Effect of acquisition technique on radiation dose and image quality in multidetector row computed tomography coronary angiography with submillimeter collimation

被引:77
作者
Gerber, TC
Stratmann, BP
Kuzo, RS
Kantor, B
Morin, RL
机构
[1] Mayo Clin, Div Cardiovasc Dis, Jacksonville, FL 32224 USA
[2] Mayo Clin, Dept Diagnost Radiol, Jacksonville, FL 32224 USA
[3] Mayo Clin, Div Cardiovasc Dis, Rochester, MN USA
关键词
computed tomography coronary angiography; image quality; multidetector row computed tomography; radiation dose;
D O I
10.1097/01.rli.0000170628.69792.cb
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: We sought to examine effects of tube voltage and current on radiation dose and image quality for minimally invasive coronary angiography with a 16-slice multidetector row computed tomography (MDCT) scanner. Materials and Methods: We scanned the phantom used in the American College of Radiology Computed Tomography Accreditation Program at tube voltages of 80 and 120 kVp at 550, 650, and 750 mAs(eff), with and without a reduction in radiation dose by electrocardiographically (ECG) controlled tube current modulation (ECG pulsing). Results: Without ECG pulsing, the effective dose was 3 to 13 mSv. On average, a 50% increase in tube voltage led to increased radiation dose (215%), contrast-to-noise ratio (150%), and decreased image noise (-48%). On average, a 17% increase in mAs(eff) led to increased radiation dose (17%) and contrast-to-noise ratio (4%) and decreased image noise (-9%). Dose reduction by ECG pulsing (simulated heart rate, 70 beats per minute) was 28%. With ECG pulsing, noise in images reconstructed during ventricular systole was double that in images reconstructed during ventricular diastole. Conclusions: These quantitative findings about the relationships among scan acquisition parameters, radiation dose, and image quality have practical implications for using ECG pulsing to reduce radiation doses in MDCT coronary angiography.
引用
收藏
页码:556 / 563
页数:8
相关论文
共 28 条
  • [1] In-plane coronary arterial motion velocity: Measurement with electron-beam CT
    Achenbach, S
    Ropers, D
    Holle, J
    Muschiol, G
    Daniel, WG
    Moshage, W
    [J]. RADIOLOGY, 2000, 216 (02) : 457 - 463
  • [2] Risk of cancer from diagnostic X-rays:: estimates for the UK and 14 other countries
    Berrington de González, A
    Darby, S
    [J]. LANCET, 2004, 363 (9406) : 345 - 351
  • [3] Bongartz G, 2004, EUROPEAN GUIDELINES
  • [4] Estimated radiation risks potentially associated with full-body CT screening
    Brenner, DJ
    Elliston, CD
    [J]. RADIOLOGY, 2004, 232 (03) : 735 - 738
  • [5] *C RAD CONTR PROGR, 2002, RES REL COMP TOM CT
  • [6] Multisegment and halfscan reconstruction of 16-slice computed tomography for detection of coronary artery stenoses
    Dewey, M
    Laule, M
    Krug, L
    Schnapauff, D
    Rogalla, P
    Rutsch, WF
    Hamm, B
    Lembcke, A
    [J]. INVESTIGATIVE RADIOLOGY, 2004, 39 (04) : 223 - 229
  • [7] Heart rate adaptive optimization of spatial and temporal resolution for electrocardiogram-gated multislice spiral CT of the heart
    Flohr, T
    Ohnesorge, B
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2001, 25 (06) : 907 - 923
  • [8] FLOHR T, 2004, SHIFTING PARADIGMS C
  • [9] GAMMEX RMI, 2004, ACR CT ACCREDITATION, P464
  • [10] Techniques and parameters for estimating radiation exposure and dose in cardiac computed tomography
    Gerber, TC
    Kuzo, RS
    Morin, RL
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2005, 21 (01) : 165 - 176