CT of pediatric vascular stents used to treat congenital heart disease

被引:23
作者
Eichhorn, Joachim G. [1 ,2 ]
Jourdan, Claudia [2 ,3 ]
Hill, Sharon L. [4 ]
Raman, Subha V. [5 ]
Cheatham, John P. [4 ]
Long, Frederick R. [6 ]
机构
[1] Univ Heidelberg, Childrens Hosp, Dept Pediat Cardiol, D-69120 Heidelberg, Germany
[2] Ohio State Univ, Coll Med, Dept Radiol, Columbus, OH 43210 USA
[3] Univ Heidelberg, Childrens Hosp, Dept Pediat Radiol, D-6900 Heidelberg, Germany
[4] Columbus Childrens Hosp, Ctr Heart, Columbus, OH USA
[5] Ohio State Univ, Coll Med, Davis Heart & Lung Res Inst, Div Cardiol, Columbus, OH 43210 USA
[6] Columbus Childrens Hosp, Childrens Radiol Inst, Columbus, OH USA
关键词
congenital heart disease; MDCT; pediatrics; phantom; radiation exposure; stents;
D O I
10.2214/AJR.07.3194
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of our study was to assess the visibility of lumen narrowing of pediatric vascular stents using various CT dose parameters in an in vitro model. MATERIALS AND METHODS. Ten steel stents of varying designs and sizes commonly used in the treatment of congenital heart disease were implanted in polyvinyl chloride ( PVC) tubes and three of the 10 stents were partially obstructed with wax by filling 25% ( mild) to 60% ( moderate) of the lumen with contrast material. On a 64-MDCT scanner, the stents were scanned at tube voltages ( kVp) of 80, 100, and 120 and at tube currents (mA) of 40, 80, 120, and 160. CT measurements of inner-stent diameter, strut thickness, and percent lumen (in-stent) stenoses were compared with biplane fluoroscopy of digital angiography. RESULTS. The stent diameter and percent stenosis on all CT images were consistently smaller than measured on digital angiography but were highly correlated (r = 0.97; p < 0.0001) with improvement as stent diameter increased (93% agreement with digital angiography for 4-mm stent, up to 99% for 25-mm stent; p = 0.001). Moderate stenosis could be assessed better than mild stenosis ( 99% vs 91% agreement with digital angiography; p = 0.003). Increasing exposure settings improved CT correlation of all measurements for mA up to 120 and kVp up to 100 (98.1% agreement). Higher settings did not improve accuracy (93.9% for 160 mA at 120 kVp; p = 0.03). CONCLUSION. CT is feasible to assess lumen narrowing of pediatric vascular stents at a wide range of tube settings. The study suggests that it is possible to lower the radiation exposure settings without loss in image quality or accuracy in detecting in-stent stenoses.
引用
收藏
页码:1241 / 1246
页数:6
相关论文
共 30 条
[1]   Patient-specific dose and radiation risk estimation in pediatric cardiac catheterization [J].
Bacher, K ;
Bogaert, E ;
Lapere, R ;
De Wolf, D ;
Thierens, H .
CIRCULATION, 2005, 111 (01) :83-89
[2]   Dose reduction in pediatric CT: A rational approach [J].
Boone, JM ;
Geraghty, EM ;
Seibert, JA ;
Wootton-Gorges, SL .
RADIOLOGY, 2003, 228 (02) :352-360
[3]   Estimates of the cancer risks from pediatric CT radiation are not theoretical: Comment on "Point/Counterpoint: In x-ray computed tomography, technique factors should be selected appropriate to size. Against the Proposition" [Med. Phys. 28, 1543-1545 (2001)] [J].
Brenner, DJ ;
Elliston, CD ;
Hall, EJ ;
Berdon, WE .
MEDICAL PHYSICS, 2001, 28 (11) :2387-2388
[4]   Usefulness of multislice computed tomography angiography to evaluate intravascular stents and transcatheter occlusion devices in patients with d-transposition of the great arteries after mustard repair [J].
Cook, SC ;
McCarthy, M ;
Daniels, CJ ;
Cheatham, JP ;
Raman, SV .
AMERICAN JOURNAL OF CARDIOLOGY, 2004, 94 (07) :967-969
[5]   Rings, slings and other vascular abnormalities - Ultrafast computed tomography and magnetic resonance angiography in pediatric cardiology [J].
Eichhorn, J ;
Fink, C ;
Delorme, S ;
Ulmer, H .
ZEITSCHRIFT FUR KARDIOLOGIE, 2004, 93 (03) :201-208
[6]   Assessment of in-stent stenosis in small children with congenital heart disease using mufti-detector computed tomography: A validation study [J].
Eichhorn, JG ;
Long, FR ;
Hill, SL ;
O'Donovan, J ;
Chisolm, JL ;
Fernandez, SA ;
Cheatham, JP .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2006, 68 (01) :11-20
[7]   Multislice computed tomography as an adjunct to the management of an in-stent stenosis in an infant with congenital heart disease: Imaging for the future [J].
Eichhorn, JG ;
Long, FR ;
Hill, SL ;
Cheatham, JP .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2006, 67 (03) :477-481
[8]   Conventional and CT angiography in children: dosimetry and dose comparisons [J].
Frush, Donald P. ;
Yoshizumi, Terry .
PEDIATRIC RADIOLOGY, 2006, 36 (Suppl 2) :154-158
[9]   Pediatric thoracic CT angiography [J].
Frush, DP ;
Herlong, JR .
PEDIATRIC RADIOLOGY, 2005, 35 (01) :11-25
[10]   Computed tomography and radiation risks: What pediatric health care providers should know [J].
Frush, DP ;
Donnelly, LF ;
Rosen, NS .
PEDIATRICS, 2003, 112 (04) :951-957