What is really measured on a d33-meter?

被引:9
作者
Erhart, J
Burianová, L
机构
[1] Tech Univ Liberec, Dept Phys, CZ-46117 Liberec 1, Czech Republic
[2] Tech Univ Liberec, Int Ctr Piezoelect Res, CZ-46117 Liberec, Czech Republic
关键词
piezoelectric properties; PZT;
D O I
10.1016/S0955-2219(01)00030-9
中图分类号
TQ174 [陶瓷工业]; TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Mechanical boundary conditions of PZT sample are not well defined in the measurement process on a d(33)-meter due to the misfit in size or shape of the jaws and of the sample. Result of such d(33) measurement is usually an effective value d(33)(eff). Mechanical stress/strain conditions are simulated for PZT discs/cylinders in a d(33)-meter by the finite element method (ANSYS((R)) software) and an effective d(33)(eff) is calculated and compared to the original input d(33) value as well as to the values measured on a d(33)-meter (type ZJ-3C). Relative d(33)(eff)/d(33) value seems to be more sensitive to the disc aspect ratio than to the PZT material type tested. Although such strong dependence on the aspect ratio has been found in simulations, it is not correlated with experiments. Coefficient d(33)(eff) is not significantly sensitive to the position of a d(33)-meter metallic jaws with respect to the disc center. (C) 2001 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:1413 / 1415
页数:3