Metal artefact reduction with cone beam CT: an in vitro study

被引:113
作者
Bechara, B. B. [1 ]
Moore, W. S. [1 ]
McMahan, C. A. [2 ]
Noujeim, M. [1 ]
机构
[1] Univ Texas Hlth Sci Ctr San Antonio, Div Oral & Maxillofacial Radiol, San Antonio, TX 78229 USA
[2] Univ Texas Hlth Sci Ctr San Antonio, Dept Pathol, San Antonio, TX 78229 USA
关键词
cone beam computed tomography; artefact; noise; MULTIDETECTOR CT; IMPLANTS; ALGORITHM; HARDWARE; SCANS;
D O I
10.1259/dmfr/80899839
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Background: Metal in a patient's mouth has been shown to cause artefacts that can interfere with the diagnostic quality of cone beam CT. Recently, a manufacturer has made an algorithm and software available which reduces metal streak artefact (Picasso Master 3D (R) machine; Vatech, Hwaseong, Republic of Korea). Objectives: The purpose of this investigation was to determine whether or not the metal artefact reduction algorithm was effective and enhanced the contrast-to-noise ratio. Methods: A phantom was constructed incorporating three metallic beads and three epoxy resin-based bone substitutes to simulate bone next to metal. The phantom was placed in the centre of the field of view and at the periphery. 10 data sets were acquired at 50-90 kVp. The images obtained were analysed using a public domain software ImageJ (NIH Image, Bethesda, MD). Profile lines were used to evaluate grey level changes and area histograms were used to evaluate contrast. The contrast-to-noise ratio was calculated. Results: The metal artefact reduction option reduced grey value variation and increased the contrast-to-noise ratio. The grey value varied least when the phantom was in the middle of the volume and the metal artefact reduction was activated. The image quality improved as the peak kilovoltage increased. Conclusion: Better images of a phantom were obtained when the metal artefact reduction algorithm was used. Dentomaxillofacial Radiology (2012) 41, 248-253. doi: 10.1259/dmfr/80899839
引用
收藏
页码:248 / 253
页数:6
相关论文
共 23 条
[1]  
[Anonymous], 1998, Applied Regression Analysis
[2]   Development of a compact computed tomographic apparatus for dental use [J].
Arai, Y ;
Tammisalo, E ;
Iwai, K ;
Hashimoto, K ;
Shinoda, K .
DENTOMAXILLOFACIAL RADIOLOGY, 1999, 28 (04) :245-248
[3]   Artifacts in CT: Recognition and avoidance [J].
Barrett, JF ;
Keat, N .
RADIOGRAPHICS, 2004, 24 (06) :1679-1691
[4]  
Bianchi S., 2001, DENTOMAXILLOFAC RAD, V30, pS28
[5]  
Chen Quan, 2009, Zhonghua Kou Qiang Yi Xue Za Zhi, V44, P217
[6]   Beam hardening artefacts occur in dental implant scans with the NewTom® cone beam CT but not with the dental 4-row multidetector CT [J].
Draenert, F. G. ;
Coppenrath, E. ;
Herzog, P. ;
Muller, S. ;
Mueller-Lisse, U. G. .
DENTOMAXILLOFACIAL RADIOLOGY, 2007, 36 (04) :198-203
[7]   METALLIC HIP IMPLANTS - CT WITH MULTIPLANAR RECONSTRUCTION [J].
FISHMAN, EK ;
MAGID, D ;
ROBERTSON, DD ;
BROOKER, AF ;
WEISS, P ;
SIEGELMAN, SS .
RADIOLOGY, 1986, 160 (03) :675-681
[8]   Position of the impacted third molar in relation to the mandibular canal. Diagnostic accuracy of cone beam computed tomography compared with panoramic radiography [J].
Ghaeminia, H. ;
Meijer, G. J. ;
Soehardi, A. ;
Borstlap, W. A. ;
Mulder, J. ;
Berge, S. J. .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2009, 38 (09) :964-971
[9]   AN ALGORITHM FOR THE REDUCTION OF METAL CLIP ARTIFACTS IN CT RECONSTRUCTIONS [J].
GLOVER, GH ;
PELC, NJ .
MEDICAL PHYSICS, 1981, 8 (06) :799-807
[10]   CT SCANS THROUGH METAL SCANNING TECHNIQUE VERSUS HARDWARE COMPOSITION [J].
HARAMATI, N ;
STARON, RB ;
MAZELSPERLING, K ;
FREEMAN, K ;
NICKOLOFF, EL ;
BARAX, C ;
FELDMAN, F .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1994, 18 (06) :429-434