Modifying conventional high-performance liquid chromatography systems to achieve fast separations with Fused-Core columns: A case study

被引:39
|
作者
Alexander, A. J. [1 ]
Waeghe, T. J. [2 ]
Himes, K. W. [1 ]
Tomasella, F. P. [1 ]
Hooker, T. F. [1 ]
机构
[1] Bristol Myers Squibb Co, Analyt Res & Dev, New Brunswick, NJ 08903 USA
[2] MAC MOD Analyt Inc, Chadds Ford, PA 19317 USA
关键词
Fast LC; Fused-Core (R); Extra-column dispersion; Extra-column band broadening; Extra-column volume; MASS-TRANSFER KINETICS; PHARMACEUTICAL ANALYSIS; SILICA PARTICLES; PEAK-CAPACITY; SHELL; HPLC; PRESSURE; LC;
D O I
10.1016/j.chroma.2011.06.026
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The theoretical increase in performance from the use of high efficiency columns with conventional HPLC equipment is generally not observed due to the design limitations of such equipment, particularly with respect to extra-column dispersion (ECD). This study examines the impact of ECD from a Waters Alliance 2695 system on the performance of 2.7 mu m HALO (R) C-18 Fused-Core superficially porous particle columns of various dimensions. The Alliance system was re-configured in different ways to reduce extra-column volume (ECV) and the ECD determined in each case as a function of flow rate up to a maximum of 2 mL/min. The results obtained showed a progressive decrease in ECD as the ECV was reduced, irrespective of the flow rate employed. However, this decrease in ECD was less than theoretically expected for the lower ECV configurations. The inability to reduce the actual extra-column dispersion further was attributed to additional dispersion associated with the design/volume of the auto-injector. This was confirmed by making sample injections with a low dispersion manual injection valve, instead of auto-injection, for the two lowest ECV configurations studied. In each case, the measured and predicted ECD values were in good agreement. The auto-injector module is an integral part of the Alliance 2695 instrument and cannot be easily modified. However, even with autosampler injection, for a 3 mm ID x 100 mm Fused-Core (R) column approximately 70% of the maximum plate count (similar to 84% of the resolution or more) could still be obtained in isocratic separations for solutes with k >= similar to 4.5 when using the lowest ECV configuration. This study also highlights some of the problems inherent in trying to measure accurately the true extra-column dispersion of a chromatographic system and compares the results obtained to those theoretically predicted. Using this same lowest volume instrument configuration, two real-world pharmaceutical methods were scaled to separations that are similar to 3-3.5-fold faster, while still maintaining comparable data quality (resolution and signal-to-noise ratios). (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:5456 / 5469
页数:14
相关论文
共 50 条
  • [31] Comparison of performance of high-performance liquid chromatography columns packed with superficially and fully porous 2.5 μm particles using kinetic plots
    Fanigliulo, Ameriga
    Cabooter, Deirdre
    Bellazzi, Guido
    Tramarin, Davide
    Allieri, Brigida
    Rottigni, Alessandro
    Desmet, Gert
    JOURNAL OF SEPARATION SCIENCE, 2010, 33 (23-24) : 3655 - 3665
  • [32] Application of ultra-performance columns in high-performance liquid chromatography for determination of albendazole and its metabolites in turkeys
    Grabowski, Tomasz
    Jaroszewski, Jerzy Jan
    Swierczewska, Anna
    Sawicka, Renata
    Maslanka, Tomasz
    Markiewicz, Wlodzimierz
    Ziolkowski, Hubert
    BIOMEDICAL CHROMATOGRAPHY, 2011, 25 (10) : 1159 - 1167
  • [33] Fused-core silica column ultra-performance liquid chromatography-ion trap tandem mass spectrometry for determination of global DNA methylation status
    Yang, Ill
    Fortin, Marie C.
    Richardson, Jason R.
    Buckley, Brian
    ANALYTICAL BIOCHEMISTRY, 2011, 409 (01) : 138 - 143
  • [34] On-line coupling of sol-gel-generated immunoaffinity columns with high-performance liquid chromatography
    Cichna, M
    Markl, P
    Knopp, D
    Niessner, R
    JOURNAL OF CHROMATOGRAPHY A, 2001, 919 (01) : 51 - 58
  • [35] Practical assessment of frictional heating effects and thermostat design on the performance of conventional (3 μm and 5 μm) columns in reversed-phase high-performance liquid chromatography
    Fallas, Morgane M.
    Hadley, Mark R.
    McCalley, David V.
    JOURNAL OF CHROMATOGRAPHY A, 2009, 1216 (18) : 3961 - 3969
  • [36] Effect of the mobile phase on antibody-based enantiomer separations of amino acids in high-performance liquid chromatography
    Hofstetter, O
    Lindstrom, H
    Hofstetter, H
    JOURNAL OF CHROMATOGRAPHY A, 2004, 1049 (1-2) : 85 - 95
  • [37] Development and application of a high-performance liquid chromatography method using monolithic columns for the analysis of ecstasy tablets
    Mc Fadden, Kim
    Gillespie, John
    Carney, Brian
    O'Driscoll, Daniel
    JOURNAL OF CHROMATOGRAPHY A, 2006, 1120 (1-2) : 54 - 60
  • [38] A column capacity study of single, serial, and parallel linked rod monolithic high performance liquid chromatography columns
    Gray, MJ
    Slonecker, PJ
    Dennis, G
    Shalliker, RA
    JOURNAL OF CHROMATOGRAPHY A, 2005, 1096 (1-2) : 92 - 100
  • [39] Evaluation of monolithic columns for determination of formaldehyde and acetaldehyde in sugar cane spirits by High-Performance Liquid Chromatography
    Pires Penteado, Jose C.
    Sobral, Ana C.
    Masini, Jorge C.
    ANALYTICAL LETTERS, 2008, 41 (09) : 1674 - 1681
  • [40] Fast Separation of Capsaicinoids from Peppers by Reversed Phase Ultra-Performance Liquid Chromatography: Comparation with Traditional High-Performance Liquid Chromatography Methods
    Barbero, Gerardo F.
    Liazid, Ali
    Ferreiro-Gonzalez, Marta
    Palma, Miguel
    Barroso, Carmelo G.
    INTERNATIONAL JOURNAL OF FOOD PROPERTIES, 2016, 19 (05) : 984 - 992