Design, fabrication and testing of a micro-Venturi tube for fluid manipulation in a microfluidic system

被引:13
作者
Yu, H. [1 ,2 ]
Li, D. [1 ]
Roberts, R. C. [2 ]
Xu, K. [1 ]
Tien, N. C. [2 ]
机构
[1] Tianjin Univ, State Key Lab Precis Measuring Technol & Instrume, Tianjin 300072, Peoples R China
[2] Case Western Reserve Univ, Dept Elect Engn & Comp Sci, Cleveland, OH 44106 USA
基金
中国国家自然科学基金;
关键词
PUMP;
D O I
10.1088/0960-1317/22/3/035010
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this paper, a micro-Venturi tube fabricated with polydimethylsiloxane (PDMS) is studied for interstitial fluid (ISF) transdermal extraction and fluid manipulation in a microfluidic system toward the application of continuous glucose monitoring. The fabrication structure parameters of the PDMS Venturi tube were theoretically analyzed and experimentally validated against the output vacuum efficiency of the Venturi structure. The optimization methods of the Venturi structure were also discussed. In addition, an optimized micro-Venturi structure was proposed and fabricated. A vacuum pressure of less than 86 kPa had been achieved when an external pressure of 240 kPa was applied to this optimized Venturi tube. Both experimental and mathematical results demonstrate the potential applicability of the micro-Venturi tube in ISF transdermal extraction and fluid manipulation.
引用
收藏
页数:9
相关论文
共 13 条
[1]  
Burdick Jonathan, 2005, Diabetes Technol Ther, V7, P448, DOI 10.1089/dia.2005.7.448
[2]  
Koo K, 2007, TRANSDUCERS EUROSENS, P1335
[3]  
Koo K, 2006, WORLD C MED PHYS BIO, V1, P310
[4]  
Koops H, European Patents, Patent No. [EP1403903, 1403903]
[5]   Proposal of a miniaturized Orbitron pump for MEMS applications [J].
Koops, HWP .
Nanotechnology II, 2005, 5838 :38-42
[6]   Transdermal monitoring of glucose and other analytes using ultrasound [J].
Kost, Joseph ;
Mitragotri, Samir ;
Gabbay, Robert A. ;
Pishko, Michael ;
Langer, Robert .
NATURE MEDICINE, 2000, 6 (03) :347-350
[7]   On-chip vacuum generated by a micromachined Knudsen pump [J].
McNamara, S ;
Gianchandani, YB .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2005, 14 (04) :741-746
[8]  
Muntz E P, 2003, NASA TECH BRIEF, V27, P72
[9]   Clinical evaluation of the GlucoWatch® biographer:: a continual, non-invasive glucose monitor for patients with diabetes [J].
Tierney, MJ ;
Tamada, JA ;
Potts, RO ;
Jovanovic, L ;
Garg, S .
BIOSENSORS & BIOELECTRONICS, 2001, 16 (9-12) :621-629
[10]  
Wang Ping M, 2005, Diabetes Technol Ther, V7, P131, DOI 10.1089/dia.2005.7.131