new continuous blood pressure monitoring system in critically ill patients
    Heard, SO
    Lisbon, A
    Toth, I
    Ramasubramanian, R
    [J]. JOURNAL OF CLINICAL ANESTHESIA, 2000, 12 (07) : 509 - 518
  • [3] Cuff-less and noninvasive measurements of arterial blood pressure by pulse transit time
    Poon, C. C. Y.
    Zhang, Y. T.
    [J]. 2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 5877 - 5880
  • [4] THOMAS JG, 1955, J PHYSIOL-LONDON, V129, pP75
  • [5] CLINICAL-EVALUATION OF CONTINUOUS NONINVASIVE BLOOD-PRESSURE MONITORING - ACCURACY AND TRACKING CAPABILITIES
    YOUNG, CC
    MARK, JB
    WHITE, W
    DEBREE, A
    VENDER, JS
    FLEMING, A
    [J]. JOURNAL OF CLINICAL MONITORING, 1995, 11 (04): : 245 - 252