Commissioning of a micro multi-leaf collimator and planning system for stereotactic radiosurgery

被引:138
|
作者
Cosgrove, VP [1 ]
Jahn, U [1 ]
Pfaender, M [1 ]
Bauer, S [1 ]
Budach, V [1 ]
Wurm, RE [1 ]
机构
[1] Free Univ Berlin, Klinikum Charite, Klin Strathlentherapie, D-10117 Berlin, Germany
关键词
micro multi-leaf collimator; MLC; conformal therapy; dosimetry; radiosurgery planning;
D O I
10.1016/S0167-8140(99)00020-1
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: A computer controlled micro multi-leaf collimator, m3 mMLC, has been commissioned for conformal, fixed-held radiosurgery applications. Measurements were made to characterise the basic dosimetric properties of the m3, such as leaf transmission, leakage and beam penumbra. In addition, the geometric and dosimetric accuracy of the m3 was verified when used in conjunction with a BrainSCAN v3.5 stereotactic planning system. Materials and methods: The m3 was detachably mounted to a Varian Clinac 2100C accelerator delivering 6 MV X-rays. Leaf transmission, leakage, penumbra and multiple, conformal fixed field dose distributions were measured using calibrated film in solid water. Beam data were collected using a diamond detector in a scanning water tank and planned dose distributions were verified using LIF TLDs and film. A small, shaped phantom was also constructed to confirm field shaping accuracy using portal images. Results: Mean transmission through the closed multi-leaves was 1.9 +/- 0.1% and leakage between leaves was 2.8 +/- 0.15%. Between opposing leaves abutting along the central beam-axis transmission was similar to 15 +/- 3%, but was reduced to a mean of 4.5 +/- 0.6% by moving the abutmen position 4.5 cm off-axis. Beam penumbrae were effectively constant as a function of increasing square field size and asymmetric fields and was seen to vary non-linearly when shaped to diagonal, straight edges. TMR, OAR and relative output beam data measurements of circular m3 fields were comparable to conventional, circular stereotactic collimators. Multiple, conformal field dose distributions were calculated with good spatial and dosimetric accuracy, with the planned 90% isodose curves agreeing with measurements to within 1-2 mm and to +/- 3% at isocentre. Portal films agreed with planned beams eye-view field shaping to within 1 mm. Conclusions: The m3 micro multi-leaf collimator is a stable, high precision field-shaping device suitable for small-field, radiosurgery applications. Dose distributions can be accurately calculated by a planning system using only a few beam data parameters. (C) 1999 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:325 / 336
页数:12
相关论文
共 50 条
  • [41] Clinical Feasibility of Multi-Leaf Collimator Based Dynamic Wedge
    Jeong, C.
    Kwak, J.
    Alm, S.
    Kim, J.
    Park, J.
    Yoon, S.
    Cho, B.
    MEDICAL PHYSICS, 2015, 42 (06) : 3372 - 3372
  • [42] Integration of the function generation synthesis into the design of a multi-leaf collimator
    Gezgin, Erkin
    Yazici, Mustafa Volkan
    JOURNAL OF THE BRAZILIAN SOCIETY OF MECHANICAL SCIENCES AND ENGINEERING, 2020, 42 (07)
  • [43] Multi-leaf collimator designs: the clinical significance of linear accelerators
    Cakir, Aydin
    Bilge, Hatice
    TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2012, 27 (01): : 46 - 54
  • [44] Preliminary inter-leaf transmission characterization of a neutron multi-leaf collimator
    Farr, J
    Yudelev, M
    Maughan, R
    Forman, J
    Blosser, E
    Horste, T
    MEDICAL PHYSICS, 2002, 29 (06) : 1327 - 1327
  • [45] Comparison of algorithms for static multi-leaf collimator field segmentation
    Ni, Dongdong
    Wu, Xu
    Chen, Jianqi
    Jing, Jia
    Liu, Caixia
    INFORMATION ENGINEERING FOR MECHANICS AND MATERIALS SCIENCE, PTS 1 AND 2, 2011, 80-81 : 1091 - 1095
  • [46] An analytical approach for optimizing the leaf design of a multi-leaf collimator in a linear accelerator
    Topolnjak, R.
    van der Heide, U. A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (11): : 3007 - 3021
  • [47] Automated Analysis of Leaf Position Accuracy for Multi-Leaf Collimator Quality Assurance
    Du, W.
    Casey, K.
    Kudchadker, R.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [48] The effect of multi-leaf collimator leaf width on VMAT treatment plan quality
    Peiris, Gregory Sadharanu
    Whelan, Brendan
    Hardcastle, Nicholas
    Sheehy, Suzie Lynn
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025,
  • [49] Impact of dose rates on the position accuracy of multi-leaf collimator
    Zhang, Yuhai
    Li, Yuemin
    Xia, Huosheng
    Wang, Jie
    RADIATION PHYSICS AND CHEMISTRY, 2012, 81 (12) : 1813 - 1816
  • [50] Characterization of an extendable multi-leaf collimator for clinical electron beams
    O'Shea, Tuathan P.
    Ge, Yuanyuan
    Foley, Mark J.
    Faddegon, Bruce A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (23): : 7621 - 7638