Background: Balancing techniques in total knee arthroplasty are often based on surgeons' subjective judgment. However, newer technologies have allowed for objective measurements of soft tissue balancing. This study compared the use of sensor technology to the 30-year surgeon experience regarding (1) compartment loads, (2) soft tissue releases, and (3) component rotational alignments. Methods: Patients received either sensor-guided soft tissue balancing (n = 10) or manual gap balancing (n = 12). Wireless, intraoperative sensor tibial inserts were used to measure intracompartmental loads. The surgeon was blinded to values in the manual gap-balancing cohort. In the sensor cohort, the surgeon was unblinded, and implant trials were placed after normal releases were performed to guide further ligament releases after femoral and tibial resections, as needed. Load measurements were taken at 10 degrees, 45 degrees, and 90 degrees. Results: The sensor cohort had lower medial and lateral compartment loading at 10 degrees, 45 degrees, and 90 degrees. The sensor group had lower mean differences in intercompartment loading at 10 degrees (-5.6 vs -51.7 lbs), 45 degrees (-9.8 vs -45.9 lbs), and 90 degrees (-4.3 vs -27 lbs) compared to manually balanced patients. There were 10 additional soft tissue releases in the sensor cohort (2 initial ones before sensor use), compared to 2 releases in the gap-balanced cohort. In the gap-balanced cohort, tibial trays were positioned at a mean 9 degrees external rotation, compared to a mean 1 degrees internal rotation in the sensor-guided cohort. Conclusion: Sensor-balanced total knee arthroplasties provide objective feedback to perform releases and potentially improve knee balancing and rotational alignment. Future work may clarify whether these changes are beneficial for our patients. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:S102 / S105
页数:4
相关论文
共 16 条
[1]  
Anastasiadis A., 2010, Journal of Medical Engineering & Technology, V34, P448, DOI 10.3109/03091902.2010.517898
[2]   A Fresh Look at Soft-Tissue Balancing [J].
Delanois, Ronald E. ;
Elmallah, Randa K. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2015, 97A (20)
[3]   Gap Balancing versus Measured Resection Technique for Total Knee Arthroplasty [J].
Dennis, Douglas A. ;
Komistek, Richard D. ;
Kim, Raymond H. ;
Sharma, Adrija .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (01) :102-107
[4]  
Gustke KA, 2014, BONE JOINT J, V96B, P78, DOI [10.1302/0301-620X.96B11.34339$2.00, 10.1302/0301-620X.96B11.34339]
[5]   Increased satisfaction after total knee replacement using sensor-guided technology [J].
Gustke, K. A. ;
Golladay, G. J. ;
Roche, M. W. ;
Jerry, G. J. ;
Elson, L. C. ;
Anderson, C. R. .
BONE & JOINT JOURNAL, 2014, 96B (10) :1333-1338
[6]  
Gustke KA, 2014, ADV ORTHOP, V628695, P2014
[7]   A New Method for Defining Balance: Promising Short-Term Clinical Outcomes of Sensor-Guided TKA [J].
Gustke, Kenneth A. ;
Golladay, Gregory J. ;
Roche, Martin W. ;
Elson, Leah C. ;
Anderson, Christopher R. .
JOURNAL OF ARTHROPLASTY, 2014, 29 (05) :955-960
[8]   Factors predisposing to claims and compensations for patient injuries following total hip and knee arthroplasty [J].
Jarvelin, Jutta ;
Hakkinen, Unto ;
Rosenqvist, Gunnar ;
Remes, Ville .
ACTA ORTHOPAEDICA, 2012, 83 (02) :190-196
[9]   Can Surgeons Predict What Makes a Good TKA?: Intraoperative Surgeon Impression of TKA Quality Does Not Correlate With Knee Society Scores [J].
Lee, Gwo-Chin ;
Lotke, Paul A. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (01) :159-165
[10]   Ten Year Follow-Up of Gap Balanced, Rotating Platform Total Knee Arthroplasty in Patients Under 60 Years of Age [J].
Lee, Jason H. ;
Barnett, Steven L. ;
Patel, Jay J. ;
Nassif, Nader A. ;
Cummings, Dennis J. ;
Gorab, Robert S. .
JOURNAL OF ARTHROPLASTY, 2016, 31 (01) :132-136