A retrieval analysis of explanted Durom metal-on-metal hip arthroplasties

被引:15
作者
Matthies, Ashley K. [1 ]
Henckel, Johann [1 ]
Skinner, John A. [2 ]
Hart, Alister J. [1 ]
机构
[1] Univ London Imperial Coll Sci Technol & Med, Dept Musculoskeletal Surg, London SW7 2AZ, England
[2] Royal Natl Orthopaed Hosp, Stanmore HA7 4LP, Middx, England
关键词
Metal-on-metal; Hip arthroplasty; Retrieval analysis; ACETABULAR COMPONENT; WEAR; REPLACEMENT; FAILURE; DESIGN;
D O I
10.5301/HIP.2011.8885
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Given the recent reports of high failure rates, an improved understanding of the mechanism of failure of large diameter metal-on-metal hip arthroplasties is essential. We present clinical data and tribological analysis of a consecutive series of 74 failed large diameter metal-on-metal hips, comparing the Durom (Zimmer) with the Birmingham hip resurfacing (BHR) (Smith and Nephew). We retrospectively analysed pre-, intra-, and post-operative clinical data and measured the linear wear and component form of the explanted components using a roundness measuring machine. A significantly higher proportion of hips in the Durom group failed as a result of acetabular loosening (p=0.001) and this was supported by evidence of reduced bone in-growth on the backside of the cup. Comparison of roundness measurement revealed that the Durom hip was significantly lower wearing than the BHR (p<0.05) but the Durom femoral components were subject to significantly greater form errors (p<0.001), the pole of the head being flattened by up to 31 microns. Although the Durom hip is low wearing, reduced sphericity of the femoral component may have resulted in equatorial bearing, leading to an increased frictional torque at the cup-bone interface, preventing bone in-growth, and culminating in acetabular loosening. This supports recent clinical findings of high revision rates as a result of acetabular loosening for the Durom metal-on-metal hip system.
引用
收藏
页码:724 / 731
页数:8
相关论文
共 24 条
[1]  
American Society for Testing and Materials (ASTM), F203305 ASTM
[2]  
[Anonymous], 2010, 7 NAT JOINT REG ENGL
[3]  
[Anonymous], 6 NAT JOINT REG ENGL
[4]  
[Anonymous], 5 NAT JOINT REG ENGL
[5]  
Australian Orthopaedic Association National Joint Replacement Registry, 2009, AUSTR ORTH ASS NAT J
[6]   The Durom Large Diameter Head acetabular component EARLY RESULTS WITH A LARGE-DIAMETER METAL-ON-METAL BEARING [J].
Berton, C. ;
Girard, J. ;
Krantz, N. ;
Migaud, H. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2010, 92B (02) :202-208
[7]   Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement [J].
De Haan, R. ;
Pattyn, C. ;
Gill, H. S. ;
Murray, D. W. ;
Campbell, P. A. ;
De Smet, K. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (10) :1291-1297
[8]   A hip joint simulator study of the performance of metal-on-metal joints - Part II: Design [J].
Dowson, D ;
Hardaker, C ;
Flett, M ;
Isaac, GH .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :124-130
[9]   The relationship between the angle of version and rate of wear of retrieved metal-on-metal resurfacings A PROSPECTIVE, CT-BASED STUDY [J].
Hart, A. J. ;
Ilo, K. ;
Underwood, R. ;
Cann, P. ;
Henckel, J. ;
Lewis, A. ;
Cobb, J. ;
Skinner, J. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (03) :315-320
[10]   The painful metal-on-metal hip resurfacing [J].
Hart, A. J. ;
Sabah, S. ;
Henckel, J. ;
Lewis, A. ;
Cobb, J. ;
Sampson, B. ;
Mitchell, A. ;
Skinner, J. A. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2009, 91B (06) :738-744