Slow equilibration of reversed-phase columns for the separation of ionized solutes

被引:34
作者
Marchand, DH [1 ]
Williams, LA
Dolan, JW
Snyder, LR
机构
[1] Univ Wisconsin, Dept Chem, River Falls, WI 54022 USA
[2] BASi NW Lab, McMinnville, OR 97128 USA
[3] Rheodyne LLC, LC Resources Grp, Walnut Creek, CA 94596 USA
基金
美国国家卫生研究院;
关键词
column equilibration; ionized solutes;
D O I
10.1016/S0021-9673(03)01216-0
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Reversed-phase columns that have been stored in buffer-free solvents can exhibit pronounced retention-time drift when buffered, low-pH mobile phases are used with ionized solutes. Whereas non-ionized compounds exhibit constant retention times within 20 min of the beginning of mobile phase flow, the retention of ionized compounds can continue to change (by 20% or more) for several hours. If mobile phase pH is changed from low to high and back again, an even longer time may be required before the column reaches equilibration at low pH. The speed of column equilibration for ionized solutes can vary significantly among different reversed-phase columns and is not affected by flow rate. (C) 2003 Elsevier B.V. All rights reserved.