Miniature, high-resolution, quadrupole mass-spectrometer array

被引:80
作者
Orient, OJ [1 ]
Chutjian, A [1 ]
Garkanian, V [1 ]
机构
[1] CALTECH,JET PROP LAB,CTR SPACE MICROELECT TECHNOL,PASADENA,CA 91109
关键词
D O I
10.1063/1.1147947
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A miniature quadrupole mass spectrometer array consisting of 16 rods in a 4x4 array is reported. Each rod is 25 mm in length and 2 mm in diameter. The ionizer is of a miniature Nier-type design, and the detector is a channel-type electron multiplier. Operating frequencies are 5.3, 7.1, and 12.9 MHz. The mass range demonstrated herein is 1-300 u; and the resolution of the system is 0.1-0.5 u (full width at half-maximum), or m/Delta m=600. The present sensitivity is calculated and measured to be approximately 1x10(12) counts/Torr s. (C) 1997 American Institute of Physics.