Modifying conventional high-performance liquid chromatography systems to achieve fast separations with Fused-Core columns: A case study

被引:39
|
作者
Alexander, A. J. [1 ]
Waeghe, T. J. [2 ]
Himes, K. W. [1 ]
Tomasella, F. P. [1 ]
Hooker, T. F. [1 ]
机构
[1] Bristol Myers Squibb Co, Analyt Res & Dev, New Brunswick, NJ 08903 USA
[2] MAC MOD Analyt Inc, Chadds Ford, PA 19317 USA
关键词
Fast LC; Fused-Core (R); Extra-column dispersion; Extra-column band broadening; Extra-column volume; MASS-TRANSFER KINETICS; PHARMACEUTICAL ANALYSIS; SILICA PARTICLES; PEAK-CAPACITY; SHELL; HPLC; PRESSURE; LC;
D O I
10.1016/j.chroma.2011.06.026
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The theoretical increase in performance from the use of high efficiency columns with conventional HPLC equipment is generally not observed due to the design limitations of such equipment, particularly with respect to extra-column dispersion (ECD). This study examines the impact of ECD from a Waters Alliance 2695 system on the performance of 2.7 mu m HALO (R) C-18 Fused-Core superficially porous particle columns of various dimensions. The Alliance system was re-configured in different ways to reduce extra-column volume (ECV) and the ECD determined in each case as a function of flow rate up to a maximum of 2 mL/min. The results obtained showed a progressive decrease in ECD as the ECV was reduced, irrespective of the flow rate employed. However, this decrease in ECD was less than theoretically expected for the lower ECV configurations. The inability to reduce the actual extra-column dispersion further was attributed to additional dispersion associated with the design/volume of the auto-injector. This was confirmed by making sample injections with a low dispersion manual injection valve, instead of auto-injection, for the two lowest ECV configurations studied. In each case, the measured and predicted ECD values were in good agreement. The auto-injector module is an integral part of the Alliance 2695 instrument and cannot be easily modified. However, even with autosampler injection, for a 3 mm ID x 100 mm Fused-Core (R) column approximately 70% of the maximum plate count (similar to 84% of the resolution or more) could still be obtained in isocratic separations for solutes with k >= similar to 4.5 when using the lowest ECV configuration. This study also highlights some of the problems inherent in trying to measure accurately the true extra-column dispersion of a chromatographic system and compares the results obtained to those theoretically predicted. Using this same lowest volume instrument configuration, two real-world pharmaceutical methods were scaled to separations that are similar to 3-3.5-fold faster, while still maintaining comparable data quality (resolution and signal-to-noise ratios). (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:5456 / 5469
页数:14
相关论文
共 50 条
  • [1] Fast high performance liquid chromatography separations for proteomic applications using Fused-Core® silica particles
    Schuster, Stephanie A.
    Boyes, Barry E.
    Wagner, Brian M.
    Kirkland, Joseph J.
    JOURNAL OF CHROMATOGRAPHY A, 2012, 1228 : 232 - 241
  • [2] Fast analysis of curcuminoids from turmeric (Curcuma longa L.) by high-performance liquid chromatography using a fused-core column
    Felipe Osorio-Tobon, J.
    Carvalho, Pedro I. N.
    Fernandez Barbero, Gerardo
    Nogueira, Gislaine Chrystina
    Rostagno, Mauricio Ariel
    de Almeida Meireles, Maria Angela
    FOOD CHEMISTRY, 2016, 200 : 167 - 174
  • [3] Fast analysis of capsaicinoids in Naga Jolokia extracts (Capsicum chinense) by high-performance liquid chromatography using fused core columns
    Stipcovich, Tea
    Barbero, Gerardo F.
    Ferreiro-Gonzalez, Marta
    Palma, Miguel
    Barroso, Carmelo G.
    FOOD CHEMISTRY, 2018, 239 : 217 - 224
  • [4] Monolithic silica columns for high-efficiency separations by high-performance liquid chromatography
    Ishizuka, N
    Kobayashi, H
    Minakuchi, H
    Nakanishi, K
    Hirao, K
    Hosoya, K
    Ikegami, T
    Tanaka, N
    JOURNAL OF CHROMATOGRAPHY A, 2002, 960 (1-2) : 85 - 96
  • [5] Monolithic columns in high-performance liquid chromatography
    Guiochon, Georges
    JOURNAL OF CHROMATOGRAPHY A, 2007, 1168 (1-2) : 101 - 168
  • [6] Comprehensive two-dimensional liquid chromatography separations of pharmaceutical samples using dual Fused-Core columns in the 2nd dimension
    Alexander, Anthony J.
    Ma, Lianjia
    JOURNAL OF CHROMATOGRAPHY A, 2009, 1216 (09) : 1338 - 1345
  • [7] A validated method for the determination of selected phenolics in olive oil using high-performance liquid chromatography with coulometric electrochemical detection and a fused-core column
    Bayram, Banu
    Ozcelik, Beraat
    Schultheiss, Gerhard
    Frank, Jan
    Rimbach, Gerald
    FOOD CHEMISTRY, 2013, 138 (2-3) : 1663 - 1669
  • [8] Fast and simultaneous determination of phenolic compounds and caffeine in teas, mate, instant coffee, soft drink and energetic drink by high-performance liquid chromatography using a fused-core column
    Rostagno, M. A.
    Manchon, N.
    D'Arrigo, M.
    Guillamon, E.
    Villares, A.
    Garcia-Lafuente, A.
    Ramos, A.
    Martinez, J. A.
    ANALYTICA CHIMICA ACTA, 2011, 685 (02) : 204 - 211
  • [9] Core-Shell Columns in High-Performance Liquid Chromatography: Food Analysis Applications
    Preti, Raffaella
    INTERNATIONAL JOURNAL OF ANALYTICAL CHEMISTRY, 2016, 2016
  • [10] Kinetic performance comparison of a capillary monolithic and a fused-core column in micro-scale liquid chromatography
    Dioszegi, Tuende A.
    Raynie, Douglas E.
    JOURNAL OF CHROMATOGRAPHY A, 2012, 1261 : 107 - 112