共 5 条
Advanced Morphological 3D Magnetic Resonance Observation of Cartilage Repair Tissue (MOCART) Scoring Using a New Isotropic 3D Proton-Density, Turbo Spin Echo Sequence With Variable Flip Angle Distribution (PD-SPACE) Compared to an Isotropic 3D Steady-State Free Precession Sequence (True-FISP) and Standard 2D Sequences
被引:35
|作者:
Welsch, Goetz H.
[1
,2
]
Zak, Lukas
[3
]
Mamisch, Tallal C.
[4
]
Paul, Dominik
[5
]
Lauer, Lars
[5
]
Mauerer, Andreas
[2
]
Marlovits, Stefan
[3
]
Trattnig, Siegfried
机构:
[1] Med Univ Vienna, Dept Radiol, MR Ctr, High Field MRA 1090, Vienna, Austria
[2] Univ Hosp Erlangen, Dept Trauma Surg, Erlangen, Germany
[3] Med Univ Vienna, Dept Trauma Surg, Ctr Joint & Cartilage, Vienna, Austria
[4] Univ Bern, Dept Orthoped Surg, Bern, Switzerland
[5] Siemens Healthcare, Erlangen, Germany
关键词:
PD-SPACE;
True-FISP;
MOCART;
cartilage repair;
AUTOLOGOUS CHONDROCYTE IMPLANTATION;
ARTICULAR-CARTILAGE;
3.0;
TESLA;
KNEE;
MRI;
DEFECTS;
TRANSPLANTATION;
ABNORMALITIES;
D O I:
10.1002/jmri.22399
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
Purpose: To evaluate a new isotropic 3D proton-density, turbo-spin-echo sequence with variable flip-angle distribution (PD-SPACE) sequence compared to an isotropic 3D true-fast-imaging with steady-state-precession (True-FISP) sequence and 2D standard MR sequences with regard to the new 3D magnetic resonance observation of cartilage repair tissue (MOCART) score. Materials and Methods: Sixty consecutive MR scans on 37 patients (age: 32.8 +/- 7.9 years) after matrix-associated autologous chondrocyte transplantation (MACT) of the knee were prospectively included. The 3D MOCART score was assessed using the standard 2D sequences and the multiplanar-reconstruction (MPR) of both isotropic sequences. Statistical, Bonferroni-corrected correlation as well as subjective quality analysis were performed. Results: The correlation of the different sequences was significant for the variables defect fill, cartilage interface, bone interface, surface, subchondral lamina, chondral osteophytes, and effusion (Pearson coefficients 0.514-0.865). Especially between the standard sequences and the 3D True-FISP sequence, the variables structure, signal intensity, subchondral bone, and bone marrow edema revealed lower, not significant, correlation values (0.242-0.383). Subjective quality was good for all sequences (P >= 0.05). Artifacts were most often visible on the 3D True-FISP sequence (P < 0.05). Conclusion: Different isotropic sequences can be used for the 3D evaluation of cartilage repair with the benefits of isotropic 3D MRI, MPR, and a significantly reduced scan time, where the 3D PD-SPACE sequence reveals the best results.
引用
收藏
页码:180 / 188
页数:9
相关论文