Consistency of the estimated target weights and ECW/TBW using BIA after hemodialysis in patients between standing and lying-down positions

被引:8
作者
Choi, Gwangho [1 ]
Yoon, Ho Joong [1 ]
Song, Young Jin [1 ]
Jeong, Hae Min [1 ]
Gu, Jae Eon [1 ]
Han, Miyeun [2 ]
Kim, Seok Hyung [1 ]
Yoon, Jong-Woo [1 ]
Kim, Hyunsuk [1 ]
机构
[1] Hallym Univ, Chuncheon Sacred Heart Hosp, Dept Internal Med, Med Ctr, Chuncheon Si 24253, Gangwon Do, South Korea
[2] Hallym Univ, Hangang Sacred Heart Hosp, Dept Internal Med, Med Ctr, Seoul 07247, South Korea
基金
新加坡国家研究基金会;
关键词
Hemodialysis; Bioimpedance analysis; Estimated target weight; BIOELECTRICAL-IMPEDANCE ANALYSIS; BODY-COMPOSITION; EXTRACELLULAR WATER; DIALYSIS PATIENTS; DRY-WEIGHT; BIOIMPEDANCE; FLUID; OVERHYDRATION; INDEX; RISK;
D O I
10.1186/s12882-022-02737-3
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Background As hemodialysis is administered with the patient lying down, the distribution of body fluid is stable in the lying position, which is why this position is recommended for bioimpedance analysis (BIA). Although the InBody S10 is widely used for hemodialysis patients in the lying position, clinicians must make the measurements in person. In contrast, patients can use the InBody 770 to obtain measurements by themselves in the standing position, which may be more convenient. Therefore, this study compared the measurements of hemodialysis patients' estimated target weight and ECW/TBW obtained lying down using the S10 to those obtained in the standing position using the 770. Methods This study was conducted among maintenance hemodialysis patients at Chuncheon Sacred Heart Hospital in October 2020. Measurements from 56 patients before and after hemodialysis were obtained using the 2 machines. Each (S10 or 770) estimated target weight, both pre- and post-hemodialysis, was considered ideal when the ECW/TBW ratio was 0.380. R-2 was calculated and the Bland-Altman test was performed. Results The patients' median age was 64 years old, and 51% were men. The actual ultrafiltration was 2 kg, and the mean TBW change measured using the InBody devices was 1.5 L (R-2 = 0.718) for the S10 and 1.7 L (R-2 = 0.616) for the 770. The estimated target weight at pre- and post-hemodialysis showed a remarkably high correlation with the patients' actual pre- and post-hemodialysis weight (R-2 > 0.095). The correlation between these measurements (lying vs. standing) before and after hemodialysis was also very close (R-2 = 1.0000). In addition, ECW/TBW had a good correlation (R2 >= 0.970) The Bland-Altman test of dry weight and ECW/TBW yielded similar results. Conclusions This study showed that patients' estimated target weights in the lying position using the InBody S10 device and in the standing position using the InBody 770 device were consistent in both pre- and post-hemodialysis states.
引用
收藏
页数:12
相关论文
共 33 条
  • [1] Bioimpedance Can Solve Problems of Fluid Overload
    Abbas, Samer R.
    Zhu, Fansan
    Levin, Nathan W.
    [J]. JOURNAL OF RENAL NUTRITION, 2015, 25 (02) : 234 - 237
  • [2] Abbas SR, J RENAL NUTR
  • [3] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [4] Validation of InBody 770 bioelectrical impedance analysis compared to a four-compartment model criterion in young adults
    Brewer, Gabrielle J.
    Blue, Malia N. M.
    Hirsch, Katie R.
    Saylor, Hannah E.
    Gould, Lacey M.
    Nelson, Alyson G.
    Smith-Ryan, Abbie E.
    [J]. CLINICAL PHYSIOLOGY AND FUNCTIONAL IMAGING, 2021, 41 (04) : 317 - 325
  • [5] A whole-body model to distinguish excess fluid from the hydration of major body tissues
    Chamney, Paul W.
    Wabel, Peter
    Moissl, Ulrich M.
    Mueller, Manfred J.
    Bosy-Westphal, Anja
    Korth, Oliver
    Fuller, Nigel J.
    [J]. AMERICAN JOURNAL OF CLINICAL NUTRITION, 2007, 85 (01) : 80 - 89
  • [6] Charra B, 1996, NEPHROL DIAL TRANSPL, V11, P16
  • [7] Chung YJ., 2020, SURG METABOLISM NUTR, V11, P53, DOI [10.18858/smn.2020.11.2.53, DOI 10.18858/SMN.2020.11.2.53]
  • [8] Hypoalbuminemia is also a marker of fluid excess determined by bioelectrical impedance parameters in dialysis patients
    Cigarran, Secundino
    Barril, Guillermina
    Cirugeda, Antonio
    Bernis, Carmen
    Aguilera, Abelardo
    Sanz, Paloma
    Herraez, Isabel
    Alegre, Laura
    Selgas, Rafael
    [J]. THERAPEUTIC APHERESIS AND DIALYSIS, 2007, 11 (02) : 114 - 120
  • [9] Value of bioimpedance analysis estimated "dry weight" in maintenance dialysis patients: a systematic review and meta-analysis
    Covic, Adrian
    Ciumanghel, Adi-Ionut
    Siriopol, Dimitrie
    Kanbay, Mehmet
    Dumea, Raluca
    Gavrilovici, Cristina
    Nistor, Ionut
    [J]. INTERNATIONAL UROLOGY AND NEPHROLOGY, 2017, 49 (12) : 2231 - 2245
  • [10] The role of bioimpedance and biomarkers in helping to aid clinical decision-making of volume assessments in dialysis patients
    Davies, Simon J.
    Davenport, Andrew
    [J]. KIDNEY INTERNATIONAL, 2014, 86 (03) : 489 - 496